Cook Medical Zenith Flex AAA Endovascular Graft Iliac Leg… recallZ-3144-2011 · Cook, Inc. · initiated Apr 15, 2011
Terminated All recalls posted in September 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cook, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm initiated this action in response to information that there may be potential damage in the braided suture which is used to attach the external stents to the graft material.
FDA root cause category: Process control
What the company is doing
The firm sent URGENT PRODUCT RECALL notices dated 4/15/2011 to Risk Managers/Recall Administration (Chief of Vascular Surgery and or Interventional Radiology). The letters explained the problem with the devices and the action to be taken. Consignees were instructed to return the recalled devices to Cook Medical for credit or replacement. Customers with questions or concerns can contact the firm at 800-457-2235.
Recall details
- Recall number
- Z-3144-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 15, 2011
- Posted by the FDA
- Sep 6, 2011
- Recalling firm
- Cook, Inc.
- Quantity
- 13 units
- Product code
- MIH
- FDA submissions
- P020018
- Distribution
- US, Canada, Japan
Product description
Cook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-16-90-ZT, REF G48463, 16Fr ID/ 6.0 mm OD, Sterile.
Code information (models, lots, serial numbers)
F2629502
Other devices under the same FDA submission
36 devices cleared under P020018 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Zenith G32551William A. Cook Australia Pty. Ltd. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT5 deaths44 reports
- Zenith G32543William A. Cook Australia Pty. Ltd. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT6 deaths18 reports
- Zenith G32554William A. Cook Australia Pty. Ltd. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 deaths17 reports
- Zenith G32542William A. Cook Australia Pty. Ltd. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 death15 reports
- Zenith G32539William A. Cook Australia Pty. Ltd. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT14 reports
- Zenith G32537William A. Cook Australia Pty. Ltd. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 death13 reports
- Zenith G32538William A. Cook Australia Pty. Ltd. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT4 deaths13 reports
- Zenith G32544William A. Cook Australia Pty. Ltd. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 deaths9 reports
- Zenith G32533William A. Cook Australia Pty. Ltd. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 death8 reports
- Zenith G32532William A. Cook Australia Pty. Ltd. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT7 reports
- Zenith G32550William A. Cook Australia Pty. Ltd. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 deaths7 reports
- Zenith G32534William A. Cook Australia Pty. Ltd. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 deaths6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 58613).
- Z-3130-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-24-82-ZT, REF G48407, 18 Fr ID/ 7.1 mm OD, Sterile
- Z-3131-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-24-96-ZT, REF G48408, 18 Fr ID/ 7.1 mm OD, Sterile
- Z-3132-2011 · TerminatedCook Medical, Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-26-96-ZT, REF G48413, 18 Fr ID/ 7.1 mm OD, Sterile.
- Z-3133-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-26-125-ZT, REF G48415, 18 Fr ID/ 7.1 mm OD, Sterile.
- Z-3134-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-28-96-ZT, REF G48418, 20 Fr ID/ 7.7 mm OD, Sterile
- Z-3135-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-30-82-ZT, REF G48422, 20 Fr ID/ 7.7 mm OD
- Z-3136-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-32-82-ZT, REF G48427, 20 Fr ID/ 7.7 mm OD, Sterile.
- Z-3137-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-32-96-ZT, REF G48428, 20 Fr ID/ 7.7 mm OD, Sterile.
- Z-3138-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Main Body, REF TFFB-36-95-ZT, REF G48432, 22 Fr ID/ 8.5 mm OD, Sterile.
- Z-3139-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-10-71-ZT, REF G48444, 14 Fr ID/ 5.3 mm OD, Sterile.
- Z-3140-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-12-56-ZT, REF G48449, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3141-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-14-56-ZT, REF G48455, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3142-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-16-56-ZT, REF G48461, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3143-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-16-73-ZT, REF G48462, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3145-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-18-56-ZT, REF G48465, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3146-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-18-73-ZT, REF G48466, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3147-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-56-ZT, REF G48469, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3148-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-73-ZT, REF G48470, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3149-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-90-ZT, REF G48471, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3150-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-56-ZT, REF G48473, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3151-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-22-73-ZT, REF G48474, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3152-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-73-ZT, REF G48476, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3153-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-20-56-ZT, REF G48477, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3154-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-24-73-ZT, REF G48478, 16Fr ID/ 6.0 mm OD, Sterile.
- Z-3155-2011 · TerminatedCook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-24-90-ZT, REF G48479, 16 Fr. ID/ 6.0 OD, Sterile.
- Z-3156-2011 · TerminatedCook Medical Zenith Flex AAA Main Body Extension, REF ESBE-22-29, REF G32179, 18 Fr. ID/ 6.90 OD, Sterile.
- Z-3157-2011 · TerminatedCook Medical Zenith Flex AAA Main Body Extension, REF ESBE-30-39, REF G32183, 20 Fr. ID/ 7.7 OD, Sterile.
- Z-3158-2011 · TerminatedCook Medical Zenith Flex AAA Main Body Extension, REF ESBE-32-39, REF G32184, 20 Fr. ID/ 7.7 mm OD, Sterile.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page