Medical device recalls: January 2022
The FDA posted 23 medical device recall records in January 2022, from 11 recall events by 8 companies. 6 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Philips North America Llc · 3 events, 6 records
- Microbiologics Inc · 2 events, 2 records
- GE Healthcare, LLC · 1 event, 7 records
- Baxter Healthcare Corporation · 1 event, 2 records, 2 Class I
- ev3 Inc. · 1 event, 2 records, 2 Class I
- Philips Respironics, Inc. · 1 event, 2 records, 2 Class I
- Covidien Llc · 1 event, 1 record
- Siemens Medical Solutions USA, Inc · 1 event, 1 record
FDA root cause
- Device Design · 11
- No Marketing Application · 5
- Mixed-up of materials/components · 2
- Labeling mix-ups · 2
- Labeling design · 2
- Software design · 1
Class I recalls (6)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-0529-2022 · Baxter Healthcare Corporation · posted Jan 29, 2022 · Open, Classified
There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using…
- Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product code 35700BAX2.Class IZ-0530-2022 · Baxter Healthcare Corporation · posted Jan 29, 2022 · Open, Classified
There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using…
- Z-0493-2022 · Philips Respironics, Inc. · posted Jan 22, 2022 · Open, Classified
Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers. A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming material in the devices and repair…
- Z-0494-2022 · Philips Respironics, Inc. · posted Jan 22, 2022 · Open, Classified
Philips has identified a specific lot of non-conforming material manufactured by one of its suppliers. A Philips supplier incorrectly used polyester-based polyurethane (PE-PUR) sound abatement foam, a non-conforming material in the devices and repair…
- Z-0450-2022 · ev3 Inc. · posted Jan 19, 2022 · Terminated
Reports have been received of tip damage during use of the devices which resulted in some instances of tip detachment and embolization.
- Z-0451-2022 · ev3 Inc. · posted Jan 19, 2022 · Terminated
Reports have been received of tip damage during use of the devices which resulted in some instances of tip detachment and embolization.
Class II recalls (15)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-0518-2022 · GE Healthcare, LLC · posted Jan 24, 2022 · Terminated
The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
- Z-0519-2022 · GE Healthcare, LLC · posted Jan 24, 2022 · Terminated
The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
- Z-0520-2022 · GE Healthcare, LLC · posted Jan 24, 2022 · Terminated
The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
- Z-0521-2022 · GE Healthcare, LLC · posted Jan 24, 2022 · Terminated
The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
- Z-0522-2022 · GE Healthcare, LLC · posted Jan 24, 2022 · Terminated
The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
- Z-0523-2022 · GE Healthcare, LLC · posted Jan 24, 2022 · Terminated
The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
- Z-0524-2022 · GE Healthcare, LLC · posted Jan 24, 2022 · Terminated
The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.
- Z-0506-2022 · Philips North America Llc · posted Jan 19, 2022 · Open, Classified
After the device was cleared by the FDA, Philips made changes to the labeling that do not fall within the existing FDA clearance. These devices with the modified labeling cannot be distributed or sold without new 510(k) clearance.
- Z-0507-2022 · Philips North America Llc · posted Jan 19, 2022 · Open, Classified
After the device was cleared by the FDA, Philips made changes to the labeling that do not fall within the existing FDA clearance. These devices with the modified labeling cannot be distributed or sold without new 510(k) clearance.
- Z-0508-2022 · Philips North America Llc · posted Jan 19, 2022 · Open, Classified
After the device was cleared by the FDA, Philips made changes to the labeling that do not fall within the existing FDA clearance. These devices with the modified labeling cannot be distributed or sold without new 510(k) clearance.
- Z-0509-2022 · Philips North America Llc · posted Jan 19, 2022 · Open, Classified
After the device was cleared by the FDA, Philips made changes to the labeling that do not fall within the existing FDA clearance. These devices with the modified labeling cannot be distributed or sold without new 510(k) clearance.
- Z-0505-2022 · Philips North America Llc · posted Jan 18, 2022 · Open, Classified
After the device was cleared by the FDA, Philips made changes to the labeling that do not fall within the existing FDA clearance. These devices with the modified labeling cannot be distributed or sold without new 510(k) clearance.
- 1. Sensis, material # 10764561Class IIZ-0499-2022 · Siemens Medical Solutions USA, Inc · posted Jan 14, 2022 · Open, Classified
Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic circumstances, the CO (Cardiac…
- Z-0495-2022 · Covidien Llc · posted Jan 13, 2022 · Terminated
Customers reported that the device jaws were difficult to open or would not open following application on tissue. Use of a device with this potential assembly issue could result in the potential for bleeding, unintended tissue injury or loss, and delay in…
- Z-0476-2022 · Philips North America Llc · posted Jan 10, 2022 · Open, Classified
Wireless foot switch connection issues causing interruption of Fluoroscopy and exposure
Class III recalls (2)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-0544-2022 · Microbiologics Inc · posted Jan 31, 2022 · Terminated
Potential of products packaged for lot 323-108-7 were packaged in pouches labeled as 726-57-10.
- Z-0543-2022 · Microbiologics Inc · posted Jan 28, 2022 · Terminated
Potential that products packaged for lot 726-57-10 were packaged in pouches labeled as 323-108-7.