Trilogy 100 Ventilator, Model/Catalog Number 1054260… recallZ-1896-2010 · Respironics, Inc. · initiated Jun 10, 2010
Terminated All recalls posted in June 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Respironics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It is possible that the battery could electrically short, resulting in a potential for fire or a thermal event.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Philips Respironics has contacted all of its consignees by phone and instructed them to remove the potentially affected external battery packs from their inventory and any that may be on their patients/end users. These consignees have been instructed to quarantine batteries they have removed away from flammable materials. Product return/replacement instructions were provided. For further information, contact Philips Respironics at 1-877-387-3311.
Recall details
- Recall number
- Z-1896-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jun 10, 2010
- Posted by the FDA
- Jun 24, 2010
- Recalling firm
- Respironics, Inc.
- Quantity
- 23 units
- Product code
- CBK
- FDA submissions
- K083526
- Distribution
- Worldwide Distribution -- United States (MT, IL, IA, NY, NC, TX, CA and FL), France, South Korea and Turkey.
Product description
Trilogy 100 Ventilator, Model/Catalog Number 1054260, U1054260 (Domestic). Philips Respironics, Monroeville, PA 15146. Provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
Code information (models, lots, serial numbers)
Serial Numbers: TV110051146, TV110051147, TV110051347, TV110051335, TV110050717, TV110052245, TV110051033, TV110051130, TV110050733, TV110051133, TV110051336, TV110051144, TV110051241, TV110051236, TV110051344, TV110051333, TV110051338, TV110051340, TV110051352, TV110051348
Other devices under the same FDA submission
23 devices cleared under K083526 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Trilogy100 Ventilator, InternationalRespironics, Inc. · Oximeter8 deaths2,493 reports
- Trilogy 00Respironics, Inc. · CONTINUOUS, VENTILATOR, HOME USE15 deaths2,059 reports
- Trilogy Trilogy100 Ventilator, GermanyRespironics, Inc. · Oximeter696 reports
- Trilogy Trilogy 100 Ventilator, ChinaRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE1 death447 reports
- Trilogy100 Ventilator, Japan-RentalRespironics, Inc. · Medical device data system2 deaths334 reports
- Trilogy Trilogy100 Ventilator, Latin AmericaRespironics, Inc. · Oximeter154 reports
- Trilogy 100 Trilogy100 Ventilator, Bluetooth, ItalyRespironics, Inc. · Medical device data system146 reports
- Trilogy 00Respironics, Inc. · CONTINUOUS, VENTILATOR, HOME USE1 death135 reports
- Trilogy Trilogy 100 Ventilator, BrazilRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE2 deaths127 reports
- Trilogy 00Respironics, Inc. · Oximeter4 deaths119 reports
- Trilogy Trilogy100 Ventilator, AustraliaRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE97 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.