Ovation Hip Stem, REF 201-0104, Offset Stem Inserter… recallZ-1673-2011 · Ortho Development Corporation · initiated Jan 11, 2011
Terminated All recalls posted in March 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho Development Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Component of total hip arthroplasty may break during surgical procedure.
FDA root cause category: Process design
What the company is doing
Ortho Development Corporation notified Field representatives of the issue and conditions that could lead to failure by e-mail on December 14, 2010. Redesigned product is being produced for replacement. For questions regarding this recall call 801-619-3450.
Recall details
- Recall number
- Z-1673-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jan 11, 2011
- Posted by the FDA
- Mar 16, 2011
- Recalling firm
- Ortho Development Corporation
- Quantity
- 33 units
- Product code
- LZO
- FDA submissions
- K062775
- Distribution
- Nationwide Distribution including AZ, CA, CO, FL, MI, NV, OR, and UT.
Product description
Ovation Hip Stem, REF 201-0104, Offset Stem Inserter Neutral, Non-Sterile, Ortho Development, Draper, Utah 84020. The Ovation Hip System is intended for use in total hip arthroplasty. The hip stem inserter is an instrument used to seat the hip stem in the femur. There is a notch in the top of the hip stem, and the hip stem inserter has a corresponding tip at the end of the instrument. The tip of the inserter is designed to fit into the notch of the hip stem, as a guide. Once this is inserted, the surgeon will tap on the opposite end of the inserter with a hammer to seat the press-fitted stem assembly.
Code information (models, lots, serial numbers)
All lots
Other devices under the same FDA submission
5 devices cleared under K062775 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Ovation Hip Stem 101-0007Ortho Development Corporation · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED4 reports
- Ovation Hip Stem 101-0006Ortho Development Corporation · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- Ovation Hip Stem 101-0005Ortho Development Corporation · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- Ovation Hip Stem 101-0003Ortho Development Corporation · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
- Ovation Hip Stem 101-1008Ortho Development Corporation · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 57747).
- Z-1670-2011 · TerminatedOvation Hip Stem, REF 201-0002, Hip Stem Inserter, Non-Sterile, Ortho Development, Draper, Utah 84020. The Ovation Hip System is intended for use in total hip arthroplasty. The hip…
- Z-1671-2011 · TerminatedOvation Hip Stem, REF 201-0102, Offset Stem Inserter RT, Non-Sterile, Ortho Development, Draper, Utah 84020. The Ovation Hip System is intended for use in total hip arthroplasty. T…
- Z-1672-2011 · TerminatedOvation Hip Stem, REF 201-0103, Offset Stem Inserter LT, Non-Sterile, Ortho Development, Draper, Utah 84020. The Ovation Hip System is intended for use in total hip arthroplasty. T…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page