AnastoClip GC 8cm (Medium) 4008-06Recalls & reported problems
Lemaitre Vascular, Inc. · AnastoClip GC 8cm (Medium) · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00840663109579), received between Mar 20, 2022 and Aug 6, 2025 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (4 reports).
2 other recalls cover devices cleared under the same FDA submission (K091987); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Form Staple4
- Loose or Intermittent Connection2
- Misfire2
- Difficult to Open or Close1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Infiltration into Tissue1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryAug 6, 2025Failure to Form Staple; Mechanical Jam; Failure to Form Staple; Mechanical Jam
IT WAS REPORTED THAT DURING A BVT PROCEDURE, THE ANASTOCLIP GC CLIP APPLIER JAMMED AND STUCK ONTO THE PATIENT'S ARM AND COULD NOT GET IT TO RELEASE THE CLIP OR OFF THE PATIENT'S ARM. THEY HAD TO CUT OUT PART OF THE VEIN. DEVICE WAS CHECKED PRIOR TO USE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE OPERATION. NO OTHER ISSUES OR INJURY WERE REPORTED TO THE PATIENT.
Patient problems: Infiltration into Tissue
MAUDE report 22729546 ↗MalfunctionFeb 19, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT THE ANASTOCLIP GC CLIPS KEEPS MISFIRING. THE CLIP DEPLOYS WITHOUT COMPLETELY CLOSING. THE PRODUCT WAS NOT DIRECTLY USED ON THE PATIENT AND NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21417808 ↗MalfunctionFeb 19, 2025Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT THE ANASTOCLIP GC CLIPS WERE MISFIRING. THE CLIP DEPLOYS WITHOUT COMPLETELY CLOSING SO THEY FALL OFF. DEVICE WAS NOT DIRECTLY USED ON PATIENT. NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21417759 ↗MalfunctionAug 21, 2024Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT THE ANASTOCLIP GC CLIPS WERE NOT CLOSING AND GRIPPING PROPERLY WHEN USED DURING THE CASE. IT HAD TO BE REMOVED THEM FROM THE DURA DURING A TETHERED CORD PROCEDURE. NO INJURY WAS REPORTED TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20021675 ↗MalfunctionMay 12, 2023Loose or Intermittent Connection; Misfire; Loose or Intermittent Connection; Misfire
IT WAS REPORTED THAT THE ANASTOCLIP GC MEDIUM CLIPS WERE MISFIRING AND THEN COMING OUT OF THE VESSEL HANGING/LOOSE AFTER BEING DEPLOYED. THERE WERE NO PATIENT INJURY. TWO DEVICES WERE INVOLVED IN THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16919011 ↗MalfunctionMay 12, 2023Loose or Intermittent Connection; Misfire; Loose or Intermittent Connection; Misfire
IT WAS REPORTED THAT THE ANASTOCLIP GC MEDIUM CLIPS WERE MISFIRING AND THEN COMING OUT OF THE VESSEL HANGING/LOOSE AFTER BEING DEPLOYED. THERE WERE NO PATIENT INJURY. TWO DEVICES WERE INVOLVED IN THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16919006 ↗MalfunctionMar 20, 2022Difficult to Open or Close; Difficult to Open or Close
CLIP WILL FALL OUT OR CLIP WILL BE PARTIALLY CRIMPED AFTER ACTIVATING THE APPLIER. FIVE OR SIX CLIPS WILL APPLY PERFECTLY AND THEN A SERIES OF 2-3 THAT FALL OUT UNCRIMPED OR DO NOT CLOSE COMPLETELY AFTER SQUEEZING THE APPLIER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13830813 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K091987. One submission often covers many models, so check the product and lot details in each record.
- Z-0165-2011 · Sep 22, 2010 · TerminatedLeMaitre Vascular AnastoClip GC Vessel Closure System-Large Size Models: 4007-03 Large size 3 Pack 4008-03 Large size 10 Pack 4007-07 Large size 1 Pack Manufactured by LeMatire Vas
- Z-0164-2011 · Sep 22, 2010 · TerminatedLeMaitre Vascular AnastoClip GC Vessel Closure System-Medium Size Models: 4007-02 Medium size 3 Pack 4008-02 Medium size 10 Pack 4007-06 Medium size 1 Pack Manufactured by LeMatire
Other Clip, Implantable devices
- LIGAMAX EL5MLEthicon Endo-Surgery, LLC · CLIP, IMPLANTABLE3 deaths3,808 reports
- Endo Clip III 176630Covidien LP · CLIP, IMPLANTABLE1 death1,959 reports
- Epix® Universal Clip Applier CA500Applied Medical Resources Corporation · Clip, Implantable1 recall1 death1,183 reports
- Endo Clip II 176657Covidien LP · CLIP, IMPLANTABLE1,183 reports
- LIGACLIP ER320Ethicon Endo-Surgery, LLC · CLIP, IMPLANTABLE1,115 reports
- Direct Drive® Clip Applier CA090Applied Medical Resources Corporation · Clip, Implantable1 death806 reports
- LAPRA-TY XC200Ethicon Inc. · CLIP, IMPLANTABLE791 reports
- Endo Clip 176625Covidien LP · CLIP, IMPLANTABLE1 death768 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page