Medical device recalls: July 2007
The FDA posted 143 medical device recall records in July 2007, from 61 recall events by 51 companies. 143 have not been classified yet.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Abbott Diagnostic International, Ltd. · 4 events, 9 records
- Terumo Cardiovascular Systems Corp · 2 events, 16 records
- Baxter Healthcare Corp. · 2 events, 5 records
- Radiometer America Inc · 2 events, 3 records
- Diagnostica Stago, Inc. · 2 events, 2 records
- Medtronic Perfusion Systems · 2 events, 2 records
- Smiths Medical PM, Inc. · 2 events, 2 records
- Thoratec Corp · 2 events, 2 records
FDA root cause
- Other · 123
- Device Design · 8
- Software design · 5
- Nonconforming Material/Component · 2
- Labeling design · 2
- Component design/selection · 2
- Process control · 1
Not yet classified (143)
The FDA has posted these records but has not published a class for them yet.
- Z-1111-2007 · Baxter Healthcare Corp. · posted Jul 28, 2007 · Terminated
Falsified Repair/Test and Inspection records: Pumps submitted for repair or service may not have been properly repaired or serviced before being returned to the customer.
- Z-1112-2007 · Baxter Healthcare Corp. · posted Jul 28, 2007 · Terminated
Falsified Repair/Test and Inspection records: Pumps submitted for repair or service may not have been properly repaired or serviced before being returned to the customer.
- Z-1113-2007 · Baxter Healthcare Corp. · posted Jul 28, 2007 · Terminated
Falsified Repair/Test and Inspection records: Pumps submitted for repair or service may not have been properly repaired or serviced before being returned to the customer.
- Z-1114-2007 · Baxter Healthcare Corp. · posted Jul 28, 2007 · Terminated
Falsified Repair/Test and Inspection records: Pumps submitted for repair or service may not have been properly repaired or serviced before being returned to the customer.
- Z-1108-2007 · Thoratec Corp · posted Jul 27, 2007 · Terminated
Component Failure; Device Failure without warning; due to premature wear-out of the compressor motor, VAD support for the patient's circulatory system may fail without warning.
- Z-1109-2007 · Thoratec Corp · posted Jul 27, 2007 · Terminated
Incorrect collect nut assembly (inadvertent use of the incorrect (Black) collect nut) during surgical implantation (wrong size collect nut) to secure the inflow cannula to the PVAD has resulted in patient deaths
- Z-1110-2007 · Medtronic Perfusion Systems · posted Jul 26, 2007 · Terminated
Pyrogen Levels detected above specification: The product contains a pyrogen level above Medtronic''s internal specifications while the product is labeled as nonpyrogenic.
- Z-0240-2008 · Hill-Rom, Inc. · posted Jul 24, 2007 · Terminated
False Latch/Failure to Latch-The side rail may not latch, or the latch may not fully engage, resulting in the siderail being able to be lowered without pulling on the release handle.
- Z-1080-2007 · Synovis Surgical Inovation Div. of · posted Jul 24, 2007 · Terminated
There is a potential for the pouch to become damaged during shipping, compromising the sterility of the device inside.
- Z-1081-2007 · Synovis Surgical Inovation Div. of · posted Jul 24, 2007 · Terminated
There is a potential for the pouch to become damaged during shipping, compromising the sterility of the device inside.
- Z-1082-2007 · Synovis Surgical Inovation Div. of · posted Jul 24, 2007 · Terminated
There is a potential for the pouch to become damaged during shipping, compromising the sterility of the device inside.
- Z-1083-2007 · Synovis Surgical Inovation Div. of · posted Jul 24, 2007 · Terminated
There is a potential for the pouch to become damaged during shipping, compromising the sterility of the device inside.
- Z-1084-2007 · Synovis Surgical Inovation Div. of · posted Jul 24, 2007 · Terminated
There is a potential for the pouch to become damaged during shipping, compromising the sterility of the device inside.
- Z-1085-2007 · Synovis Surgical Inovation Div. of · posted Jul 24, 2007 · Terminated
There is a potential for the pouch to become damaged during shipping, compromising the sterility of the device inside.
- Z-1086-2007 · Thommen Medical, AG · posted Jul 24, 2007 · Terminated
The undercut on the SPI EASY abutments which engages the stabilization ring of the implant is missing. Although the abutments can still be placed on the implant by the user the parts do not fulfill the quality expectations of the firm.
- Z-1094-2007 · Radiometer America Inc · posted Jul 24, 2007 · Terminated
Software anomaly. Protein removing program incorporated in the operating software for Versions: V3.833 and V5.24, of the ABL700 and ABL800 Blood Gas Analyzers, may allow for protein build up on the electrode membranes of blood gas analyzers. This…
- Z-1095-2007 · Radiometer America Inc · posted Jul 24, 2007 · Terminated
Software anomaly. Protein removing program incorporated in the operating software for Versions: V3.833 and V5.24, of the ABL700 and ABL800 Blood Gas Analyzers, may allow for protein build up on the electrode membranes of blood gas analyzers. This…
- Z-1096-2007 · Abbott Diagnostic International, Ltd. · posted Jul 24, 2007 · Terminated
During an internal study, the Architect AUSAB assay was found to read preparations of the WHO standard up to 53% higher than the WHO assigned value. This has the potential to cause elevated results and may impact the interpretation for a small percentage…
- Z-1097-2007 · Abbott Diagnostic International, Ltd. · posted Jul 24, 2007 · Terminated
During an internal study, the Architect AUSAB assay was found to read preparations of the WHO standard up to 53% higher than the WHO assigned value. This has the potential to cause elevated results and may impact the interpretation for a small percentage…
- Z-1098-2007 · Abbott Diagnostic International, Ltd. · posted Jul 24, 2007 · Terminated
During an internal study, the Architect AUSAB assay was found to read preparations of the WHO standard up to 53% higher than the WHO assigned value. This has the potential to cause elevated results and may impact the interpretation for a small percentage…
- Z-1099-2007 · Power Medical Interventions, Inc · posted Jul 24, 2007 · Terminated
Malfunction - improper staple formation
- Z-1100-2007 · Sunnex Inc. · posted Jul 24, 2007 · Terminated
Surgical and/or exam lamp holder may malfunction, discolor, melt and smolder.
- Z-1101-2007 · Sunnex Inc. · posted Jul 24, 2007 · Terminated
Surgical and/or exam lamp holder may malfunction, discolor, melt and smolder.
- Z-1102-2007 · Veridex, LLC · posted Jul 24, 2007 · Terminated
An increased incidence of aggregates was observed in Lot 0036 and 0036A. The presence of aggregates will cause customers to invalidate patient results.
- Z-1103-2007 · Louisville Apl Diagnostics Inc · posted Jul 24, 2007 · Terminated
Product mislabeled; conjugate IgG placed in bottles labeled as conjugate IgM and vice versa.
- Z-1104-2007 · Louisville Apl Diagnostics Inc · posted Jul 24, 2007 · Terminated
Product mislabeled; conjugate IgG placed in bottles labeled as conjugate IgM and vice versa.
- Z-1057-2007 · Gebauer Company · posted Jul 23, 2007 · Terminated
Mold contamination
- Z-1058-2007 · Gebauer Company · posted Jul 23, 2007 · Terminated
Mold contamination
- Z-1059-2007 · Gebauer Company · posted Jul 23, 2007 · Terminated
Mold contamination
- Z-1060-2007 · Gebauer Company · posted Jul 23, 2007 · Terminated
Mold contamination
- Z-1061-2007 · Gebauer Company · posted Jul 23, 2007 · Terminated
Mold contamination
- Z-1062-2007 · Gebauer Company · posted Jul 23, 2007 · Terminated
Mold contamination
- Z-1063-2007 · Encore Medical, Lp · posted Jul 23, 2007 · Terminated
Sterility/Package Integrity : Device packaging may potentially be unable to ensure sterility when exposed to shipping conditions.
- Z-1064-2007 · Encore Medical, Lp · posted Jul 23, 2007 · Terminated
Sterility/Package Integrity : Device packaging may potentially be unable to ensure sterility when exposed to shipping conditions.
- Z-1065-2007 · Encore Medical, Lp · posted Jul 23, 2007 · Terminated
Sterility/Package Integrity: Device packaging may potentially be unable to ensure sterility when exposed to shipping conditions.
- Z-1066-2007 · Encore Medical, Lp · posted Jul 23, 2007 · Terminated
Sterility/Package Integrity: Device packaging may potentially be unable to ensure sterility when exposed to shipping conditions.
- Z-1089-2007 · Smiths Medical PM, Inc. · posted Jul 23, 2007 · Terminated
Turn off unexpectedly/ Not turn on : monitors turned off unexpectedly sounding the critical failure alarm. Advisor Vital Signs monitors that were off, would not turn back on.
- Z-1090-2007 · Medtronic Perfusion Systems · posted Jul 23, 2007 · Terminated
Device failure requiring use of back up systems (malfunctioning printed circuit board (PCB))
- Z-1115-2007 · Kensey Nash Corp · posted Jul 23, 2007 · Terminated
Defective component (power supply)
- Z-1116-2007 · B. Braun Medical, Inc. · posted Jul 23, 2007 · Terminated
Cracks at the Catheter Tip
- Z-1117-2007 · Liko, Inc. · posted Jul 23, 2007 · Terminated
Sling bar may have a center bolt break at the adapter while in use
- Z-1118-2007 · Liko, Inc. · posted Jul 23, 2007 · Terminated
Sling bar may have a center bolt break at the adapter while in use
- Z-1119-2007 · Biomedicare, Inc. · posted Jul 23, 2007 · Terminated
Lack of validation for sterilization cycle: This recall was initiated because the validation of the sterilization process for these devices has not been completed. As a result, Biomedicare cannot assure the sterility of fixtures distributed to date.
- Z-1056-2007 · Abbott Diagnostic International, Ltd. · posted Jul 20, 2007 · Terminated
Incorrect configuration: kits contain reagent bottles in the order S-W-T-P instead of the correct order, W-S-T-P.
- Z-1092-2007 · Advanced Medical Optics, Inc. · posted Jul 19, 2007 · Terminated
In response to information received 05/25/2007 from the U.S. Centers for Disease Control and Prevention (CDC) regarding eye infections from Acanthamoeba, a naturally occurring water-borne organism which can contribute to serious corneal infections. CDC…
- Z-1093-2007 · Advanced Medical Optics, Inc. · posted Jul 19, 2007 · Terminated
In response to information received 05/25/2007 from the U.S. Centers for Disease Control and Prevention (CDC) regarding eye infections from Acanthamoeba, a naturally occurring water-borne organism which can contribute to serious corneal infections. CDC…
- Z-1091-2007 · Baxter Healthcare Corp. · posted Jul 18, 2007 · Terminated
A software anomaly is causing newly upgraded Colleague Triple Channel Infusion Pumps to alarm, display an error code (16:310:867:0002) and stop the infusion. This occurs during user programming with all three channels simultaneously infusing fluids. In…
- Z-1011-2007 · Teleflex Creek Dr · posted Jul 17, 2007 · Terminated
Suture Diameter/Strength specification failure: The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage.
- Z-1012-2007 · Teleflex Creek Dr · posted Jul 17, 2007 · Terminated
The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage.
- Z-1013-2007 · Teleflex Creek Dr · posted Jul 17, 2007 · Terminated
Suture Diameter/Strength specification failure: The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage.
- Z-1014-2007 · Teleflex Creek Dr · posted Jul 17, 2007 · Terminated
Suture Diameter/Strength specification failure: The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage.
- Z-1015-2007 · Teleflex Creek Dr · posted Jul 17, 2007 · Terminated
Suture Diameter/Strength specification failure: The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage.
- Z-1016-2007 · Teleflex Creek Dr · posted Jul 17, 2007 · Terminated
The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage.
- Z-1017-2007 · Teleflex Creek Dr · posted Jul 17, 2007 · Terminated
Suture Diameter/Strength specification failure: The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage.
- Z-1018-2007 · Teleflex Creek Dr · posted Jul 17, 2007 · Terminated
Suture Diameter/Strength specification failure: The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage.
- Z-1019-2007 · Teleflex Creek Dr · posted Jul 17, 2007 · Terminated
Suture Diameter/Strength specification failure: The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage.
- Z-1042-2007 · Lifescan Inc · posted Jul 13, 2007 · Terminated
Damaged Vials/Inaccurate test results: a manufacturing equipment problem on one of the packaging lines had the potential to puncture vials, causing a single hole in the side of the vial below the rim. This result of this defect may produce inaccurately…
- Z-0989-2007 · GE OEC Medical Systems, Inc · posted Jul 12, 2007 · Terminated
The Air Kerma Rate displayed value and cumulative air kerma displayed value could exceed the allowable error under certain imaging conditions.
- Z-0990-2007 · GE OEC Medical Systems, Inc · posted Jul 12, 2007 · Terminated
The Air Kerma Rate displayed value and cumulative air kerma displayed value could exceed the allowable error under certain imaging conditions.
- Z-1067-2007 · Abbott Diagnostic International, Ltd. · posted Jul 12, 2007 · Terminated
An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted then the background effect is…
- Z-1068-2007 · Abbott Diagnostic International, Ltd. · posted Jul 12, 2007 · Terminated
An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted then the background effect is…
- Z-1069-2007 · Abbott Diagnostic International, Ltd. · posted Jul 12, 2007 · Terminated
An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted then the background effect is…
- Z-1070-2007 · Abbott Diagnostic International, Ltd. · posted Jul 12, 2007 · Terminated
An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted then the background effect is…
- Z-1071-2007 · Wako Chemicals USA, Inc. · posted Jul 12, 2007 · Terminated
Diagnostic reagent for clinical laboratory testing was incorrectly labeled.
- Z-1074-2007 · Varian Medical Systems Inc · posted Jul 12, 2007 · Terminated
The wedge tray may slip out of the Clinac interface mount during gantry rotation.
- Z-1075-2007 · Cardinal Health · posted Jul 12, 2007 · Terminated
Under isolated conditions, the device may experience issues with O2 Calibration, which may result in either a nuisance alarm or a delivered FiO2 of up to 5% below the specified tolerance.
- Z-1076-2007 · Ossur · posted Jul 12, 2007 · Terminated
Some units of the Total Knee Model 2100 produced from August 1, 2006 through October 4, 2006 may contain a faulty linkage component.
- Z-1077-2007 · Dental Equipment LLC · posted Jul 12, 2007 · Terminated
Potential for light shield to melt due to light assemblies not properly rated to handle the VAC.
- Z-0992-2007 · Ev3, Inc · posted Jul 11, 2007 · Terminated
Amplatz Goose Neck Snare Kit's catheter contained in the kit may be 4 Fr instead of the intended 6 Fr.
- Z-1008-2007 · The Anspach Effort, Inc. · posted Jul 11, 2007 · Terminated
The white film portion of the sterile pouch may become brittle prior to the expiry date compromising the sterility of the product.
- Z-1023-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1024-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1025-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1026-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1027-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1028-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1029-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1030-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1031-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1032-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1033-200 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1034-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1035-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1036-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1037-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1038-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1039-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Z-1040-2007 · Abbott Laboratories · posted Jul 11, 2007 · Terminated
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
- Wallach Cryotip T-00337 Anorectal Tip used with the Cryosurgical LL100 or LL50X Cryosurgical SystemsZ-1045-2007 · Wallach Surgical Devices, Inc. · posted Jul 11, 2007 · Terminated
The tip may separate from the probe.
- Z-1046-2007 · Wallach Surgical Devices, Inc. · posted Jul 11, 2007 · Terminated
The tip may separate from the probe.
- Z-1047-2007 · Boston Scientific CRM Corp · posted Jul 11, 2007 · Terminated
Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective…
- Z-1048-2007 · Boston Scientific CRM Corp · posted Jul 11, 2007 · Terminated
Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective…
- Z-1049-2007 · Boston Scientific CRM Corp · posted Jul 11, 2007 · Terminated
Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective…
- Z-1050-2007 · Boston Scientific CRM Corp · posted Jul 11, 2007 · Terminated
Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective…
- Z-1051-2007 · Boston Scientific CRM Corp · posted Jul 11, 2007 · Terminated
Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective…
- Z-1052-2007 · Boston Scientific CRM Corp · posted Jul 11, 2007 · Terminated
Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective…
- Z-1053-2007 · Boston Scientific CRM Corp · posted Jul 11, 2007 · Terminated
Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective…
- Z-1054-2007 · Boston Scientific CRM Corp · posted Jul 11, 2007 · Terminated
Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective…
- Z-1055-2007 · Boston Scientific CRM Corp · posted Jul 11, 2007 · Terminated
Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective…
- Z-0874-2007 · Terumo Cardiovascular Systems Corp · posted Jul 10, 2007 · Terminated
The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).
- Z-0875-2007 · Terumo Cardiovascular Systems Corp · posted Jul 10, 2007 · Terminated
The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).
- Z-0876-2007 · Terumo Cardiovascular Systems Corp · posted Jul 10, 2007 · Terminated
The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).
- Z-0877-2007 · Terumo Cardiovascular Systems Corp · posted Jul 10, 2007 · Terminated
The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).
- Z-0878-2007 · Terumo Cardiovascular Systems Corp · posted Jul 10, 2007 · Terminated
The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).
- Z-0879-2007 · Terumo Cardiovascular Systems Corp · posted Jul 10, 2007 · Terminated
The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).
- Z-0880-2007 · Terumo Cardiovascular Systems Corp · posted Jul 10, 2007 · Terminated
The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).
- Z-0881-2007 · Terumo Cardiovascular Systems Corp · posted Jul 10, 2007 · Terminated
The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).
- Z-0882-2007 · Terumo Cardiovascular Systems Corp · posted Jul 10, 2007 · Terminated
The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).
- Z-0883-2007 · Terumo Cardiovascular Systems Corp · posted Jul 10, 2007 · Terminated
The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).
- Z-0884-2007 · Terumo Cardiovascular Systems Corp · posted Jul 10, 2007 · Terminated
The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).
- Z-0885-2007 · Terumo Cardiovascular Systems Corp · posted Jul 10, 2007 · Terminated
The bond may fail and a leak or separation of the device occur at the distal end (where this device connects to the arterial line).
- Z-0888-2007 · Ross Products Division Abbott Laboratories · posted Jul 10, 2007 · Terminated
The enteral pump set may have a 1-3 mm slit in the silicone tubing located between the blue connector and the stretch band Safe-T-Valve.
- GE Healthcare Innova 2100 IQ Digital Fluoroscopic Imaging System (Versatile cardiac and vascular). ↗Z-1043-2007 · General Electric Med Systems LLC · posted Jul 10, 2007 · Terminated
Caution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage of Innova Sense patient contouring…
- Z-1044-2007 · General Electric Med Systems LLC · posted Jul 10, 2007 · Terminated
Caution on the non-detection of small invasive objects/devices, which project out from the skin surfaces (e.g. biopsy or other needles) while using the Innova Sense software option designed for patient contouring. Usage of Innova Sense patient contouring…
- Z-0999-2007 · Medtronic Inc. Cardiac Rhythm Managment · posted Jul 7, 2007 · Terminated
Pacing rate; Software anomaly affecting Vitatron dual chamber C-series and T-series pacemakers, related to automatic retrograde conduction testing, can inhibit the device from pacing if the patient's intrinsic rate falls below the programmed lower rate.
- Z-1000-2007 · Medtronic Inc. Cardiac Rhythm Managment · posted Jul 7, 2007 · Terminated
Pacing rate: Software anomaly affecting Vitatron dual chamber C-series and T-series pacemakers, related to automatic retrograde conduction testing, can inhibit the device from pacing if the patient's intrinsic rate falls below the programmed lower rate.
- Z-1002-2007 · Terumo Cardiovascular Systems Corp · posted Jul 7, 2007 · Terminated
Roller Pump: Due to a circuit board malfunction the roller pumps may stop during use without displaying an error message, may stop during use with an overspeed error posted on the local display, or may fail to start/restart resulting in the pump(s) being…
- Z-1003-2007 · Terumo Cardiovascular Systems Corp · posted Jul 7, 2007 · Terminated
Roller Pump: Due to a circuit board malfunction the roller pumps may stop during use without displaying an error message, may stop during use with an overspeed error posted on the local display, or may fail to start/restart resulting in the pump(s) being…
- Z-1004-2007 · Terumo Cardiovascular Systems Corp · posted Jul 7, 2007 · Terminated
Roller Pump: Due to a circuit board malfunction the roller pumps may stop during use without displaying an error message, may stop during use with an overspeed error posted on the local display, or may fail to start/restart resulting in the pump(s) being…
- Z-1005-2007 · Terumo Cardiovascular Systems Corp · posted Jul 7, 2007 · Terminated
Due to a circuit board malfunction the roller pumps may stop during use without displaying an error message, may stop during use with an overspeed error posted on the local display, or may fail to start/restart resulting in the pump(s) being non-functional.
- Z-1007-2007 · Medtronic MiniMed · posted Jul 7, 2007 · Terminated
Exposure to Magnetic Resonance Imaging (MRI) has resulted in damage to the component that monitors and controls movement of the motor in the MiniMed Paradigm insulin infusion pump. Although there were alarms as a result of the damage, some users cleared…
- Z-0886-2007 · Medtronic Emergency Response Systems, Inc. · posted Jul 4, 2007 · Terminated
Defibrillator internal battery cell may contain a short which could reduce battery capacity and cause the device to shut down without warning while operating on backup (DC) battery power.
- Z-0887-2007 · Medtronic Emergency Response Systems, Inc. · posted Jul 4, 2007 · Terminated
Defibrillator internal battery cell may contain a short which could reduce battery capacity and cause the device to shut down without warning while operating on backup (DC) battery power.
- Z-0966-2007 · Invacare Corporation · posted Jul 4, 2007 · Terminated
Software error may permit the chair to recline too far.
- Z-0991-2007 · Smiths Medical PM, Inc. · posted Jul 4, 2007 · Terminated
Smiths Medical is conducting a recall of 93 Pneupac VR1 Emergency & Transport Ventilators following the identification of a problem which may exist with the demand valve that could cause the ventilator to stop cycling & to deliver continuous flow output to…
- Z-0993-2007 · Biomet, Inc. · posted Jul 4, 2007 · Terminated
The syringe assembly has an obstruction in the tube that prevents the plunger from being depressed.
- Z-0994-2007 · Biomet, Inc. · posted Jul 4, 2007 · Terminated
The syringe assembly has an obstruction in the tube that prevents the plunger from being depressed.
- Z-0995-2007 · Oasis Medical Inc · posted Jul 4, 2007 · Terminated
This voluntary recall has been initiated due to the product puncturing the packing during rough handling. Once the packaging is damaged, the product is no longer sterile as labeled.
- Z-0996-2007 · Oasis Medical Inc · posted Jul 4, 2007 · Terminated
This voluntary recall has been initiated due to the product puncturing the packing during rough handling. Once the packaging is damaged, the product is no longer sterile as labeled.
- Z-0997-2007 · Oasis Medical Inc · posted Jul 4, 2007 · Terminated
This voluntary recall has been initiated due to the product puncturing the packing during rough handling. Once the packaging is damaged, the product is no longer sterile as labeled.
- Z-0998-2007 · Oasis Medical Inc · posted Jul 4, 2007 · Terminated
This voluntary recall has been initiated due to the product puncturing the packing during rough handling. Once the packaging is damaged, the product is no longer sterile as labeled.
- Z-0717-2007 · Carestream Health, Inc. · posted Jul 3, 2007 · Terminated
The x-ray systems do not indicate the field size when the system's detector is turned off and the collimator is in manual mode. This results in the systems being noncompliant with the field indication performance standards of section 1020.31(e)(2)& (3) of…
- Z-0853-2007 · Diagnostica Stago, Inc. · posted Jul 3, 2007 · Terminated
A dose-hook effect is observed starting from VWF:AG levels of 300%, while the package insert indicates that no dose-hook effect has been observed with VWF:Ag levels up to 800%.
- Z-0928-2007 · OEC Medical Systems, Inc · posted Jul 3, 2007 · Terminated
X-ray systems did not have the required labeling to indicate Air Kerma Rate and cumulative Air Kerma during fluoroscopic procedures
- Z-0958-2007 · Abbott Diagnostic International, Ltd. · posted Jul 3, 2007 · Terminated
A decrease in mP values (calibration curve Net Polarization value) has been observed with this reagent lot at all calibrator levels and may produce the following results: Calibration Error Code 148 (Polarization Too Small), Calibration Error Code 172…
- Z-0975-2007 · Radiometer America Inc · posted Jul 3, 2007 · Terminated
The clinitubes are not able to be filled due to contamination with silicone during production.
- Z-0979-2007 · Diagnostica Stago, Inc. · posted Jul 3, 2007 · Terminated
Based on the confirmed clotting times in the upper part of the assigned ranges of STA-Coag Control ABN control plasmas when used with STA-Neoplastine CI Plus, use of this lot may cause difficulty during the validation of the STA-Coag Control ABN.
- Z-0987-2007 · Philips Medical Systems · posted Jul 3, 2007 · Terminated
Potential for erroneous measurements which may lead to incorrect diagnosis based upon the presence of an inaccurate, undetectable error in clinical measurement or calculation. If the user exits an imaging mode while Active Annotation is in use, the…
- Z-0988-2007 · Medispectra Inc. · posted Jul 3, 2007 · Terminated
Power supply may fail and render the unit inoperable.
- Z-1001-2007 · Bayer Healthcare LLC · posted Jul 3, 2007 · Terminated
Incorrect Unit of Measure displayed : The meters report results in the wrong unit of measurement for Canadian users.
- Z-1006-2007 · Resmed Corporation · posted Jul 3, 2007 · Terminated
S8 devices manufactured between July 2004 and May 15, 2006 are susceptible to fatigue in the connection of the AC appliance inlet connector and the power supply (power cord) leading to possible failure (short circuit, electrical failure, sparks, smoke).
- Z-1010-2007 · Brainlab AG · posted Jul 3, 2007 · Terminated
Target Alignment Error : A mechanical component/software incompatibility caused by a specific combination of the BrainLAB Target Positioner version 40700-3A for Leksell headrings and BrainLAB planning software resulted in a 1.25 mm shift in target area…
- Z-1041-2007 · Abbott Laboratories, Inc · posted Jul 3, 2007 · Terminated
Product's hinged lid can fall, if not fully opened as designed, representing a potential injury hazard to the operator.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.