Skip to content

ExactaMix EVA Container, 500mL, REF: 738, Sterile R, Baxa… recallZ-1964-2011 · Baxa Corporation · initiated Mar 8, 2011

Terminated All recalls posted in April 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxa Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

TPN bags may leak during filling operation, due to bag manufacturing error.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The firm, BAXA, sent a "ExactaMix Bag Product Notice" dated March 8, 2011 to its customers. The notice described the product, problem and actions to be taken. The customers were instructed to do the following: If they do not intend to use the manual addition port, verify that the cap is firmly seated onto the port by gently tugging on the cap. If they do intend to use the manual addition port, they should snap off the additive port cap prior to compounding the bag. If they encounter a faulty additive port, contact Baxa Technical Support to receive credit and/or replacement on defective product only. To return affected bags, follow the instructions provided in the notice if the bags were purchased directly from Baxa or purchased from wholesaler, distributor or other 3rd party vendor. Additional copies of the return form can be retrieved by going to www.baxa.com/ebug OR by calling Baxa Technical Support at 800-678-2292. If you have any questions or concerns, contact the Sr. Vice President, Quality and Regulatory at (303) 400-2838 or Baxa Technical Support at 800-678-2292.

Recall details

Recall number
Z-1964-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 8, 2011
Posted by the FDA
Apr 12, 2011
Recalling firm
Baxa Corporation
Quantity
1,611,150 bags
Product code
LHI
FDA submissions
K900585
Distribution
Worldwide distribution: USA (nationwide) and countries including: Argentina, Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, Hong Kong, Ireland, Japan, Malaysia, Mexico, New Zealand, Panama, Puerto Rico, Singapore, South Korea, Taiwan, UK, Uruguay, Venezuela.

Product description

ExactaMix EVA Container, 500mL, REF: 738, Sterile R, Baxa Corporation, Englewood, CO 80112. A plastic container used to hold large-volume sterile solutions to be administered to a patient through an intravascular administration set

Code information (models, lots, serial numbers)

All codes sold since 7/23/2009.

Other devices under the same FDA submission

8 devices cleared under K900585 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 58304).