Allied Medical LLC AHP300Recalls & reported problems

Allied Medical, LLC · Transport Ventilator with Internal Compressor · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is the Allied Medical AHP300, a transport ventilator with a built-in compressor. The 29 reports mostly describe it delivering less air or less pressure than expected, with screens freezing or alarms going off unexpectedly, sometimes while a patient was on it. A few mention a leak in the internal compressor and patients getting too little air. No deaths and no injuries are reported, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 00026072000014), received between Apr 30, 2024 and Mar 12, 2025 — classified as 29 malfunctions.

The most frequently reported specific device problem is “Inaccurate Delivery” (23 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 25 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202625
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inaccurate Delivery23
  • Insufficient Device Problem Information6

Patient problems

  • No Clinical Signs, Symptoms or Conditions23
  • Hypoventilation5
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 12, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    THE UNIT FAILED WHILE ON A PATIENT. IT HAD A BATTERY LESS THAN A YEAR OLD IN IT AND I HAVE PUT A NEW BATTERY IN IT SENSE. IT SHOWS A FULL BATTERY UNTIL YOU TURN IT ON. THEN IT THROWS THE CRITICAL BATTERY ALARM AND WILL NOT FUNCTION. DEFINITELY NOT A BATTERY ISSUE.

    Patient problems: Insufficient Information

    MAUDE report 21590094 ↗
  • MalfunctionOct 1, 2024Inaccurate Delivery; Inaccurate Delivery

    DID NOT PASS 21% O2 PUMP ASSISTED TIDAL VOLUME AS RECEIVED. LOW TIDAL VOLUME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20348552 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    FAILED PRESSURE, PRESSURE ALARMS, APNEA, LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20274165 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    FROM THE PARAMEDIC WHO WITNESSED THE MALFUNCTION: THE VENTILATOR HAD STRANGE ISSUE OF THE PIP SUDDENLY SHOWING 6 WOTHOUT CHANGING FROM THAT NUMBER A WORSENING IN PATIENT CONDITION. I HAD TO TAP THE SCREEN WHICH MADE IT START TO WORK AGAIN. IT HAPPENED AGAIN ON THE TRANSPORT AND AGAIN TAPPING IT WORKED. THE VENT WAS PULLED OFFLINE.

    Patient problems: Hypoventilation

    MAUDE report 20274042 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING AS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20274034 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    FAILED CPAP AND APNEA, LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20274024 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING AS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20274015 ↗
  • MalfunctionSep 20, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    LEAK IN INTERNAL COMPRESSOR CAUSING BIPAP AND CPAP NOT TO WORK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273950 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    FAILED PC, CPAP, APNEA, LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273891 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING AS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273879 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING AS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273875 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    LOW TIDAL VOLUMES DURING MANUFACTURER MAINTENANCE TESTING AS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273873 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    FAIL CPAP, LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273871 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING AS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273821 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    VERY LOW TIDAL VOLUMES AND NO AUDIBLE ALARM DURING MANUFACTURER MAINTENANCE TESTING

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273819 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING AS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273809 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    FAILED PC, CPAP, LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273784 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    FAILED CPAP TEST AND LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273767 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    FAILED O2 22%, LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273751 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    LOW TIDAL VOLUME DURING MANUFACTURER MAINTENANCE TESTING AS RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273729 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    LOW TIDAL VOLUME.

    Patient problems: Hypoventilation

    MAUDE report 20273621 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    UNIT IS NOT DELIVERING VOLUMES OR PRESSURE AS INTENDED

    Patient problems: Hypoventilation

    MAUDE report 20273610 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    LOW TIDAL VOLUME.

    Patient problems: Hypoventilation

    MAUDE report 20273601 ↗
  • MalfunctionSep 20, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    PRESSURE CONTROL MODE PEAK INSPIRATORY PRESSURE LIMITS ARE OUT OF TOLERANCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273567 ↗
  • MalfunctionSep 20, 2024Inaccurate Delivery; Inaccurate Delivery

    LOW TIDAL VOLUME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20273556 ↗

Other Ventilator, Emergency, Powered (Resuscitator) devices