Shiley Disposable Cannula Low Pressure Cuffed Tracheostomy… recallZ-1462-2010 · Nellcor Puritan Bennett Inc. (dba Covidien Ltd) · initiated Apr 13, 2010
Terminated All recalls posted in May 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Nellcor Puritan Bennett Inc. (dba Covidien Ltd) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Inflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.
FDA root cause category: Process design
What the company is doing
Covidien issued an "Urgent Medical Device Recall" notification dated April 13, 2010. Consignees were alerted to the affected device. If an affected product is already in use, the firm recommends that the tube be replaced as soon as the patient's physician determines it is clinically appropriate. All affected product is to be returned to the firm. For further Information, contact Covidien at 1-800-635-5267.
Recall details
- Recall number
- Z-1462-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 13, 2010
- Posted by the FDA
- May 6, 2010
- Recalling firm
- Nellcor Puritan Bennett Inc. (dba Covidien Ltd)
- Quantity
- 19,777 units
- Product code
- JOH
- FDA submissions
- K962173
- Distribution
- Worldwide Distribution -- United States (including Guam and Puerto Rico), Australia, Belgium, Brazil, Canada, Chile, Colombia, Honduras, Japan, Mexico, Panama, Singapore and Uruguay.
Product description
Shiley Disposable Cannula Low Pressure Cuffed Tracheostomy Tube, 4 DCT, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper airway obstructions, provide long term ventilation, support and/or manage tracheal/bronchial secretions.
Code information (models, lots, serial numbers)
Lot Numbers: 0810002426 through 0908000766
Other devices under the same FDA submission
52 devices cleared under K962173 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Shiley 6FENCovidien LP · Tube tracheostomy and tube cuff1 recall106 reports
- Shiley 8DCTCovidien LP · Tube tracheostomy and tube cuff2 deaths104 reports
- Shiley 8DFENCovidien LP · Tube tracheostomy and tube cuff1 recall26 reports
- Shiley 6CFNCovidien LP · Tube tracheostomy and tube cuff67 reports
- Shiley 6DCFSCovidien LP · Tube tracheostomy and tube cuff1 recall16 reports
- Shiley 8FENCovidien LP · Tube tracheostomy and tube cuff2 deaths64 reports
- Shiley 6DFENCovidien LP · Tube tracheostomy and tube cuff1 recall5 reports
- Shiley 10DFENCovidien LP · Tube tracheostomy and tube cuff1 recall4 reports
- Shiley 4DCFSCovidien LP · Tube tracheostomy and tube cuff1 recall2 reports
- Shiley 6DCFNCovidien LP · Tube tracheostomy and tube cuff1 recall1 report
- Shiley 8DCFNCovidien LP · Tube tracheostomy and tube cuff1 recall1 report
- Shiley 10DCFSCovidien LP · Tube tracheostomy and tube cuff1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 55354).
- Z-1463-2010 · TerminatedShiley Disposable Cannula Low Pressure Cuffed Tracheostomy Tube, 6 DCT, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended f…
- Z-1464-2010 · TerminatedShiley Disposable Cannula Low Pressure Cuffed Tracheostomy Tube, 8 DCT, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended f…
- Z-1465-2010 · TerminatedShiley Disposable Cannula Low Pressure Cuffed Tracheostomy Tube, 10 DCT, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended …
- Z-1466-2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 4 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypa…
- Z-1467-2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 6 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypa…
- Z-1468-2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 8 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypa…
- Z-1469-2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 10 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for byp…
- Z-1470-2010 · TerminatedShiley Disposable Cannula Fenestrated Low Pressure Cuffed Tracheostomy Tube, 4 DFEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA…
- Z-1471-2010 · TerminatedShiley Disposable Cannula Fenestrated Low Pressure Cuffed Tracheostomy Tube, 6 DFEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA…
- Z-1472-2010 · TerminatedShiley Disposable Cannula Fenestrated Low Pressure Cuffed Tracheostomy Tube, 8 DFEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA…
- Z-1473-2010 · TerminatedShiley Disposable Cannula Fenestrated Low Pressure Cuffed Tracheostomy Tube, 10 DFEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, C…
- Z-1476-2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 6 FEN-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton,…
- Z-1477-2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 8 FEN-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton,…
- Z-1478-2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube, 4 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper …
- Z-1479-2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube, 6 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper …
- Z-1480-2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube, 8 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper …
- Z-1481-2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube, 10 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper…
- Z-1488-2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 4 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intend…
- Z-1489-2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 6 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intend…
- Z-1490-2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 8 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intend…
- Z-1491-2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 10 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Inten…
- Z-1492-2010 · TerminatedShiley Cuffed Pediatric Tracheostomy Tube, 4.0 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper a…
- Z-1493-2010 · TerminatedShiley Cuffed Pediatric Tracheostomy Tube, 4.5 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper a…
- Z-1494-2010 · TerminatedShiley Cuffed Pediatric Tracheostomy Tube, 5.0 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper a…
- Z-1495-2010 · TerminatedShiley Cuffed Pediatric Tracheostomy Tube, 5.5 PDC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper a…
- Z-1496-2010 · TerminatedShiley Cuffed Long Pediatric Tracheostomy Tube, 5.0 PLC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of up…
- Z-1497-2010 · TerminatedShiley Cuffed Long Pediatric Tracheostomy Tube, 5.5 PLC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of up…
- Z-1498-2010 · TerminatedShiley Cuffed Long Pediatric Tracheostomy Tube, 6.0 PLC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of up…
- Z-1499-2010 · TerminatedShiley Cuffed Long Pediatric Tracheostomy Tube, 6.5 PLC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of up…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page