ZIMMER TRABECULAR METAL" SHOULDER INSTRUMENTATION, DISTAL… recallZ-0571-2011 · Zimmer Inc. · initiated Sep 10, 2010
Terminated All recalls posted in December 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The results of an investigation into complaints revealed that Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Humeral implants engaging in the canal earlier than expected could lead to incomplete seating of the implant, delay in surgery, humeral fracture, or the need for cement in a press-fit application.
FDA root cause category: Device Design
What the company is doing
Zimmer, Inc. sent an URGENT DEVICE CORRECTION letter dated September 10, 2010 to surgeons, risk managers, and distributors. The letter identified the product, the problem, and the action to be taken. Sales Representatives were instructed to: a. provide the Dear Surgeon (Urgent Device Correction) letter and revised surgical technique directly to surgeons and risk managers at facilities using the device or b. provide a list of facilities and surgeons using the TM Humeral and TM Reverse shoulder systems to the Zimmer Recall Coordinator, and c. destroy previous versions of the surgical technique, and d. inform users of when the new surgical technique, instrument case, distal reamers, and distal pilots will be shipped to them based on communication from Zimmer, Inc. Handling of returned product: HCPs were notified to return all of the distal pilots in the scope back to Zimmer which were to be be destroyed once the recall has been completed. They will also be instructed to either return the reamers back to Zimmer or to remove these reamers from the TM Reverse shoulder systems sets and place into the Bigliani/Flatow Complete Shoulder Solution set. New surgical techniques will also be introduced and the user will be asked to remove previous versions of the surgical technique from all points of use. This removal will be verified by Zimmer Sales Representatives. For questions regarding this recall call (574) 371-8852.
Recall details
- Recall number
- Z-0571-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Sep 10, 2010
- Posted by the FDA
- Dec 9, 2010
- Recalling firm
- Zimmer Inc.
- Quantity
- 9,790 total distal pilots in distribution.
- Product code
- KWT
- FDA submissions
- K050761, K052906, K060704
- Distribution
- Worldwide Distribution - USA, Nationwide, Switzerland and Eschbach, Germany, Switzerland, Australia, China, India, Japan, Korea, Malaysia, Singapore, Taiwan, Brazil, Venezuela, El Salvador, Colombia, Canada, and Thailand.
Product description
ZIMMER TRABECULAR METAL" SHOULDER INSTRUMENTATION, DISTAL PILOT used for implantation of the Zimmer Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Shoulder Systems: Catalog Number 00-4310-008/014-17. The distal pilot manual surgical instrument is used in the surgical procedure to implant the Zimmer Trabecular Metal Humeral and Zimmer Trabecular Metal Reverse Shoulder Systems. The distal pilot is attached to the proximal reamer during proximal reaming and to the proximal provisional during trialing. In both of these cases the distal pilot is used to provide alignment for the proximal devices. The Zimmer Trabecular Metal Shoulder System consists of conventional and reverse, semi- and nonconstrained (resurfacing) shoulder prostheses for total or hemi-arthroplasty applications. The Zimmer Trabecular Metal Reverse Shoulder System is a treatment option for patients with severe rotator cuff (RC) deficiency. The stem was designed so that its bearing surfaces, as compared to a traditional humeral stem, is such that the humeral component contains a concave ultra-high molecular weight polyethylene (UHMWPE) liner that articulates with a hemispheric metallic glenoid component. These two systems are used for: - the treatment of severe pain or significant disability in degenerative, - rheumatoid, or traumatic disease of the glenohumeral joint; - ununited humeral head fractures of long duration; - irreducible 3-and 4-part proximal humeral fractures; - avascular necrosis of the humeral head, or other difficult clinical management problems where arthrodesis or resectional arthroplasty is not acceptable.
Code information (models, lots, serial numbers)
Catalog Number 00-4310-008/014-17 All lot numbers.
Other devices under the same FDA submission
34 devices cleared under K050761, K052906, K060704 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trabecular Metal® 00-4349-036-11Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED17 reports
- Trabecular Metal® 00-4349-012-13Zimmer, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED14 reports
- Trabecular Metal® 00-4349-036-03Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED9 reports
- Trabecular Metal® 00-4349-014-13Zimmer, Inc. · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED8 reports
- Trabecular Metal® 00-4349-036-00Zimmer, Inc. · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED7 reports
- Trabecular Metal® 00-4349-066-06Zimmer, Inc. · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED7 reports
- Trabecular Metal® 00-4349-036-06Zimmer, Inc. · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED6 reports
- Trabecular Metal® 00-4349-040-11Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED6 reports
- Trabecular Metal® 00-4309-029-00Zimmer, Inc. · PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED5 reports
- Trabecular Metal® 00-4349-065-06Zimmer, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED5 reports
- Trabecular Metal™ 00-4309-041-00Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED4 reports
- Trabecular Metal® 00-4349-039-12Zimmer, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 56731).
- Z-0570-2011 · TerminatedZIMMER TRABECULAR METAL SHOULDER INSTRUMENTATION, DISTAL PILOT used for implantation of the Zimmer Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Shoulder Systems:…
- Z-0573-2011 · TerminatedTRABECULAR METAL" SHOULDER REAMER CASE used for implantation of the Zimmer Trabecular Metal Humeral and/or Zimmer Trabecular Metal Reverse Shoulder Systems: Catalog Number 00-4312-…
- Z-0574-2011 · TerminatedZimmer Trabecular Metal Humeral Stem Surgical Technique. Catalog Number: 97-4309-002-00, Rev. 3 The Zimmer Trabecular Metal Shoulder System consists of conventional and reverse, s…
- Z-0575-2011 · TerminatedZimmer Trabecular Metal Humeral Stem 4-Part Fracture Surgical Technique. Catalog Number 97-4309-004-00, Rev. 3. The Zimmer Trabecular Metal Shoulder System consists of convention…
- Z-0576-2011 · TerminatedZimmer Trabecular Metal Reverse Shoulder System Surgical Technique. Catalog Number 97-4309-003-00, Rev. 4. The Zimmer Trabecular Metal Reverse Shoulder System is a treatment option…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page