Neo-Tee T-Piece Resuscitator 10-50811Recalls & reported problems
Mercury Enterprises, Inc. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10641043508114), received between Dec 5, 2017 and Jan 7, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device Operates Differently Than Expected” (1 report).
2 FDA recall records name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K093913); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Dec 10, 2024Class IOpen, Classified
Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.
Z-0821-2025 · Root cause: Process change control · Recall details · FDA record ↗
- Oct 22, 2015Class IITerminated
Incorrect use of T-Piece resuscitators will result in complete occlusion of the patient circuit and consequent failure to ventilate the patient.
Z-0639-2016 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Operates Differently Than Expected1
- Device Slipped1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 7, 2026
WOULD NEVER ALLOW YOU TO SET A PEEP [POSITIVE END-EXPIRATORY PRESSURE]. TRIED SEVERAL DIFFERENT RADIANT WARMERS TO MAKE SURE, FLOW WAS SET TO 10LPM, ABLE TO GET A PIP [PEAK INSPIRATORY PRESSURE] LEVEL SET, BUT WHEN YOU TURN THE DIAL FOR PEEP IT STAYS AT 0 AND WILL NOT ALLOW FOR PEEP SO UNABLE TO GIVE CPAP [CONTINUOUS POSITIVE AIRWAY PRESSURE] TO NEWBORN INFANT. I TRIED SEVERAL WITH THIS LOT NUMBER AND THREE OF THEM WOULD NOT SET THE PEEP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23988450 ↗MalfunctionApr 9, 2020Device Slipped
THE MANOMETER FALLS OUT OF THE PATIENT CIRCUIT, THERE IS NOTHING HOLDING IT IN PLACE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9943310 ↗MalfunctionDec 5, 2017Device Operates Differently Than Expected
UNABLE TO SET PIP OR PEEP ON DEVICE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7082183 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K093913. One submission often covers many models, so check the product and lot details in each record.
- Z-0004-2025 · Aug 30, 2024 · Open, ClassifiedNeo-Tee T-Piece Resuscitator. Part Numbers 1050805, 1050832. For pediatric emergency respiratory support.
Other Ventilator, Emergency, Powered (Resuscitator) devices
- Pneupac P310NUSIcu Medical, Inc. · Ventilator, emergency, powered (resuscitator)4 deaths1,035 reports
- Pneupac 120003Icu Medical, Inc. · Ventilator, emergency, powered (resuscitator)2 deaths523 reports
- Neo-Tee T-Piece Resuscitator 10-50805Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)3 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50832Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50810Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50809Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls0 reports
- Neo-Tee T-Piece Resuscitator 10-50808Mercury Enterprises, Inc. · Ventilator, Emergency, Powered (Resuscitator)2 recalls0 reports
- Pneupac 130003Icu Medical, Inc. · Ventilator, emergency, powered (resuscitator)1 death77 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page