Skip to content

AnastoClip GC 8cm (Medium) 4008-06Recalls & reported problems

Lemaitre Vascular, Inc. · AnastoClip GC 8cm (Medium) · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00840663109579), received between Mar 20, 2022 and Aug 6, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Failure to Form Staple” (4 reports).

2 other recalls cover devices cleared under the same FDA submission (K091987); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Form Staple4
  • Loose or Intermittent Connection2
  • Misfire2
  • Difficult to Open or Close1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Infiltration into Tissue1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 6, 2025Failure to Form Staple; Mechanical Jam; Failure to Form Staple; Mechanical Jam

    IT WAS REPORTED THAT DURING A BVT PROCEDURE, THE ANASTOCLIP GC CLIP APPLIER JAMMED AND STUCK ONTO THE PATIENT'S ARM AND COULD NOT GET IT TO RELEASE THE CLIP OR OFF THE PATIENT'S ARM. THEY HAD TO CUT OUT PART OF THE VEIN. DEVICE WAS CHECKED PRIOR TO USE. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE OPERATION. NO OTHER ISSUES OR INJURY WERE REPORTED TO THE PATIENT.

    Patient problems: Infiltration into Tissue

    MAUDE report 22729546 ↗
  • MalfunctionFeb 19, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT THE ANASTOCLIP GC CLIPS KEEPS MISFIRING. THE CLIP DEPLOYS WITHOUT COMPLETELY CLOSING. THE PRODUCT WAS NOT DIRECTLY USED ON THE PATIENT AND NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21417808 ↗
  • MalfunctionFeb 19, 2025Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT THE ANASTOCLIP GC CLIPS WERE MISFIRING. THE CLIP DEPLOYS WITHOUT COMPLETELY CLOSING SO THEY FALL OFF. DEVICE WAS NOT DIRECTLY USED ON PATIENT. NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21417759 ↗
  • MalfunctionAug 21, 2024Failure to Form Staple; Failure to Form Staple

    IT WAS REPORTED THAT THE ANASTOCLIP GC CLIPS WERE NOT CLOSING AND GRIPPING PROPERLY WHEN USED DURING THE CASE. IT HAD TO BE REMOVED THEM FROM THE DURA DURING A TETHERED CORD PROCEDURE. NO INJURY WAS REPORTED TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20021675 ↗
  • MalfunctionMay 12, 2023Loose or Intermittent Connection; Misfire; Loose or Intermittent Connection; Misfire

    IT WAS REPORTED THAT THE ANASTOCLIP GC MEDIUM CLIPS WERE MISFIRING AND THEN COMING OUT OF THE VESSEL HANGING/LOOSE AFTER BEING DEPLOYED. THERE WERE NO PATIENT INJURY. TWO DEVICES WERE INVOLVED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16919011 ↗
  • MalfunctionMay 12, 2023Loose or Intermittent Connection; Misfire; Loose or Intermittent Connection; Misfire

    IT WAS REPORTED THAT THE ANASTOCLIP GC MEDIUM CLIPS WERE MISFIRING AND THEN COMING OUT OF THE VESSEL HANGING/LOOSE AFTER BEING DEPLOYED. THERE WERE NO PATIENT INJURY. TWO DEVICES WERE INVOLVED IN THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16919006 ↗
  • MalfunctionMar 20, 2022Difficult to Open or Close; Difficult to Open or Close

    CLIP WILL FALL OUT OR CLIP WILL BE PARTIALLY CRIMPED AFTER ACTIVATING THE APPLIER. FIVE OR SIX CLIPS WILL APPLY PERFECTLY AND THEN A SERIES OF 2-3 THAT FALL OUT UNCRIMPED OR DO NOT CLOSE COMPLETELY AFTER SQUEEZING THE APPLIER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13830813 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K091987. One submission often covers many models, so check the product and lot details in each record.

  • Z-0165-2011 · Sep 22, 2010 · Terminated
    LeMaitre Vascular AnastoClip GC Vessel Closure System-Large Size Models: 4007-03 Large size 3 Pack 4008-03 Large size 10 Pack 4007-07 Large size 1 Pack Manufactured by LeMatire Vas
  • Z-0164-2011 · Sep 22, 2010 · Terminated
    LeMaitre Vascular AnastoClip GC Vessel Closure System-Medium Size Models: 4007-02 Medium size 3 Pack 4008-02 Medium size 10 Pack 4007-06 Medium size 1 Pack Manufactured by LeMatire

Other Clip, Implantable devices

All Clip, Implantable devices →

Sources and updates