Regenerex (TM) Ringloc +, Solid Acetablular Cup w/Apical… recallZ-2451-2011 · Biomet, Inc. · initiated Apr 14, 2011
Terminated All recalls posted in June 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
FDA root cause category: Employee error
What the company is doing
Biomet Orthopedics sent an URGENT MEDICAL DEVICE RECALL NOTICE DATED April 14, 2011, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Immediately locate and remove the affected product. Follow the instructions on the enclosed "FAX Back Response Form" and fax a copy of the Response Form to 574-372-1683 prior to return of product. Customers were to confirm receipt of the notice by calling 800-348-9500, ext 3755 or 3756. For any questions regarding this recall call 574-371-3755 or 574-371-3756.
Recall details
- Recall number
- Z-2451-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 14, 2011
- Posted by the FDA
- Jun 8, 2011
- Recalling firm
- Biomet, Inc.
- Quantity
- 16
- Product code
- LPH
- FDA submissions
- K070369
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Australia, Netherlands and Japan.
Product description
Regenerex (TM) Ringloc +, Solid Acetablular Cup w/Apical Hole, 62 mm Size 25, w/Lock Ring w/Plug, sterile, REF PT-104062, Biomet Orthopedics, Warsaw, IN. Intended Use: The Regenerex Ringloc is intended to replace the natural acetabulum damaged by disease, trauma or revision of previous arthroplasty. Indications For Use: 1 . Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2. Rheumatoid arthritis 3. Correction of functional deformity 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques. 5. Revision procedures where other treatment or devices have failed.
Code information (models, lots, serial numbers)
Lot numbers M299860 and M520400.
Other devices under the same FDA submission
5 devices cleared under K070369 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- REGENEREX RINGLOC + HIP SYSTEM PT-116056Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous1 report
- REGENEREX RINGLOC + HIP SYSTEM PT-116054Biomet Orthopedics, LLC · Prosthesis, hip, femoral component, cemented, metal1 report
- RINGLOC + HIP SYSTEM 106024Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
- RingLoc + Hip System 106021Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 report
- Regenerex® Ringloc®+ Hip System PT-106068Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained, uncemented, metal / polymer, non-porous, calcium phosphate1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 58663).
- Z-2421-2011 · TerminatedRingloc + Acetabular Shell limited hole finned 46 mm Size 22 porous coat, for uncemented use, sterile, REF 16-104146, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plas…
- Z-2422-2011 · TerminatedRingloc + Acetabular Shell limited hole finned 48 mm Size 22 porous coat, for uncemented use, sterile, REF 16-104148, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plas…
- Z-2423-2011 · TerminatedRingloc + Acetabular Shell limited hole finned 50 mm Size 23 porous coat, for uncemented use, sterile, REF 16-104150, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plas…
- Z-2424-2011 · TerminatedRingloc + Acetabular Shell limited hole finned 52 mm Size 23 porous coat, for uncemented use, sterile, REF 16-104152, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plas…
- Z-2425-2011 · TerminatedRingloc + Acetabular Shell limited hole finned 54 mm Size 24 porous coat, for uncemented use, sterile, REF 16-104154, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plas…
- Z-2426-2011 · TerminatedRingloc + Acetabular Shell limited hole finned 56 mm Size 24 porous coat, for uncemented use, sterile, REF 16-104156, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plas…
- Z-2427-2011 · TerminatedRingloc + Acetabular Shell limited hole finned 58 mm Size 25 porous coat, for uncemented use, sterile, REF 16-104158, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plas…
- Z-2428-2011 · TerminatedRingloc + Acetabular Shell limited hole finned 60 mm Size 25 porous coat, for uncemented use, sterile, REF 16-104160, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plas…
- Z-2429-2011 · TerminatedRingloc + Acetabular Shell limited hole finned 62 mm Size 26 porous coat, for uncemented use, sterile, REF 16-104162, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plas…
- Z-2430-2011 · TerminatedRingloc + Acetabular Shell limited hole finned 64 mm Size 26 porous coat, for uncemented use, sterile, REF 16-104164, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plas…
- Z-2431-2011 · TerminatedRingloc + Acetabular Shell limited hole finned 66 mm Size 27 porous coat, for uncemented use, sterile, REF 16-104166, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plas…
- Z-2432-2011 · TerminatedRingloc + Acetabular Shell limited hole finned 70 mm Size 28 porous coat, for uncemented use, sterile, REF 16-104170, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plas…
- Z-2433-2011 · TerminatedRingloc + Acetabular Shell limited hole, 46 mm Size 22 porous coat, for uncemented use, sterile, REF 16-116046, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Spr…
- Z-2434-2011 · TerminatedRingloc + Acetabular Shell limited hole, 48 mm Size 22, porous coat, for uncemented use, sterile, REF 16-116048, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2435-2011 · TerminatedRingloc + Acetabular Shell limited hole, 50 mm Size 23, porous coat, for uncemented use, sterile, REF 16-116050, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2436-2011 · TerminatedRingloc + Acetabular Shell limited hole, 52 mm Size 23, porous coat, for uncemented use, sterile, REF 16-116052, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2437-2011 · TerminatedRingloc + Acetabular Shell limited hole, 54 mm Size 24, porous coat, for uncemented use, sterile, REF 16-116054, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2438-2011 · TerminatedRingloc + Acetabular Shell limited hole, 56 mm Size 24, porous coat, for uncemented use, sterile, REF 16-116056, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2439-2011 · TerminatedRingloc + Acetabular Shell limited hole, 58 mm Size 25, porous coat, for uncemented use, sterile, REF 16-116058, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2440-2011 · TerminatedRingloc + Acetabular Shell limited hole, 60 mm Size 25, porous coat, for uncemented use, sterile, REF 16-116060, Biomet Orthopedics, Warsaw, IN.
- Z-2441-2011 · TerminatedRingloc + Acetabular Shell limited hole, 62 mm Size 26, porous coat, for uncemented use, sterile, REF 16-116062, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2442-2011 · TerminatedRingloc + Acetabular Shell limited hole, 64 mm Size 26, porous coat, for uncemented use, sterile, REF 16-116064, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2443-2011 · TerminatedRingloc + Acetabular Shell limited hole, 66 mm Size 27, porous coat, for uncemented use, sterile, REF 16-116066, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2444-2011 · TerminatedRingloc + Acetabular Shell limited hole, 68 mm Size 27, porous coat, for uncemented use, sterile, REF 16-116068, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2445-2011 · TerminatedRingloc + Acetabular Shell limited hole, 70 mm Size 28, porous coat, for uncemented use, sterile, REF 16-116070, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2446-2011 · TerminatedRingloc + Acetabular Shell limited hole, 72 mm Size 28, porous coat, for uncemented use, sterile, REF 16-116072, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2447-2011 · TerminatedRingloc + Acetabular Shell limited hole, 76 mm Size 28, porous coat, for uncemented use, sterile, REF 16-116076, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous Plasma Sp…
- Z-2448-2011 · TerminatedRegenerex (TM) Ringloc +, Solid Acetablular Cup w/Apical Hole, 50 mm Size 22, w/Lock Ring w/Plug, sterile, REF PT-104050, Biomet Orthopedics, Warsaw, IN. Intended Use: The Porous …
- Z-2449-2011 · TerminatedRegenerex (TM) Ringloc +, Solid Acetablular Cup w/Apical Hole, 52 mm Size 23, w/Lock Ring w/Plug, sterile, REF PT-104052, Biomet Orthopedics, Warsaw, IN. Intended Use: The Regenere…
- Z-2450-2011 · TerminatedRegenerex (TM) Ringloc +, Solid Acetablular Cup w/Apical Hole, 56 mm Size 24, w/Lock Ring w/Plug, sterile, REF PT-104056, Biomet Orthopedics, Warsaw, IN. Intended Use: The Regenere…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page