Toggleloc Artificial ligament fixation device ZIPLOOP… recallZ-1953-2010 · Biomet, Inc. · initiated Mar 31, 2010
Terminated All recalls posted in July 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm has received complaints that the saddle is too long. The saddle is supposed to be between 7/8 and 1 1/8 inches long. The saddle in the complaints was up to 1.9 inches long.
FDA root cause category: Process control
What the company is doing
Biomet Sports Medicine issued an "Urgent Medical Device Recall Notice" dated March 31, 2010 to user facilities and distributors. A second notification containing updated information was issued June 2, 2010. Consignees were instructed to discontinue use of the product and return it to the firm along with a completed Response Form. For further information, contact Biomet at 1-574-371-3755 or 1-574-371-3756, Monday through Friday, 8AM to 5PM.
Recall details
- Recall number
- Z-1953-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 31, 2010
- Posted by the FDA
- Jul 1, 2010
- Recalling firm
- Biomet, Inc.
- Quantity
- 2466 both products
- Product code
- MBI
- FDA submissions
- K083070
- Distribution
- Worldwide Distribution -- United States, Canada, Australia, The Netherlands, Belgium, Italy, France, New Zealand, Austria, United Kingdom, Chile, Greece, Poland, Denmark, Turkey, Portugal, South Africa and Spain.
Product description
Toggleloc Artificial ligament fixation device ZIPLOOP Technology Implant Kit, Sterile, REF 909848, Biomet Sports Medicine Warsaw, IN. Intended for soft tissue to bone fixation.
Code information (models, lots, serial numbers)
Lot Numbers: 228150, 228160, 783930, 975220, 975230 and 975240.
Other devices under the same FDA submission
9 devices cleared under K083070 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ZIPTIGHT FIXATION SYSTEM 904759Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 recall23 reports
- ZIPTIGHT 904834Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 recall9 reports
- ToggleLoc With Ziploop 904755Biomet Orthopedics, LLC · STAPLE, FIXATION, BONE3 reports
- ToggleLoc With Ziploop 904754Biomet Orthopedics, LLC · STAPLE, FIXATION, BONE3 reports
- ZIPTIGHT FIXATION SYSTEM 909856Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE3 reports
- ToggleLoc With Ziploop 904756Biomet Orthopedics, LLC · STAPLE, FIXATION, BONE1 report
- ToggleLoc with Ziploop 909854Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 report
- ZIPTIGHT FIXATION SYSTEM 909857Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 report
- ZIPTIGHT 904835Biomet Orthopedics, LLC · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 55776).
- Z-1952-2010 · TerminatedArtificial ligament fixation device # 7PE ZIPLOOP Extended Toggle, Sterile, REF 904755, Biomet Sports Medicine Warsaw, IN. Intended for soft tissue to bone fixation.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page