ORTHOFIX Firebird Spinal Fixation System, Pedicle Screw… recallZ-1608-2011 · Orthofix Inc · initiated Oct 22, 2010

Terminated All recalls posted in March 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Orthofix Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a possibility that certain Anti-Splay Caps may not mate properly during assembly of the construct to the Screw Body with Multi-Axial Reduction Screw Assemblies.

FDA root cause category: Other

What the company is doing

The firm, Orthofix, sent an "URGENT: PRODUCT RECALL" letter dated October 22, 2010 to all customers. The letter describes the product, problem and the action to be taken by the customers. The customers were instructed to: 1) identify and remove the affected product 2) complete the enclosed Tracking and Verification Form via fax to: Orthofix Inc. at 214-937-2764, and 3) contact their Account Services Representative at 1-888-298-5700 and obtain a return authorization number to return product to Orthofix, Inc. The letter also states that Orthofix Inc., will provide replacement Anti-Splay Caps via priority delivery. If you have any questions regarding the removal and return of this product to Orthofix Inc., please contact your Account Service Representative at 1-888-298-5700. Should you have any other questions or concerns, please contact the Director of Regulatory Affairs at 214-937-2061 or by email: [email protected].

Recall details

Recall number
Z-1608-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 22, 2010
Posted by the FDA
Mar 8, 2011
Recalling firm
Orthofix Inc
Quantity
166 Units
Product code
NKB
FDA submissions
K081684
Distribution
Nationwide distribution: USA including states of: CA, CO, FL, LA , MD, MI, MO, NJ, NY, OH, TN, VA, and WA.

Product description

ORTHOFIX Firebird Spinal Fixation System, Pedicle Screw Spinal System, Reduction Case, Part #68-0111, Lot #16518-LC18. Intended for use in conjunction with Multi-Axial Reduction Screw Assemblies. During assembly of a construct, the Anti-Splay Cap locks onto the Screw Body to prevent the Screw Body tabs from splaying (spreading apart). The Anti-Splay Cap is subsequently removed from the Screw Body after assembly of the construct. The Anti-Splay Cap is a stainless steel machined instrument. The instrument is used within the Firebird Reduction Kit (Domestic, 44-9050) and the ICON Reduction Kit (International, 68-0010). Product Code: NKB-Orthosis, Spinal Pedicle Fixation, For Degenerative Disc Disease. Class II.

Code information (models, lots, serial numbers)

Part #68-0111, Lot #16518-LC18.

Other devices under the same FDA submission

3 devices cleared under K081684 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.