i-STAT Analyzer recallZ-0593-2011 · Abbott Point Of Care Inc. · initiated Sep 20, 2010
Terminated All recalls posted in December 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Point Of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
i-STAT Analyzers batteries may overheat due to a circuitry failure, resulting in the analyzer becoming uncomfortably hot in the area of the battery component.
FDA root cause category: Other
What the company is doing
Abbott Point of Care issued Urgent Recall Notice letters dated September 20, 2010 to customers. The letter informs users to of the possibility of the unit becoming hot to touch, to remove the unit if this is the case, not to remove batteries unless instructed to do by APOC and to forward notification of the recall to customers if it has been further distributed. Abbott Point of Care can be contacted at 800 366-8020, option 1, about this action. Recall extended in May 2011 to included distribution dates of 9/20/2010-6/6/2011.
Recall details
- Recall number
- Z-0593-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Sep 20, 2010
- Posted by the FDA
- Dec 10, 2010
- Recalling firm
- Abbott Point Of Care Inc.
- Quantity
- 06F16-05: 45; 06F16-10: 22289; 310000: 4393; 310001: 1139; 310002: 11489.
- Product code
- CHL
- FDA submissions
- K940918
- Distribution
- Worldwide Distrbution: Throughout USA and the countries of Argentina, Australia, Austria, Belgium.Bolivia, Bosnia-Herz, Canada. Croatia, Chile, Colombia, Costa Rica, Denmark, Ecuador, El Salvador, Finland, Georgian Republic, Greece, Honduras, France, Germany, Italy, Netherlands, New Zealand, Hong Kong, India, Ireland, Israel, Japan, Kingdom of Saudi Arabia, Korea, Kuwait, Mauritius, Mexico, Nigeria, Norway, Panama, Paraguay, Peoples Republic of China, Peru, Poland, Portugal, Qatar, Russia, Singapore, South Africa, Spain, Sweden, UK, Sultanate of Oman, Switzerland, Taiwan, R.O.C., Thailand, Turkey, UAE, Uruguay, and Yemen.
Product description
i-STAT Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. the 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2 and the i-stat portable analyzer as part of the i-STAT system, are intended for use in the in vitro quantization of arterial, venous or capillary whole blood.
Code information (models, lots, serial numbers)
Product numbers: 06F16-05, 06F16-10, 310000, 310001, 310002; all lots
Other devices under the same FDA submission
5 devices cleared under K940918 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- i-STAT 03P88-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH105 reports
- i-STAT 03P76-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall46 reports
- i-STAT 03P77-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall10 reports
- i-STAT 03P79-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH23 reports
- i-STAT 03P78-25Abbott Point Of Care Inc. · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH16 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 56832).
- Z-0594-2011 · TerminatedFUSO Analyzer 300F Abbott Point of Care, Princeton, NJ 08540 Model 200. the 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2 and the i…
- Z-0595-2011 · Terminatedi-STAT1 Abaxis Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2 and …
- Z-0596-2011 · TerminatedMilitary Kit i-STAT1; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2 and the…
- Z-0597-2011 · Terminatedi-STAT Portable Clinical Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2…
- Z-0598-2011 · TerminatedMilitary Kit Portable Clinical Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, p…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.