R3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 36 MM O.D.… recallZ-2055-2011 · Smith & Nephew Inc · initiated Mar 11, 2011

Terminated All recalls posted in April 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smith & Nephew Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During the manufacturing process for several batches of R3 Ceramic Liners, the titanium rings were pressed onto the ceramic component with a higher force than allowed by manufacturing specifications. This has the potential to result in lower than expected strength for the liners.

FDA root cause category: Employee error

What the company is doing

All affected Smith & Nephew Distributors were notified via email and the hospitals were notified by letter delivered by Fed-Ex on 3/11/2011. They were instructed to immediately cease distribution, quarantine affected product for return, if product further distributed, identify customers and immediately notify them of recall. Please contact Mr. David Archer, Group Director, at 701.396.2121, ext. 5373, if you have questions.

Recall details

Recall number
Z-2055-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 11, 2011
Posted by the FDA
Apr 22, 2011
Recalling firm
Smith & Nephew Inc
Quantity
563 units
Product code
MRA
FDA submissions
P030022
Distribution
AK, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OR, PA, SC, TN, TX, WA, and WI

Product description

R3 TM CERAMIC ACETABULAR LINER, BIOLOX FORTE, 36 MM O.D. 52 MM O.D., REF 71338952, STERILE R, QTY (1), Smith & Nephew, Inc., Memphis, TN 38116, 2017/06

Code information (models, lots, serial numbers)

09FT32383, 09FT32504, 09GT33518, 09GT33519, 09GT33918, 09GT33919, 09GT34111, 09GT34112, 09KT36378, and 09MT38082

Other devices under the same FDA submission

12 devices cleared under P030022 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 58324).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page