AVOXimeter 1000, AVOXimeter 4000 recallZ-2218-2011 · International Technidyne Corp. · initiated Mar 23, 2011

Terminated All recalls posted in May 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from International Technidyne Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An incorrect calibration code that affects the total hemoglobin measurement was applied to affected Avoximeter instruments.

FDA root cause category: Device Design

What the company is doing

International Technidyne Corp. sent a letter dated March 23, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory to determine if they have any of the affected product. If they have any of the affected product, customers were instructed to stop using it and remove it from their inventory. Customers were also instructed to complete the attached Customer Account Tracking Form and return it by fax to (732-635-0144) or e-mail ([email protected]) or mail to: ITC Technical Support 20 Corporate Place South Piscataway, New Jersey 08854 For any questions call the Distributor or ITC Nexus Dx Technical Support at 800-631-5945 (US) or 732-548-5700 (International), ext 4707.

Recall details

Recall number
Z-2218-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 23, 2011
Posted by the FDA
May 19, 2011
Recalling firm
International Technidyne Corp.
Quantity
17
Product code
DQA
FDA submissions
K951485
Distribution
Worldwide Distribution - USA including TX, KS, OH, OR, FL, GA, NY, MA, and DC and the countries of Taiwan and Thailand

Product description

AVOXimeter 1000, AVOXimeter 4000 1000- portable bedside whole blood oximeter that directly measures oxyhemoglobin fraction, the total hemoglobin concentration, and oxygen content. It has enhanced software that optimizes its use int he cardiac catheterization laboratory through calculations of physiologic equations. 4000-a portable bedside whole blood CO-oximeter that directly measures total hemoglobin concentration and the relative concentrations of oxyhemoglobin carboxyhemoglobin concentration methemoglobin. In addition, oxygen content oxygen capacity and oxygen saturation indices are calculated.

Code information (models, lots, serial numbers)

AVOXimeter 1000: K922075 AVOX1000, AVOX1000E-110, AVOX1000INT  AVOXimeter 4000: K951485 AVOX4000, AVOX4000D

Other devices under the same FDA submission

1 device cleared under K951485 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.