Medical device recalls: December 2020
The FDA posted 41 medical device recall records in December 2020, from 6 recall events by 6 companies. 21 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Cook Inc. · 1 event, 21 records, 21 Class I
- Biomet, Inc. · 1 event, 16 records
- Galt Medical Corporation · 1 event, 1 record
- GE Healthcare, LLC · 1 event, 1 record
- Remote Diagnostic Technologies Ltd. · 1 event, 1 record
- ROi CPS LLC · 1 event, 1 record
FDA root cause
- Process control · 38
- Use error · 1
- Software design · 1
- Labeling Change Control · 1
Class I recalls (21)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0591-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0592-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0593-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0594-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0595-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0596-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0597-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0598-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0599-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0600-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0601-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0602-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0603-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0604-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0605-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0606-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0607-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0608-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0609-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0610-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
- Z-0611-2021 · Cook Inc. · posted Dec 23, 2020 · Terminated
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited…
Class II recalls (20)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0616-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0617-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0618-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0619-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0620-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0621-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0622-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0623-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0624-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0625-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0626-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0627-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0628-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0629-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0630-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0631-2021 · Biomet, Inc. · posted Dec 17, 2020 · Terminated
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
- Z-0568-2021 · Galt Medical Corporation · posted Dec 14, 2020 · Terminated
Due to a manufacturing issue, the catheter tip geometry is incorrect (thicker wall) resulting in a more obtuse tip angle.
- Z-0564-2021 · Remote Diagnostic Technologies Ltd. · posted Dec 11, 2020 · Terminated
A software error was detected within software version 1.3.4 for the Tempus LS device. When exiting pacer back into manual mode the pacer pulse wave may put the ECG calibration out of calibration. This software error may result in a delay in therapy, as the…
- MAC VU360, ElectrocardiographClass IIZ-0543-2021 · GE Healthcare, LLC · posted Dec 7, 2020 · Terminated
Incorrect patient identification and/or patient demographic errors.
- Z-0524-2021 · ROi CPS LLC · posted Dec 2, 2020 · Terminated
Labeling error, incorrect expiration date.