MAGNETOM Avanto, Espree, Trio A Tim System recallZ-2021-2011 · Siemens Medical Solutions USA, Inc · initiated Mar 3, 2011
Terminated All recalls posted in April 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm discovered during product monitoring that on a few systems the gradient cable connections did not meet their specifications after installation. The potential exists for a cable connection to become overheated, leading to the emission of smoke and possible material damage.
FDA root cause category: Other
What the company is doing
Siemens Medical Solutions Inc., issued a Customer Safety Advisory Notice dated November 24, 2010, to affected customers via Update Instructions MR029/10S. Affected customers will be notified via hand delivered letter by Siemens Service Engineers at the time of inspection. A future Update instruction will be released for installation of detectors in appropriate locations on systems to switch off the power electronics in the event of a malfunction in the power supply of the gradient coils. Further questions regarding this recall please call (610) 448-3237.
Recall details
- Recall number
- Z-2021-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 3, 2011
- Posted by the FDA
- Apr 18, 2011
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 1419
- Product code
- LNH
- FDA submissions
- K013586, K032428, K041112, K050200
- Distribution
- Product was distributed domestically to hospitals and imaging centers in the District of Columbia, Puerto Rico, AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, and WY(every State except VT).
Product description
MAGNETOM Avanto, Espree, Trio A Tim System. Nuclear magnetic resonance imaging.
Code information (models, lots, serial numbers)
Show the full code information (10k characters)
Model numbers 10018165, 10018221, 10018222, 7387074, 7391167. Serial numbers 30265, 30204, 35108, 30646, 35020, 25297, 30125, 30234, 30435, 30385, 26326, 30317, 30476, 35258, 30019, 30130, 30035, 30851, 30475, 30759, 26182, 30833, 25635, 30162, 25998, 35227, 30153, 35012, 30226, 30028, 30380, 30045, 30202, 25300, 25880, 30108, 25758, 30799, 30616, 30767, 30940, 26213, 30426, 27213, 35161, 31089, 25492, 26566, 35299, 25242, 25293, 30504, 35330, 30454, 30233, 30338, 30529, 30208, 30740, 30482, 30472, 30383, 26425, 30364, 30499, 30275, 30834, 26698, 30254, 26223, 26329, 30124, 30512, 30166, 30410, 26553, 25409, 35015, 26900, 30602, 30957, 30144, 30780, 30896, 30239, 30839, 30250, 30298, 30083, 30730, 30986, 30367, 30109, 35064, 30225, 31000, 30857, 30440, 30012, 30350, 35318, 35319, 35321, 26491, 30471, 30242, 30948, 30337, 30392, 30582, 30058, 35403, 25911, 30938, 30474, 30732, 27641, 30601, 25134, 35259, 25717, 25420, 20600, 27633, 35111, 26364, 30391, 27175, 25154, 25424, 30056, 30742, 26194, 30959, 30300, 30830, 30353, 30777, 30201, 35312, 35311, 26504, 30887, 27260, 30958, 31033, 35176, 30238, 30176, 35077, 30652, 35408, 27101, 30273, 35222, 30925, 30090, 25852, 30272, 25230, 30032, 25264, 35152, 25991, 30066, 25286, 31001, 26016, 26272, 30753, 27013, 30224, 30720, 35380, 30607, 30633, 31015, 35010, 25215, 25427, 25962, 27505, 30783, 26077, 30112, 31039, 25316, 35364, 26230, 27157, 25716, 30627, 26145, 30039, 30346, 25054, 26358, 35235, 35182, 26031, 35201, 25948, 30255, 35008, 35014, 25069, 25408, 26657, 26953, 27162, 25931, 26252, 27156, 25063, 26827, 26088, 25672, 25179, 25976, 25114, 25893, 25395, 30689, 30569, 30402, 25888, 30650, 25552, 25231, 30811, 30539, 30705, 30059, 30960, 30982, 35261, 35029, 25389, 25730, 25768, 26856, 30656, 30137, 30628, 30591, 35050, 30017, 30249, 30132, 30782, 35351, 26726, 25678, 30180, 30622, 27316, 25905, 25582, 25166, 30597, 30992, 25921, 25551, 25491, 30388, 25310, 30481, 30256, 30401, 30647, 26465, 27051, 25651, 30791, 30635, 25818, 26187, 30552, 26668, 35348, 30468, 30452, 30348, 25507, 30962, 27079, 26762, 26764, 30424, 30119, 25576, 30549, 26359, 31094, 30111, 30161, 30880, 30846, 30291, 30347, 30187, 30015, 30453, 30478, 25680, 30614, 30040, 25404, 25484, 27306, 26114, 26121, 25807, 30072, 35163, 30590, 30837, 26436, 30524, 35046, 26244, 27707, 25445, 35125, 30542, 35110, 25108, 30550, 30838, 30441, 35221, 30422, 30803, 35398, 30506, 35167, 25429, 25292, 30443, 30534, 30596, 25663, 25992, 27019, 25352, 25386, 35025, 26356, 31020, 25505, 30564, 30528, 30386, 25207, 26141, 25438, 25444, 26360, 30357, 25960, 26311, 26280, 30670, 26792, 25025, 30770, 26171, 25993, 25525, 30360, 30488, 25145, 30613, 30972, 30733, 30662, 30985, 25307, 25364, 30570, 25682, 30171, 35172, 25977, 30258, 30771, 27376, 30036, 30487, 30723, 30598, 30370, 30044, 30316, 30247, 31074, 25523, 35052, 25489, 26591, 26062, 26653, 30175, 30669, 31013, 25459, 25785, 30868, 30618, 26287, 26383, 30800, 30459, 27252, 35186, 26078, 30758, 30089, 30553, 25572, 30810, 26779, 35389, 30376, 25796, 35358, 26281, 30557, 30793, 30879, 31072, 25175, 26224, 35323, 27347, 30750, 26446, 35250, 30467, 30522, 26665, 30408, 30375, 30991, 35240, 30115, 30561, 26619, 30721, 31023, 30145, 25877, 30587, 30866, 35157, 30063, 30359, 30182, 30711, 26398, 26888, 30743, 26217, 25967, 26897, 27426, 25696, 25196, 30345, 25763, 25041, 30497, 26617, 30136, 30122, 30207, 35160, 30014, 30438, 30381, 30491, 30844, 31046, 30304, 30823, 30935, 30719, 30910, 30832, 30623, 30462, 30445, 31084, 31049, 30068, 35062, 25058, 35043, 30076, 30968, 25520, 25771, 30362, 26168, 30295, 30685, 27177, 26442, 26649, 27451, 30369, 30663, 27366, 31036, 30054, 30503, 30544, 30530, 30749, 30966, 30460, 27180, 60600, 30738, 25978, 25985, 30287, 30817, 35192, 35236, 25497, 26269, 30798, 31012, 30448, 30518, 31059, 25203, 30220, 30593, 30069, 25332, 30565, 30904, 30951, 25720, 26945, 26785, 34144, 31041, 30463, 30080, 30179, 30413, 25105, 30520, 30021, 26881, 30029, 26621, 30065, 30071, 25927, 30407, 30061, 30248, 30509, 26533, 26400, 30227, 35280, 35281, 25104, 31092, 35317, 25337, 26053, 35180, 30169, 30197, 30211, 31098, 30319, 30067, 25460, 27344, 30915, 30661, 30559, 25606, 30320, 30423, 30870, 26126, 35115, 30451, 35101, 35288, 35022, 25157, 26039, 30005, 30082, 25581, 25973, 26051, 266227, 30525, 25533, 25411, 25817, 25558, 35070, 26399, 35023, 30563, 30271, 30043, 30210, 25021, 30206, 30378, 30536, 35307, 30228, 25610, 30632, 26966, 26449, 30680, 31065, 26938, 30146, 30170, 25044, 30676, 30297, 30765, 26080, 35074, 26423, 30994, 30703, 25284, 26397, 31008, 26435, 26870, 27059, 30447, 25077, 26286, 27179, 25926, 30264, 30638, 30763, 35428, 35066, 25822, 30643, 25322, 25384, 26846, 25875, 25126, 27161, 30241, 30989, 30548, 35164, 27053, 35141, 30936, 26963, 30398, 25152, 30185, 30515, 35162, 35006, 25096, 30588, 35238, 27054, 30787, 30049, 26992, 25225, 30490, 35295, 30806, 30232, 30342, 30864, 30523, 30074, 30577, 25550, 25957, 25358, 30340, 25701, 30675, 25249, 30824, 30368, 30414, 27467, 30876, 35007, 25036, 25727, 25952, 30581, 30277, 30639, 30668, 30366, 30142, 30621, 20587, 27297, 30748, 30216, 30775, 25470, 30025, 26255, 26418, 30425, 30430, 30404, 25146, 30400, 30924, 35104, 30814, 30470, 35390, 30308, 25135, 26216, 26704, 31034, 30592, 30148, 30756, 26620, 26193, 30869, 25454, 30075, 31005, 30917, 30953, 30975, 30943, 30918, 30946, 25783, 25797, 35279, 25030, 35001, 20591, 25010, 25403, 30786, 35423, 30263, 30651, 31060, 31053, 30096, 25910, 27354, 30872, 30094, 25594, 30952, 30768, 30150, 30269, 26373, 35187, 26692, 30789, 26350, 31018, 25352, 26241, 30997, 25174, 30349, 30305, 26454, 30106, 30389, 30508, 35427, 35013, 25023, 26461, 27324, 30726, 30828, 27471, 26052, 35075, 25792, 25949, 30326, 25383, 26957, 27040, 30507, 27194, 31067, 30095, 30510, 25274, 30713, 30047, 30167, 30871, 30728, 30081, 30855, 27204, 30023, 26325, 30266, 30302, 30103, 35271, 35139, 30289, 35078, 30100, 30543, 30862, 30858, 30235, 30406, 25527, 30697, 30151, 35136, 35131, 31086, 27244, 30437, 25477, 30325, 30708, 30154, 30363, 30978, 30874, 30617, 30784, 35385, 26804, 30585, 30664, 30667, 30578, 26011, 27293, 31083, 30160, 30261, 30428, 25372, 30725, 30631, 30558, 30343, 30772, 30377, 35393, 30746, 30502, 30418, 25530, 26489, 26590, 26723, 30229, 30849, 35212, 30473, 26486, 26333, 25903, 30282, 25342, 30446, 25412, 30236, 30630, 26670, 26546, 35048, 26370, 30484, 27100, 30055, 30433, 30379, 30129, 30555, 26627, 30306, 30571, 25961, 30020, 30114, 30116, 25180, 25305, 25690, 30009, 25185, 26004, 30458, 30717, 25919, 30744, 30107, 30541, 30371, 25123, 30533, 25614, 26314, 30745, 30580, 31104, 30214, 25980, 26319, 25684, 27534, 30361, 30033, 30231, 26048, 26130, 26628, 30949, 30673, 30138m 27046, 30575, 26371, 35191, 25435, 30140, 26927, 27198, 25833, 35026, 30629, 25479, 26021, 30535, 30046, 30477, 30854, 25247, 31022, 30152, 30372, 25537, 27253, 30091, 27099, 31341, 30684, 30358, 35011, 26064, 30420, 35243, 26568, 26959, 25216, 26808, 26427, 27443, 30276, 30479, 30773, 35297, 26243, 25958, 35264, 30373, 26559, 30223, 35324, 30956, 30781, 30434, 30605, 35168, 30609, 30149, 30642, 30027, 30262, 30988, 30159, 30042, 26075, 35005, 35171, 25150, 35019, 30945, 31100, 30483, 25772, 30526, 30164, 25222, 30415, 26462, 27620, 27557, 30030, 25164, 27004, 25093, 30566, 26835, 30419, 30092, 30158, 30974, 30449, 25644, 27381, 30034, 30057, 26037, 25815, 30574, 30230, 30314, 26159, 35284, 26148, 26178, 30163, 30268, 30352, 25485, 30399, 26284, 30053, 25382, 30022, 26276, 25374, 25837, 25298, 27207, 27714, 25899, 30812, 25074, 25946, 26662, 30270, 31087, 25078, 27033, 30313, 30501, 25155, 26026, 30365, 27330, 35290, 30498, 26532, 25509, 27222, 30239, 30356, 25240, 30155, 31071, 30779, 35018, 25718, 26689, 31097, 35184, 35045, 27102, 35188, 30253, 25070, 25767, 26560, 30191, 30301, 35153, 26615, 25436, 30198, 26155, 25528, 20573, 25248, 30442, 27201, 25430, 26956, 30267, 27088, 27066, 25959, 25038, 30916, 30290, 25705, 30064, 35241, 30084, 30190, 30037, 30395, 25046, 26469, 30296, 30604, 25372, 30665, 30344, 26018, 31032, 30293, 25087, 35344, 26569, 30845, 31014, 25033, 26451, 30060, 35365, 25245, 26219, 30396, 35445, 30240, 30024, 26885, 30931, 25681, 30456, 30485, 30822, 25707, 30625, 25086, 35357, 25362, 25625, 25694, 30041, 25024, 35230, 30279, 26452, 30004, 30311, 31080, 30085, 35328, 26320, 35343, 35003, 25016, 35260, 25212, 27595, 35426, 30141, 25969, 30843, 25511, 26191, 30079, 30480, 26575, 30411, 30829, 35412, 25140, 25106, 26550, 27230, 30836, 30118, 27122, 35035, 35030, 25712, 26109, 25611, 35273, 25052, 30086, 30809, 35257, 27049, 30196, 30178, 30769, 35114, 25347, 35229, 35068, 35044, 25963, 26811, 35053, 30653, 35267, 35024, 25706, 25870, 27214, 25029, 30121, 30671, 35072, 35142, 25233, 25776, 35021, 30429, 35338, 25320, 35335, 35069, 30194, 35154, 25886, 35067, 35039, 25198, 30165, 35327, 35061, 30427, 30128, 35263, 35017, 25820, 35313, 26564, 35228, 25360, 31006, 25457, 30382, 25725, 25047, 30551, 30595, 30692, 25539, 25580, 25654, 26396, 26421, 27358, 30120, 30200, 26751, 30143, 30937, 30831, 35216, 35217, 35073, 30193, 27050, 25814, 30942, 30412, 30516, 30734, 30825, 30848, 30941, 31021, 31056, 35291, 35418, 25350, 25381, 25667, 25675, 26054, 26248, 26492, 26608, 26652, 26787, 26893, 27221, 27258, 27412, 26424, 30299, 30436, 30666, 30674, 30928, 25508, 26648, 26781, 26962, 30926, 30990, 26214, 20616, 26852, 30374, 25453, 30173, 25144, 26125, 26535, 30221, 27637, 30243, 35443, 35439, 35444, 30619, 25232, 26192, 30556, 30213, 25065, 25432, 30710, 25734, 30469, 30511, 26972, 30354, 25103, 25956, 30658, 35177, 35248, 26321, 30283, 25912, 35415, 30922, 30192, 30310, 31025, 26747, 25510, 30126, 30612, 25390, 25197, 30390, 30626, 35155, 26614, 27172, 30199, 30881, 35174, 25800, 25540, 30339, 30724, 25693, 30955, 26132, 30186, 25816, 30659, 31052, 26937, 30038, 30246, 25609, 30093, 30464, 26998, 30774, 27224, 30785, 27026, 35407, 35071, 35076, 30123, 30203, 30281, 30387, 30409, and 30181.
Other devices under the same FDA submission
1 device cleared under K013586, K032428, K041112, K050200 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- MAGNETOM Avanto 10684332Siemens Healthcare GmbH · System, nuclear magnetic resonance imaging3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.