Skip to content

LeMaitre Vascular AnastoClip GC Vessel Closure… recallZ-0164-2011 · Lemaitre Vascular, Inc. · initiated Sep 22, 2010

Terminated All recalls posted in October 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Lemaitre Vascular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Mislabeled size on external packaging: The contents section listed on the Tyvek lid labels of the Medium and Large sizes incorrectly identified the size of the clip as XL, while the device inside was either a Medium or Large size.

FDA root cause category: Packaging change control

What the company is doing

LeMaitre Vascular issued an Urgent AnastoClip GC Device Field Safety Notice letter dated Sepember 28, 2010 identifying the affected devices and actions to be taken by customers. The letter contains a form that is requested to be returned to LeMaitre Vascular as record of notification and reconciliation. The letter provides an option for the notified hospital to return the affected product for replacement if unused or if they simply wish to have it corrected. Customers can contact LeMaitre Customer Service for replacement at 800 628-9470.

Recall details

Recall number
Z-0164-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 22, 2010
Posted by the FDA
Oct 28, 2010
Recalling firm
Lemaitre Vascular, Inc.
Product code
HBT
FDA submissions
K091987
Distribution
Nationwide Distribution: to the states of CA, FL, GA, IL, IN, LA, MD, MS, NJ, OK, OR, and TX.

Product description

LeMaitre Vascular AnastoClip GC Vessel Closure System-Medium Size Models: 4007-02 Medium size 3 Pack 4008-02 Medium size 10 Pack 4007-06 Medium size 1 Pack Manufactured by LeMatire Vascular Inc, Burlington, MA 01803

Code information (models, lots, serial numbers)

Lot Number Expiration Date  AGC1015 2012-10 AGC1016 2012-10 AGC1017 2012-10 AGC1018 2012-11 AGC1019 2012-11 AGC1022 2012-11 AGC1023 2012-11 AGC1024 2012-12 AGC1025 2013-01 AGC1027 2013-01 AGC1028 2013-01

Other devices under the same FDA submission

2 devices cleared under K091987 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.