Mobilett XP Hybrid/Digital/CR recallZ-0718-2011 · Siemens Medical Solutions USA, Inc · initiated Oct 4, 2010
Terminated All recalls posted in December 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
battery fuse
FDA root cause category: Other
What the company is doing
SIEMENS released a field update dated October 25, 2010, via SP011/10/S to the representatives. The update described issue and corrective action plan. The representatives are to visit each account to upgrade the systems with a new battery fuse and a new holder. If you have any questions regarding this report, please contact Quality Engineer by phone 610-448-3237 or email: [email protected] and Director, Product complaints & Field Support by phone 610-448-4634 or email: [email protected].
Recall details
- Recall number
- Z-0718-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 4, 2010
- Posted by the FDA
- Dec 17, 2010
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 170 units
- Product code
- IZL
- FDA submissions
- K033238
- Distribution
- Nationwide distribution: AZ, CA, DC, MD, PA, SC and VA.
Product description
Mobilett XP Hybrid/Digital/CR Intended use: mobile x-ray system
Code information (models, lots, serial numbers)
Model numbers 1818454, 1818447, and 1818959
Other devices under the same FDA submission
2 devices cleared under K033238 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Mobilett XP 01818363Siemens Healthcare GmbH · System, x-ray, mobile5 reports
- Mobilett XP hybrid 01818454Siemens Healthcare GmbH · System, x-ray, mobile2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page