Dimension Vista V-LYTE Standard A, Catalog No K820 recallZ-0139-2011 · Siemens Healthcare Diagnostics, Inc. · initiated Aug 19, 2010

Terminated All recalls posted in October 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

High potassium content in the affected lots results in the potential for low potassium recovery for Dilution Check, QC and patient samples when run on the Dimension Vista System.

FDA root cause category: Other

What the company is doing

Siemens sent an Urgent Field Safety Notice (dated "August 2010") to customers identifying the product, the problem and the action to be taken. Customers were instructed on how to continue testing until new lots of the recalled product were available and asked to complete and fax a Field Correction Effectiveness Check to 301-631-8467 to confirm receipt of the Urgent Field Safety Notice. For technical questions or concerns customers were instructed to contact the Siemens Technical Solutions Center at 800-441-9250.

Recall details

Recall number
Z-0139-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 19, 2010
Posted by the FDA
Oct 25, 2010
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
3,706 cartons
Product code
CEM
FDA submissions
K051087
Distribution
USA, Nationwide Distribution

Product description

Dimension Vista V-LYTE Standard A, Catalog No K820 For the calibration of Na+/K+/Cl- on the Dimension vista System.

Code information (models, lots, serial numbers)

KA0C03 EXP 3/24/11, KA0D01 EXP 4/14/11, KA0E01 EXP 5/5/11, KA0E02 EXP 5/19/11, KA0F01 EXP 6/3/11, AND KA0G01 EXP 7/8/11

Other devices under the same FDA submission

4 devices cleared under K051087 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.