Skip to content

i-STAT 04P75-01RRecalls & reported problems

Abbott Point Of Care Inc. · i-STAT 1 Analyzer, Immuno Ready · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00054749003600), received between Feb 9, 2018 and Dec 19, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem1
  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 19, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    ON (B)(6) 2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED ELECTRONIC BURNING, POSSIBLY BURNT LITHIUM SMELL ON THEIR I-STAT 1 ANALYZER SN (B)(6). CUSTOMER STATES THE ANALYZER GOT VERY HOT AND HAD BURNT GOLDEN CONTACT PADS ON BACK SIDE OF IT WHERE IT WOULD DOCK ONTO THE DOWNLOADER. CUSTOMER STATES THAT ENERGIZER LITHIUM BATTERIES WERE USED. CUSTOMER THREW OUT THE BOX OF ENERGIZER LITHIUM BATTERIES. A NEW LOT WAS ACQUIRED AND RUN WITH NO ISSUES ON OTHER ANALYZERS. CUSTOMER STATES THEY ARE USING RED FUSED BATTERY CARRIERS AND ASKED IF LEAVING THEM DOCKED ON THE DOWNLOADER OVERNIGHT MAY HAVE CAUSED THE OVER-HEATING. APOC TECHNICAL SUPPORT ADVISED THAT RECHARGEABLE BATTERIES ARE RECOMMENDED FOR THE DOWNLOADER RECHARGER. ANALYZER WILL BE REPLACED AT NO CHARGE AND RETURNED FOR INVESTIGATION. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. THERE ARE NO INJURIES ASSOCIATED WITH THE EVENTS. THE PRODUCT WAS REPLACED AND RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE N

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18354821 ↗
  • MalfunctionFeb 9, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    ON (B)(6) 2018, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT ANALYZER SN (B)(4) THAT YIELDED SUSPECTED DISCREPANT GLUCOSE RESULTS ON MULTIPLE PATIENTS. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. CUSTOMER REPORTS THEY RAN CARTRIDGE LOTS H17265B AND H17234 ON A DIFFERENT ANALYZER AND THERE HAVE BEEN NO ISSUES. I-STAT(313271): GLUCOMETER: CHEMISTRY ANALYZER: RESULT: 71, 42, 50. RESULT: 50, 42, 41. RESULT: 79, 52, 34. RESULT: 61, 48, 51. RESULT: 76, 48, 54. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

    Patient problems: No Information

    MAUDE report 7260911 ↗

Other GLUCOSE OXIDASE, GLUCOSE devices

All GLUCOSE OXIDASE, GLUCOSE devices →