Skip to content

Regenerex (TM) Ringloc +, Solid Acetablular Cup w/Apical… recallZ-2452-2011 · Biomet, Inc. · initiated Apr 14, 2011

Terminated All recalls posted in June 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.

FDA root cause category: Employee error

What the company is doing

Biomet Orthopedics sent an URGENT MEDICAL DEVICE RECALL NOTICE DATED April 14, 2011, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Immediately locate and remove the affected product. Follow the instructions on the enclosed "FAX Back Response Form" and fax a copy of the Response Form to 574-372-1683 prior to return of product. Customers were to confirm receipt of the notice by calling 800-348-9500, ext 3755 or 3756. For any questions regarding this recall call 574-371-3755 or 574-371-3756.

Recall details

Recall number
Z-2452-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 14, 2011
Posted by the FDA
Jun 8, 2011
Recalling firm
Biomet, Inc.
Quantity
8
Product code
LPH
FDA submissions
K070369
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Australia, Netherlands and Japan.

Product description

Regenerex (TM) Ringloc +, Solid Acetablular Cup w/Apical Hole, 64 mm Size 26, w/Lock Ring w/Plug, sterile, REF PT-104064, Biomet Orthopedics, Warsaw, IN. Intended Use: The Regenerex Ringloc is intended to replace the natural acetabulum damaged by disease, trauma or revision of previous arthroplasty. Indications For Use: 1 . Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2. Rheumatoid arthritis 3. Correction of functional deformity 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques. 5. Revision procedures where other treatment or devices have failed

Code information (models, lots, serial numbers)

Lot number M299890.

Other devices under the same FDA submission

5 devices cleared under K070369 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 58663).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page