i-STAT 04P75-01RRecalls & reported problems
Abbott Point Of Care Inc. · i-STAT 1 Analyzer, Immuno Ready · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00054749003600), received between Feb 9, 2018 and Dec 19, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Information1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 19, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
ON (B)(6) 2023, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED ELECTRONIC BURNING, POSSIBLY BURNT LITHIUM SMELL ON THEIR I-STAT 1 ANALYZER SN (B)(6). CUSTOMER STATES THE ANALYZER GOT VERY HOT AND HAD BURNT GOLDEN CONTACT PADS ON BACK SIDE OF IT WHERE IT WOULD DOCK ONTO THE DOWNLOADER. CUSTOMER STATES THAT ENERGIZER LITHIUM BATTERIES WERE USED. CUSTOMER THREW OUT THE BOX OF ENERGIZER LITHIUM BATTERIES. A NEW LOT WAS ACQUIRED AND RUN WITH NO ISSUES ON OTHER ANALYZERS. CUSTOMER STATES THEY ARE USING RED FUSED BATTERY CARRIERS AND ASKED IF LEAVING THEM DOCKED ON THE DOWNLOADER OVERNIGHT MAY HAVE CAUSED THE OVER-HEATING. APOC TECHNICAL SUPPORT ADVISED THAT RECHARGEABLE BATTERIES ARE RECOMMENDED FOR THE DOWNLOADER RECHARGER. ANALYZER WILL BE REPLACED AT NO CHARGE AND RETURNED FOR INVESTIGATION. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. THERE ARE NO INJURIES ASSOCIATED WITH THE EVENTS. THE PRODUCT WAS REPLACED AND RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE N
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18354821 ↗MalfunctionFeb 9, 2018Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2018, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT ANALYZER SN (B)(4) THAT YIELDED SUSPECTED DISCREPANT GLUCOSE RESULTS ON MULTIPLE PATIENTS. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. CUSTOMER REPORTS THEY RAN CARTRIDGE LOTS H17265B AND H17234 ON A DIFFERENT ANALYZER AND THERE HAVE BEEN NO ISSUES. I-STAT(313271): GLUCOMETER: CHEMISTRY ANALYZER: RESULT: 71, 42, 50. RESULT: 50, 42, 41. RESULT: 79, 52, 34. RESULT: 61, 48, 51. RESULT: 76, 48, 54. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Patient problems: No Information
MAUDE report 7260911 ↗
FDA clearance and classification
The FDA regulates product code CGA (glucose oxidase, glucose) under 21 CFR 862.1345 as a Class II device. 21 CFR 862.1345 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the i-STAT 04P75-01R
Has the i-STAT 04P75-01R been recalled?
No FDA recall record in our data names this exact device identifier (DI 00054749003600), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the i-STAT 04P75-01R?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Feb 9, 2018 and Dec 19, 2023. In those reports, the most common specific device problems are “Electrical /Electronic Property Problem” (1) and “Incorrect, Inadequate or Imprecise Result or Readings” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the i-STAT 04P75-01R?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the i-STAT 04P75-01R, and is it still sold?
The FDA's device database lists Abbott Point Of Care Inc. as the company responsible for it (brand i-STAT, model 04P75-01R). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other GLUCOSE OXIDASE, GLUCOSE devices
- TRUEtrack TRUEtrack 50 Blood Glucose Test StripsTrividia Health, Inc. · Glucose Oxidase, Glucose153 reports
- HemoCue® Glucose 201 Microcuvettes 110706Hemocue AB · Glucose Oxidase, Glucose2 recalls0 reports
- VITROS 1707801Ortho-Clinical Diagnostics, Inc. · GLUCOSE OXIDASE, GLUCOSE1 recall48 reports
- Prodigy AutoCode Blood Glucose Monitoring System 051885Prodigy Diabetes Care, LLC · Glucose Oxidase, Glucose93 reports
- OneTouch Ultra Blue 022-895LifeScan U.S. LLC · GLUCOSE OXIDASE, GLUCOSE1 recall7 reports
- OneTouch Ultra Blue 020-994LifeScan U.S. LLC · GLUCOSE OXIDASE, GLUCOSE1 recall2 reports
- Alere Cholestech LDX® 10-991Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall1 report
- Alere Cholestech LDX® 99021Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page