i-STAT ACT Celite Cartridges recallZ-0399-2011 · Abbott Point Of Care Inc. · initiated Oct 6, 2010
Terminated All recalls posted in November 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Point Of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Certain lots of i-STAT Celite ACT and Kaolin ACT cartridges may be difficult to fill or will not fill when attempting to transfer blood to the sample entry port of the cartridge.
FDA root cause category: Other
What the company is doing
Abbott Point of Care, Inc. sent an URGENT RECALL NOTICE dated September 2010 on October 6, 2010 by Federal Express to all US customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of cartridges from the identified lots. If customers did not currently have replacement cartridges then they should continue to use the cartridge lot until their facility receives replacement material. Upon receipt of replacement material, customers were to discontinue use of and return the affected cartridges and switch to the replacement material. All unused cartridges will be credited and will be returned. If you have any questions regarding this information, please contact Abbott Point of Care Technical Support at 800-366-8020, Option 1 or your Abbott Point of Care representative.
Recall details
- Recall number
- Z-0399-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 6, 2010
- Posted by the FDA
- Nov 17, 2010
- Recalling firm
- Abbott Point Of Care Inc.
- Quantity
- 49,025 cartridges.
- Product code
- JBP
- FDA submissions
- K992571
- Distribution
- Worldwide Distribution
Product description
i-STAT ACT Celite Cartridges in vitro diagnostic test used to monitor moderate and high-level heparin therapy through analysis of arterial and venous whole blood samples.
Code information (models, lots, serial numbers)
List numbers: 07G01-01, 07G01-02,. 600-9006-25; Lot numbers: S10127B, S10136A, S10150A, S10151, S10155.
Other devices under the same FDA submission
1 device cleared under K992571 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- i-STAT 03P86-25Abbott Point Of Care Inc. · ACTIVATED WHOLE BLOOD CLOTTING TIME27 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 57005).
- Z-0400-2011 · Terminatedi-STAT ACT Kaolin Cartridges in vitro diagnostic test which uses fresh whole blood to monitor high-dose heparin anticoagulation frequently associated with cardiovascular surgery.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page