Posterior Referencing Instrumentation for the NEXGEN… recallZ-1993-2011 · Zimmer Inc. · initiated Mar 14, 2011
Terminated All recalls posted in April 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the instruments after cleaning and sterilization cycles and 5 resulted in a delay of surgery.
FDA root cause category: Device Design
What the company is doing
The firm, Zimmer, Inc., sent an "URGENT MEDICAL DEVICE RECALL" letter dated March 14, 2011 to its consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: Instructions for Handling of Affected Product: 1) Stop using the devices and to quarantine immediately 2) Carry out a physical count of all affect product in their territory and record data on Inventory Identification Form included with this letter-Note: The instruments must be returned even if the instrument is not exhibiting the black powder 3)Ensure they have provided acknowledgement via "email return receipt" upon receiving this notification 4) Fax copy of completed form to (574) 372-4265 5) Return the recalled product along with the completed Inventory Identification Form and the Sterilization Certification Form to: Zimmer Distribution Center, Attn: Product Service, 1777 West Center Street, Warsaw, IN 46580 A new or refurbished instrument will be provided as replacement for their returned product once the inventory is available. Instructions for Health Care Professional Notification 1) Provide a list of all hospital that currently use or have used the affected product listed in notice by utilizing the User Facility Documentation Form and the Health Care Professional Documentation Form 2) Fax a copy of the completed form(s) to: Zimmer, Inc. at (574) 372-4265 Zimmer will provide a similar notification via FedEx to these hospitals making them aware of this recall and provide a "Dear Doctor" letter for their surgeons. Important: If you have distributed these affected lots further, please provide a copy of this letter to these customers when you implement these recall instructions. For shipping assistance, questions or assistance in notifying your accounts about the recall, please contact Zimmer, Inc., at 1-800-613-6131. ADDENDUM INFO The information listed above was updated to include the additional 5
Recall details
- Recall number
- Z-1993-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 14, 2011
- Posted by the FDA
- Apr 14, 2011
- Recalling firm
- Zimmer Inc.
- Quantity
- 323+409 (in addendum)=732
- Product code
- MBH
- FDA submissions
- K031061
- Distribution
- Worldwide distribution: USA (nationwide) and countries including: Australia, Canada, Germany, Japan, New Delphi and Singapore. Addendum: Finland, Switzerland, UK, Denmark, Spain, Portugal, France, Netherlands, Italy, South Africa, Norway, Spain, and Sweden.
Product description
Posterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION EXTRAMEDULLARY MODULAR SPIKE ARM POST, REF 00-5901-074-00, NONSTERILE, Zimmer UK, Ltd.,, Zimmer Warsaw,IN This instrument is attached to the femoral implant or provisional to aid in positioning and inserting the components.
Code information (models, lots, serial numbers)
REF 00-5901-074-00, Lots 61488062, 61549223, 61584526, 61593951, 61516298 and 61517186; and according to the firm's addendum dated 6/13/11,61634588 and 61625084.
Other devices under the same FDA submission
6 devices cleared under K031061 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NexGen® 5952-15-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- NexGen® 5952-15-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall0 reports
- NexGen® 5952-15-05Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 recall0 reports
- NexGen® 5952-15-06Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- NexGen® 5952-14-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- NexGen® 5972-16-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 58225).
- Z-1985-2011 · TerminatedPosterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION, FEMORAL AND PROVISIONAL IMPACTOR/EXTRACTOR, REF 00-5901-026-00, Zimmer UK, Ltd.,, Zimmer Warsaw,IN Thi…
- Z-1986-2011 · TerminatedPosterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION POSTERIOR REFERENCING SIZER, REF 00-5901-040-00, NONSTERILE, Zimmer UK, Ltd.,, Zimmer Warsaw,IN This in…
- Z-1987-2011 · TerminatedPosterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION POSTERIOR REFERENCING SIZER Boom, REF 00-5901-040-10, NONSTERILE, Zimmer UK, Ltd.,, Zimmer Warsaw,IN Th…
- Z-1988-2011 · TerminatedPosterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION PATELLA PEG DRILL GUIDE HANDLE, REF 00-5903-060-00, NONSTERILE, Zimmer UK, Ltd.,, Zimmer Warsaw,IN This…
- Z-1989-2011 · TerminatedPosterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION QUICK-CONNECT HANDLE, REF 00-5901-034-00, NONSTERILE, Zimmer UK, Ltd.,, Zimmer Warsaw,IN This instrumen…
- Z-1990-2011 · TerminatedPosterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION MODULAR T-HANDLE, REF 00-5901-061-00, NONSTERILE, Zimmer UK, Ltd.,, Zimmer Warsaw,IN This instrument is…
- Z-1991-2011 · TerminatedPosterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION EXTRAMEDULLARY DISTAL TELESCOPING TUBE, REF 00-5901-071-00, NONSTERILE, Zimmer UK, Ltd.,, Zimmer Warsaw,…
- Z-1992-2011 · TerminatedPosterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION EXTRAMEDULLARY PROXIMAL TELESCOPING ROD, REF 00-5901-072-00, NONSTERILE, Zimmer UK, Ltd.,, Zimmer Warsaw…
- Z-1994-2011 · TerminatedPosterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION DISTAL FEMORAL RESECTION GUIDE, REF 00-5901-063-00, NONSTERILE, Zimmer UK, Ltd.,, Zimmer Warsaw,IN This…
- Z-1995-2011 · TerminatedPosterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION FEMORAL VALGUS ALIGNMENT GUIDE, REF 00-5901-067-00, NONSTERILE, Zimmer UK, Ltd.,, Zimmer Warsaw,IN This…
- Z-1996-2011 · TerminatedPosterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION PATELLA CLAMP, REF 00-5903-020-00, NONSTERILE, Zimmer UK, Ltd.,, Zimmer Warsaw,IN This instrument is at…
- Z-1997-2011 · TerminatedPosterior Referencing Instrumentation for the NEXGEN COMPLETE KNEE SOLUTION PATELLA OSTEOTOMY GUIDE, REF 00-5903-010-00, NONSTERILE, Zimmer UK, Ltd.,, Zimmer Warsaw,IN This instru…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page