Medical device recalls: August 2021
The FDA posted 10 medical device recall records in August 2021, from 8 recall events by 7 companies. 4 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Baxter Healthcare Corporation · 2 events, 2 records, 2 Class I
- Boston Scientific Corporation · 1 event, 2 records
- Exactech, Inc. · 1 event, 2 records
- BioMedical Equipment Service Co (BMES) · 1 event, 1 record, 1 Class I
- Cardinal Health 200, LLC · 1 event, 1 record, 1 Class I
- Qura S.R.L. · 1 event, 1 record
- Zeltiq Aesthetics, Inc · 1 event, 1 record
FDA root cause
- Software design · 3
- Under Investigation by firm · 2
- Component change control · 2
- Material/Component Contamination · 1
- Labeling Change Control · 1
- Component design/selection · 1
Class I recalls (4)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-2244-2021 · Cardinal Health 200, LLC · posted Aug 13, 2021 · Terminated
The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the scalpel once the safety shield is closed…
- Z-2165-2021 · Baxter Healthcare Corporation · posted Aug 11, 2021 · Terminated
Software issue: The defect creates a mismatch between linked drug identifiers (IDs) in the Dose IQ user interface (UI) and the binary drug library (BDL) loaded onto the Spectrum IQ pump.
- Z-2110-2021 · Baxter Healthcare Corporation · posted Aug 6, 2021 · Terminated
System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.
- Z-2204-2021 · BioMedical Equipment Service Co (BMES) · posted Aug 6, 2021 · Terminated
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free…
Class II recalls (6)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- CoolSculpting Elite SystemClass IIZ-2383-2021 · Zeltiq Aesthetics, Inc · posted Aug 27, 2021 · Terminated
An incorrect error messaging system that could potentially lead to: 1) Reporting a thermal event error causing a user to re-treating the affected anatomic area within 24 hours, 2) Not reporting a thermal event or any other error codes causing a user to…
- Z-2327-2021 · Boston Scientific Corporation · posted Aug 19, 2021 · Open, Classified
Potential for leaks and loosening at the patient catheter connection
- Z-2328-2021 · Boston Scientific Corporation · posted Aug 19, 2021 · Open, Classified
Potential for leaks and loosening at the patient catheter connection
- Z-2250-2021 · Exactech, Inc. · posted Aug 12, 2021 · Completed
Biolox Delta Femoral Head was packaged and labeled as 170-36-00 (36mm +0), but the device was marked as 170-36-03 (36mm +3.5).
- Exactech BIOLOX delta Femoral Head, 36m O.D. + 3.5mm, REF 170-36-03. Hip prosthesis component.Class IIZ-2251-2021 · Exactech, Inc. · posted Aug 12, 2021 · Completed
Biolox Delta Femoral Head was packaged and labeled as 170-36-00 (36mm +0), but the device was marked as 170-36-03 (36mm +3.5).
- Z-2277-2021 · Qura S.R.L. · posted Aug 12, 2021 · Terminated
Product labeled as sterile may not have been adequately sterilized. In addition, these devices were imported to the US distributor, but the devices are not 510(k) cleared in the US.
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.