Skip to content

Cook Medical Zenith Flex AAA Endovascular Graft Iliac Leg… recallZ-3145-2011 · Cook, Inc. · initiated Apr 15, 2011

Terminated All recalls posted in September 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cook, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm initiated this action in response to information that there may be potential damage in the braided suture which is used to attach the external stents to the graft material.

FDA root cause category: Process control

What the company is doing

The firm sent URGENT PRODUCT RECALL notices dated 4/15/2011 to Risk Managers/Recall Administration (Chief of Vascular Surgery and or Interventional Radiology). The letters explained the problem with the devices and the action to be taken. Consignees were instructed to return the recalled devices to Cook Medical for credit or replacement. Customers with questions or concerns can contact the firm at 800-457-2235.

Recall details

Recall number
Z-3145-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 15, 2011
Posted by the FDA
Sep 6, 2011
Recalling firm
Cook, Inc.
Quantity
6 units
Product code
MIH
FDA submissions
P020018
Distribution
US, Canada, Japan

Product description

Cook Medical Zenith Flex AAA Endovascular Graft Iliac Leg, REF TFLE-18-56-ZT, REF G48465, 16Fr ID/ 6.0 mm OD, Sterile.

Code information (models, lots, serial numbers)

F2629568, F2629575, F2629576, F2629577, F2629581, F2629594.

Other devices under the same FDA submission

36 devices cleared under P020018 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 58613).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page