i-STAT 03P75-01Recalls & reported problems
Abbott Point Of Care Inc. · i-Stat 1 Wireless Analyzer · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00054749001408), received between Aug 25, 2017 and Oct 28, 2020 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Electrical /Electronic Property Problem5
Patient problems
- No Patient Involvement4
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 28, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
MalfunctionAug 10, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
ON 06/04/2020, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT I-STAT ANALYZER SN (B)(4) WAS TOO HOT TO TOUCH AND THERE WAS DAMAGE TO THE CASE. THERE WAS NO ADDITIONAL INFORMATION AT THE TIME OF THIS REPORT. THE CUSTOMER STATED THAT THE ANALYZER WAS DOCKED ON I-STAT DOWNLOADER RECHARGER DRC-79334 AND THEY FOUND THE ANALYZER HOT TO TOUCH WHILE IN THE DOCKING STATION. APOC HAS REPLACED THE ANALYZER, DOWNLOADER RECHARGER DRC-79334, ALOLNG WITH THE RECHARGEABLE BATTERY AT NO CHARGE TO BE RETURNED FOR INVESTIGATION AS A COMPLETE UNIT. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. HOWEVER, THE CUSTOMER STATES THAT RECHARGEABLE BATTERIES WERE BEING USED AT THE TIME OF THE EVENT. THEREFORE THE ANALYZER IS UNLIKELY TO BECOME HOT TO TOUCH. THE PRODUCT WAS REPLACED AND RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.
Patient problems: No Patient Involvement
MAUDE report 10386456 ↗MalfunctionMay 20, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
MalfunctionAug 25, 2017Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
ON (B)(6)2017, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT ANALYZER SN (B)(4) WAS FEELING "EXTREMELY HOT" AND NOT ACTIVATING . THE CUSTOMER ALSO STATED THAT THERE WAS NO INDICATION OF SMOKE, OR BURNING SMELL. THERE WAS NO IMPACT TO PATIENT CARE OR MANAGEMENT AS PATIENTS WERE TESTED ON OTHER I-STAT ANALYZERS. THE I-STAT WAS OPERATING WITH RECHARGEABLE BATTERIES. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. HOWEVER, THE CUSTOMER STATES THAT RECHARGEABLE BATTERIES WERE BEING USED. THEREFORE THE ANALYZER IS UNLIKELY TO BECOME HOT TO TOUCH. THE PRODUCT WAS REPLACED AND RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.
Patient problems: No Patient Involvement
MAUDE report 6819961 ↗MalfunctionAug 25, 2017Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
ON (B)(6) 2017, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT ANALYZER (B)(4) WAS FEELING "REALLY HOT" AND NOT ACTIVATING. THE CUSTOMER ALSO STATED THAT THERE WAS NO INDICATION OF SMOKE, OR BURNING SMELL. THERE WAS NO IMPACT TO PATIENT CARE OR MANAGEMENT AS PATIENTS WERE TESTED ON OTHER I-STAT ANALYZERS. THE I-STAT WAS OPERATING WITH RECHARGEABLE BATTERIES. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. HOWEVER, THE CUSTOMER STATES THAT RECHARGEABLE BATTERIES WERE BEING USED. THEREFORE THE ANALYZER IS UNLIKELY TO BECOME HOT TO TOUCH. THE PRODUCT WAS REPLACED AND RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.
Patient problems: No Patient Involvement
MAUDE report 6819958 ↗
Other GLUCOSE OXIDASE, GLUCOSE devices
- TRUEtrack TRUEtrack 50 Blood Glucose Test StripsTrividia Health, Inc. · Glucose Oxidase, Glucose153 reports
- HemoCue® Glucose 201 Microcuvettes 110706Hemocue AB · Glucose Oxidase, Glucose2 recalls0 reports
- VITROS 1707801Ortho-Clinical Diagnostics, Inc. · GLUCOSE OXIDASE, GLUCOSE1 recall48 reports
- Prodigy AutoCode Blood Glucose Monitoring System 051885Prodigy Diabetes Care, LLC · Glucose Oxidase, Glucose93 reports
- OneTouch Ultra Blue 022-895LifeScan U.S. LLC · GLUCOSE OXIDASE, GLUCOSE1 recall7 reports
- OneTouch Ultra Blue 020-994LifeScan U.S. LLC · GLUCOSE OXIDASE, GLUCOSE1 recall2 reports
- Alere Cholestech LDX® 10-991Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall1 report
- Alere Cholestech LDX® 14-531Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page