i-STAT ACT Kaolin Cartridges recallZ-0400-2011 · Abbott Point Of Care Inc. · initiated Oct 6, 2010

Terminated All recalls posted in November 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Point Of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain lots of i-STAT Celite ACT and Kaolin ACT cartridges may be difficult to fill or will not fill when attempting to transfer blood to the sample entry port of the cartridge.

FDA root cause category: Other

What the company is doing

Abbott Point of Care, Inc. sent an URGENT RECALL NOTICE dated September 2010 on October 6, 2010 by Federal Express to all US customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of cartridges from the identified lots. If customers did not currently have replacement cartridges then they should continue to use the cartridge lot until their facility receives replacement material. Upon receipt of replacement material, customers were to discontinue use of and return the affected cartridges and switch to the replacement material. All unused cartridges will be credited and will be returned. If you have any questions regarding this information, please contact Abbott Point of Care Technical Support at 800-366-8020, Option 1 or your Abbott Point of Care representative.

Recall details

Recall number
Z-0400-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 6, 2010
Posted by the FDA
Nov 17, 2010
Recalling firm
Abbott Point Of Care Inc.
Quantity
95,325 cartridges
Product code
JBP
FDA submissions
K023582
Distribution
Worldwide Distribution

Product description

i-STAT ACT Kaolin Cartridges in vitro diagnostic test which uses fresh whole blood to monitor high-dose heparin anticoagulation frequently associated with cardiovascular surgery.

Code information (models, lots, serial numbers)

List numbers: 07G81-01, 07G81-02; Lot numbers: S10129, S10129A, S10130, S10130B, S10147A, S10147B, S10147C, S10147D, S10147E, S10148A, S10148B, S10148C, S10148D, S10148G, S10149, S10149B, S10149C, S10151A, S10151B, S10152, S10152A, S10152B, S10152C, S10152D, S10152E, S10152F, S10154A, S10154B, S10154D, S10160B, S10162.

Other devices under the same FDA submission

1 device cleared under K023582 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 57005).