Skip to content

i-STAT 03P75-06Recalls & reported problems

Abbott Point Of Care Inc. · i-Stat 1 Wireless Analyzer · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00054749002528), received between Dec 12, 2018 and Jan 17, 2023 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem6

Patient problems

  • No Patient Involvement4
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 17, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    NA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16183642 ↗
  • MalfunctionJun 29, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    ON (B)(6) 2022, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT I-STAT1 ANALYZER SN (B)(4) BEGAN BURNING AND SMOKING WHEN PLACED IN THE DOWNLOADER. THE CUSTOMER REPORTS HEARING A POP AND A SIZZLE AND THEN THE CUSTOMER REPORTS SMELLING BURNING ELECTRICAL COMPONENTS. THERE WERE NO INJURIES ASSOCIATED WITH THE EVENT. THE ANALYZER AND DOWNLOADER WERE REPLACED AT NO CHARGE AND RETURNING FOR INVESTIGATION. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. HOWEVER, THE CUSTOMER STATES THAT A RECHARGEABLE BATTERY WAS BEING USED IN THIS ANALYZER AT THE TIME OF THE EVENT. THEREFORE THE ANALYZER IS UNLIKELY TO BECOME HOT TO TOUCH. THE PRODUCT WAS REPLACED AND IS BEING RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14851161 ↗
  • MalfunctionJun 16, 2020Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    ON (B)(6) 2020, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT I-STAT ANALYZER SN (B)(4) WAS TOO HOT TO TOUCH AFTER BATTERY WAS INSTALLED & ALSO SMELL SMOKE WAS DETECTED; HOWEVER, NO FLAME WAS SEEN. THERE WAS NO ADDITIONAL INFORMATION AT THE TIME OF THIS REPORT. THE ANALYZER WAS REPLACED AT NO CHARGE AND RETURNING FOR INVESTIGATION. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. HOWEVER, THE CUSTOMER STATES THAT RECHARGEABLE BATTERIES WERE BEING USED AT THE TIME OF THE EVENT. THEREFORE, THE ANALYZER IS UNLIKELY TO BECOME HOT TO TOUCH. THE PRODUCT WAS REPLACED AND RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.

    Patient problems: No Patient Involvement

    MAUDE report 10159902 ↗
  • MalfunctionNov 19, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    ON (B)(6) 2019, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED VISIBLE SMOKE COMING FROM I-STAT1 ANALYZER (B)(4) DURING A SOFTWARE UPDATE EVENT. THERE WAS NO ADDITIONAL INFORMATION AT THE TIME OF THIS REPORT. THE ANALYZER WAS REPLACED AT NO CHARGE AND RETURNING FOR INVESTIGATION. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. HOWEVER, THE CUSTOMER STATES THAT RECHARGEABLE BATTERIES WERE BEING USED AT THE TIME OF THE EVENT. THEREFORE THE ANALYZER IS UNLIKELY TO BECOME HOT TO TOUCH. THE PRODUCT WAS REPLACED AND RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.

    Patient problems: No Patient Involvement

    MAUDE report 9342096 ↗
  • MalfunctionMay 16, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    ON (B)(6) 2019, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED VISIBLE SMOKE COMING FROM I-STAT1 ANALYZER SN (B)(4) DURING A POWER UP EVENT THERE WAS NO ADDITIONAL INFORMATION AT THE TIME OF THIS REPORT. THE ANALYZER WAS REPLACED AT NO CHARGE AND RETURNING FOR INVESTIGATION. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. HOWEVER, THE CUSTOMER STATES THAT RECHARGEABLE BATTERIES WERE BEING USED AT THE TIME OF THE EVENT. THEREFORE THE ANALYZER IS UNLIKELY TO BECOME HOT TO TOUCH. THE PRODUCT WAS REPLACED AND RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.

    Patient problems: No Patient Involvement

    MAUDE report 8615338 ↗
  • MalfunctionDec 12, 2018Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    ON (B)(6) 2018, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT I-STAT1 SN (B)(4) IS DISPLAYING CODE 86 AND IS HOT TO THE TOUCH. THE CUSTOMER REPORTS THAT THERE WAS NO IMPACT OR DELAY TO PATIENT CARE. THERE WERE NO INJURIES REPORTED. THE ANALYZER WAS REPLACED AT NO CHARGE AND RETURNING FOR INVESTIGATION. ON (B)(6) 2018 THE CUSTOMER REPORTED THAT DOWNLOADER RECHARGER SN (B)(4) ALSO BECAME HOT TO THE TOUCH THAT WAS BEING USED DURING THE EVENT. THERE WERE NO INJURIES REPORTED ON THE SECOND CALL. (B)(4) WAS REPLACED AT NO CHARGE AND ALSO RETURNING FOR INVESTIGATION. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. HOWEVER, THE CUSTOMER STATES THAT RECHARGEABLE BATTERIES WERE BEING USED AT THE TIME OF THE EVENT. THEREFORE THE ANALYZER IS UNLIKELY TO BECOME HOT TO TOUCH. THE PRODUCT WAS REPLACED AND RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.

    Patient problems: No Patient Involvement

    MAUDE report 8156728 ↗

Other GLUCOSE OXIDASE, GLUCOSE devices

All GLUCOSE OXIDASE, GLUCOSE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.