VNS Therapy Demipulse Generator, Model 103 and recallZ-1591-2010 · Cyberonics, Inc · initiated Apr 6, 2010

Terminated All recalls posted in May 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cyberonics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Battery life projection is inaccurate.

FDA root cause category: Software design

What the company is doing

Safety Alert sent April 6, 2010 to all physicians of record for Model 103 or Model 104 generators and to all hospitals with non-implanted affected product, that explains the event, risks presented and recommended actions to take. The generator can be reprogrammed by the physician the next time the patient is seen in the clinic. The issue can also be corrected during the initial implant procedure with the instructions provided. Additional questions should be directed to the firm; United States customers can contact Clinical Technical Support at 866-882-8804.

Recall details

Recall number
Z-1591-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 6, 2010
Posted by the FDA
May 10, 2010
Recalling firm
Cyberonics, Inc
Quantity
Model103: 8109 units; Model 104: 1722 units; Programmers: 8891
Product code
LYJ
FDA submissions
P970003
Distribution
Worldwide distribution.

Product description

VNS Therapy Demipulse Generator, Model 103 and VNS Therapy Demipulse Duo Generator, Model 104 VNS Therapy Programming System, Model 250 Version 7.0 and 7.1 Software

Code information (models, lots, serial numbers)

Model 103, VNS Therapy Demipulse Generator Model 104, VNS Therapy Demipulse Duo Generator Model 250, Programming System (version 7.0 or 7.1)