Skip to content

Zimmer, NexGen Complete Knee Solution LPS Flex Gender… recallZ-0533-2011 · Zimmer Inc. · initiated Sep 15, 2010

Terminated All recalls posted in December 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

These lots of Zimmer NexGen Complete Knee Solution LPS Femoral Component and Zimmer NexGen Complete Knee Solution LPS Flex Gender Femoral Component exhibit a nonconforming internal CAM radius. The condition is intermittent and not all devices have nonconforming geometry.

FDA root cause category: Process design

What the company is doing

The firm sent URGENT DEVICE CORRECTION AND REMOVAL notices to distributors and sales staff, Risk Managers, and implanting surgeons dated 9/15/2010. Sales staff and risk managers were instructed to return any product on hand. Surgeons were informed of the issue but there was no action to be taken by surgeons.

Recall details

Recall number
Z-0533-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 15, 2010
Posted by the FDA
Dec 2, 2010
Recalling firm
Zimmer Inc.
Quantity
158 all sizes of 5764
Product code
JWH
FDA submissions
K060370, K960279
Distribution
US, Germany, Spain, France, UK, Italy, Sweden, South Africa, Czech Republic, Bulgaria, Poland, Romania, and Austria.

Product description

Zimmer, NexGen Complete Knee Solution LPS Flex Gender Femoral Component, Size G Left, sterile, REF 00-5764-017-51.

Code information (models, lots, serial numbers)

61499865

Other devices under the same FDA submission

80 devices cleared under K060370, K960279 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 57178).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page