Skip to content

LeMaitre VascuTape Radiopaque Tape, LeMaitre Stent Guide… recallZ-0433-2011 · Lemaitre Vascular, Inc. · initiated Oct 22, 2010

Terminated All recalls posted in November 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Lemaitre Vascular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Box label (Top Part). Lot and Expiration Date information were switched.

FDA root cause category: Labeling Change Control

What the company is doing

Lemaitre Vascular notified customers by an Urgent: VascuTape Radiopaque Tape Device Field Safety Notice letter on 10/22/10 via USPS with delivery confirmation. The Notice advised users of the incorrect lot and lot number on the outer box and also provided an image of the box and label for ease in identifying the product in their inventory. The letter states that the use and functionality of the device are not affected by the problem; however, if customers feel uncomfortable continuing to use the affected device, they can contact Customer Service for replacement. The firm recommended discarding the outer box of the affected product. The letter contains a form that is requested to be returned to LeMaitre Vascular as record of notification and reconciliation. Accounts can contact customer service at 800-628-9470.

Recall details

Recall number
Z-0433-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 22, 2010
Posted by the FDA
Nov 22, 2010
Recalling firm
Lemaitre Vascular, Inc.
Quantity
160 units
Product code
FTY
FDA submissions
K932184
Distribution
Nationwide Distribution -- AL, CA, CT, IL, ME, MI, MS, NC, NJ, NY, PA, PR, TX, and WA.

Product description

LeMaitre VascuTape Radiopaque Tape, LeMaitre Stent Guide, 20 strips Catalog Number: 1102-20 Intended to be placed on the patient's skin to assist during imaging procedures by serving as an internal frame of reference.

Code information (models, lots, serial numbers)

Lot Number: LSG1109 Exp Date: 2015-06

Other devices under the same FDA submission

1 device cleared under K932184 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 57028).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page