Skip to content

Neo-Tee T-Piece Resuscitator 10-50808Recalls & reported problems

Mercury Enterprises, Inc. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K093913); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 10, 2024Class IOpen, Classified

    Gas powered emergency resuscitator has a circuit flow controller manufactured with an undersized spring, which may cause resuscitator not reach the high and low ends of the PIP and PEEP pressure ranges, which may cause a loss of positive pressure, impacting effective ventilation of the patient.

    Z-0821-2025 · Root cause: Process change control · Recall details · FDA record ↗

  2. Oct 22, 2015Class IITerminated

    Incorrect use of T-Piece resuscitators will result in complete occlusion of the patient circuit and consequent failure to ventilate the patient.

    Z-0639-2016 · Root cause: Device Design · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K093913. One submission often covers many models, so check the product and lot details in each record.

  • Z-0004-2025 · Aug 30, 2024 · Open, Classified
    Neo-Tee T-Piece Resuscitator. Part Numbers 1050805, 1050832. For pediatric emergency respiratory support.

Other Ventilator, Emergency, Powered (Resuscitator) devices

All Ventilator, Emergency, Powered (Resuscitator) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page