Penumbra brand Penumbra Coil System, Penumbra Coil 400… recallZ-1906-2011 · Penumbra Inc. · initiated Mar 4, 2011
Terminated All recalls posted in April 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Penumbra Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The diameter of a component of the device is out of tolerance specification; the Delivery Pusher pull wire which secures the coil in place can prolapse and allow premature detachment of the coil.
FDA root cause category: Process change control
What the company is doing
Penumbra sent an Urgent Voluntary Field Removal Notice dated March 3, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately inspect all stock of Penumbra Coil 400 and identify product from the Lots. Remove product from these Lots from inventory and place it in quarantine for return to Penumbra Inc. (Do not use product from these Lots). Complete and return the attached product identification/ return form with in 5 business days via fax to +1 510-217-6414. Penumbra employees will contact customer directly to arrange return and replacement of product. For questions call Regulatory Affairs at 510-748-3224 or e-mail [email protected].
Recall details
- Recall number
- Z-1906-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 4, 2011
- Posted by the FDA
- Apr 12, 2011
- Recalling firm
- Penumbra Inc.
- Quantity
- 2082
- Product code
- HCG
- FDA submissions
- K103305
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries in the European Union (Denmark, Germany, Italy, Netherlands, Slovakia, Slovenia, Spain, United Kingdom and Switzerland) and South Africa.
Product description
Penumbra brand Penumbra Coil System, Penumbra Coil 400, Product is manufactured and distributed by Penumbra Inc., Alameda, CA The Penumbra Coil System (Penumbra Coil 400 TM) is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous fistulae. The coils are also intended for arterial and venous embolization in the peripheral vasculature. The coils function to selectively embolize aneurysms or targeted segments of the vasculature by packing a sufficient quantity of soft platinum coils to achieve occlusion. The Penumbra Coil System is compatible with the Penumbra PX 400 Microcatheter.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
F17211, F17212, F17213, F17214, F17215, F17216, F17256, F17257, F17258, F17259, F17260, F17310, F17311, F17333, F17334, F17335, F17336, F17337, F17338, F17339, F17340, F17374, F17375, F17376, F17395, F17396, F17397, F17398, F17399, F17400, F17401, F17402, F17403, F17404, F17405, F17440, F17441, F17442, F17443, F17444, F17445, F17446, F17447, F17448, F17485, F17486, F17487, F17489, F17507, F17511, F17512, F17513, F17514, F17515, F17516, F17517, F17564, F17565, F17566, F17567, F17568, F17569, F17668, F17669, F17670, F17672, F17673, F17675, F17676, F17705, F17706, F17707, F17708, F17710, F17735, F17736, F17737, F17738, F17739, F17740, F17741, F17742, F17743, F17744, F17745, F17780, F17781, F17782, F17783, F17790, F17791, F17792, F17793, F17794, F17833, F17834, F17835, F17836, F17837, F17838, F17839, F17840, F17841, F17842, F17843, F17844, F17900, F17901, F17902, F17903, F17904, F17905, F17906, F17907, F17908, F17924, F17925, F17926, F17927, F17928 , F17929, F17930, F17931, F17932, F17933, F17934, F17935, F17936, F17966, F17967, F17968, F17969, F17970, F17971, F17972, F17973, F17974, F17975, F17976, F17977, F17999, F18000, F18001, F18002, F18021, F18022, F18023, F18024, F18041, F18042, F18043, F18044, F18045, F18046, F18047, F18066, F18068, F18073, F18074, F18075, F18076, F18090, F18091, F18092, F18093, F18094, F18121, F18122, F18123, F18124, F18125, F18126, F18127, F18128, F18129, F18146, F18147, F18148, F18149, F18150, F18151, F18152, F18153, F18179, F18180, F18181, F18182, F18183, F18184, F18185, F18186, F18194, F18195, F18196, F18197, F18231, F18232, F18233, F18234, F18235, F18236, F18239, F18240, F18241, F18242, F18243, F18244, F18245, F18246, F18247, F18248, F18249, F18291, F18292, F18293, F18294, F18295, F18307, F18308, F18309, F18336, F18337, F18338, F18373, F18374, F18375, F18387, F18388, F18389, F18400, F18401, F18415, F18416, F18417, F18449, F18450, F18451, F18457, F18458, F18459, F18460, F18461, F18479, F18480, F18481, F18482, F18483, F18484, F18485, F18486, F18487, F18488, F18513, F18514, F18515, F18516, F18517, F18518, F18519, F18520, F18521, F18550, F18551, F18552, F18553
Other devices under the same FDA submission
9 devices cleared under K103305 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- RUBY Coil RBY2C3260-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION100 reports
- Penumbra System DH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION72 reports
- Penumbra Coil 400 4006C0306Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION22 reports
- Penumbra Coil 400 4002C0620Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION7 reports
- Penumbra Coil 400 4002C0312Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION4 reports
- Penumbra Coil 400 4002C0512Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION2 reports
- Penumbra Coil 400 4002C1335Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION2 reports
- Penumbra Coil 400 4002C0415Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page