Why it was recalled
The firm has received complaints that the saddle is too long. The saddle is supposed to be between 7/8 and 1 1/8 inches long. The saddle in the complaints was up to 1.9 inches long.
FDA root cause category: Process control
What the company is doing
Biomet Sports Medicine issued an "Urgent Medical Device Recall Notice" dated March 31, 2010 to user facilities and distributors. A second notification containing updated information was issued June 2, 2010. Consignees were instructed to discontinue use of the product and return it to the firm along with a completed Response Form. For further information, contact Biomet at 1-574-371-3755 or 1-574-371-3756, Monday through Friday, 8AM to 5PM.
Recall details
- Recall number
- Z-1952-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 31, 2010
- Posted by the FDA
- Jul 1, 2010
- Recalling firm
- Biomet, Inc.
- Quantity
- 2466 both products
- Product code
- MBI
- FDA submissions
- K083070
- Distribution
- Worldwide Distribution -- United States, Canada, Australia, The Netherlands, Belgium, Italy, France, New Zealand, Austria, United Kingdom, Chile, Greece, Poland, Denmark, Turkey, Portugal, South Africa and Spain.
Product description
Artificial ligament fixation device # 7PE ZIPLOOP Extended Toggle, Sterile, REF 904755, Biomet Sports Medicine Warsaw, IN. Intended for soft tissue to bone fixation.
Code information (models, lots, serial numbers)
Lot Numbers: 047510, 050400, 050420, 089710, 089800, 182820, 192240, 192250, 192280, 219260, 219270, 219280, 228530, 228550, 228580, 228590, 228610, 228620, 295450, 361020, 361020, 392930, 392990, 483490, 483500, 529380, 527780, 529800, 529810, 582890, 582910, 627510, 648470, 648480, 671130, 671150, 671170, 671180, 671190, 671240, 706150, 706160, 782360, 782370, 821710, 821730, 943000, 943010, 943030, 952040, 952070 and 974140.
Other devices under the same FDA submission
9 devices cleared under K083070 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ZIPTIGHT FIXATION SYSTEM 904759Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 recall23 reports
- ZIPTIGHT 904834Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 recall9 reports
- ToggleLoc With Ziploop 904755Biomet Orthopedics, LLC · STAPLE, FIXATION, BONE3 reports
- ToggleLoc With Ziploop 904754Biomet Orthopedics, LLC · STAPLE, FIXATION, BONE3 reports
- ZIPTIGHT FIXATION SYSTEM 909856Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE3 reports
- ToggleLoc With Ziploop 904756Biomet Orthopedics, LLC · STAPLE, FIXATION, BONE1 report
- ToggleLoc with Ziploop 909854Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 report
- ZIPTIGHT FIXATION SYSTEM 909857Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 report
- ZIPTIGHT 904835Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 55776).
- Z-1953-2010 · TerminatedToggleloc Artificial ligament fixation device ZIPLOOP Technology Implant Kit, Sterile, REF 909848, Biomet Sports Medicine Warsaw, IN. Intended for soft tissue to bone fixation.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page