Skip to content

Vacuette 2 ml Lithium Heparin, Collection Tube… recallZ-0623-2011 · Greiner Bio-One North America, Inc. · initiated Aug 25, 2010

Terminated All recalls posted in December 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Greiner Bio-One North America, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Lithium heparin solution was contaminated by EDTA K (potassium) solution due to an operator error and a misinterpretation of a test result. Contaminated tubes will produce inaccurate test results.

FDA root cause category: Other

What the company is doing

Greiner Bio-One instructed customers to stop distribution and to complete an attached Product Disposition form and return it via fax to Greiner Bio-One. They were further instructed to destroy any product they may have. Distributors were instructed provide a list of their customers that have received the affected item via Excel file so that they can be notified of the recall.

Recall details

Recall number
Z-0623-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 25, 2010
Posted by the FDA
Dec 16, 2010
Recalling firm
Greiner Bio-One North America, Inc.
Quantity
18 cases (total 21,600 pieces)
Product code
JKA
FDA submissions
K960857
Distribution
Nationwide Distribution: USA, including the states of CA, FL, IL, IN, KS, KY, ND, NE, NH, OH, SC, SD and WI.

Product description

Vacuette 2 ml Lithium Heparin, Collection Tube, Manufactured by: Greiner-Bio One North America, Inc., 4238 Capital Drive, Monroe, NC. Venous blood collection tubes.

Code information (models, lots, serial numbers)

Item number: 454237, Lot number: B061006

Other devices under the same FDA submission

3 devices cleared under K960857 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.