i-STAT 04P75-36Recalls & reported problems
Abbott Point Of Care Inc. · i-STAT 1 Analyzer, Immuno Ready · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00054749003754), received between Feb 6, 2019 and Feb 6, 2019 — classified as 1 malfunction.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).
13 other recalls cover devices cleared under the same FDA submission (K001387); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Electrical /Electronic Property Problem1
Patient problems
- No Patient Involvement1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 6, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
ON (B)(6) 2019, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED VISIBLE SMOKE COMING FROM I-STAT1 ANALYZER SN (B)(4) AND WOULD NOT POWER UP AFTER THE EVENT THERE WAS NO ADDITIONAL INFORMATION AT THE TIME OF THIS REPORT. THE ANALYZER WAS REPLACED AT NO CHARGE AND RETURNING FOR INVESTIGATION. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. HOWEVER, THE CUSTOMER STATES THAT RECHARGEABLE BATTERIES WERE BEING USED AT THE TIME OF THE EVENT. THEREFORE THE ANALYZER IS UNLIKELY TO BECOME HOT TO TOUCH. THE PRODUCT WAS REPLACED AND RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.
Patient problems: No Patient Involvement
MAUDE report 8312463 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K001387. One submission often covers many models, so check the product and lot details in each record.
- Z-1973-2012 · Sep 8, 2011 · TerminatediStat1 Analyzer Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2, and the i-Stat 200 anal
- Z-1978-2012 · Sep 8, 2011 · TerminatedMilitary Kit Portable Clinical Analyzer ; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, P
- Z-1975-2012 · Sep 8, 2011 · TerminatedFuso Analyzer 300F; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2, and the i-Stat 200
- Z-1977-2012 · Sep 8, 2011 · Terminatedi-STAT1 Abaxis Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2, and the i-Stat
- Z-1976-2012 · Sep 8, 2011 · Terminatedi-STAT1 Abaxis Analyzer; Abbott Point of Care, Princeton, NJ 08540
- Z-1974-2012 · Sep 8, 2011 · Terminatedi-Stat Portable Clinical Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2, and
- Z-0597-2011 · Sep 20, 2010 · Terminatedi-STAT Portable Clinical Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2
- Z-0598-2011 · Sep 20, 2010 · TerminatedMilitary Kit Portable Clinical Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, p
- Z-2462-2010 · Aug 20, 2010 · TerminatedGCX Mountable Downloader-Recharger; (accessory to the i-STAT Portable Clinical Analyzer (model 300). The GCX Mountable Downloader-Recharger (GCX) is a connectivity component that i
- Z-1133-2010 · Nov 6, 2009 · TerminatedMartel Printer The Martel Printer is an accessory to the i-STAT Portable clinical Analyzer (model 300). The i-STAT Analyzer is used by trained medical professionals for running a
- Z-0865-2010 · Sep 29, 2009 · Terminatedi-STAT HL7 Interface The HL7 Interface product is an accessory to the analyzer
- Z-0812-2010 · Sep 9, 2009 · Terminatedi-STAT 1 Analyzer, Abbott Point of Care Inc., Abbott Park, IL 60064 USA, Product of USA. The product is intended to be used by trained medical professionals for use with i-STAT t
- Z-2328-2009 · Apr 22, 2009 · TerminatedMartel Printer Accessory to i-STAT Portable Clinical Analyzer, rechargeable battery pack. Printing results of clinical chemistry tests and test panels contained in the i-STAT tes
Other GLUCOSE OXIDASE, GLUCOSE devices
- TRUEtrack TRUEtrack 50 Blood Glucose Test StripsTrividia Health, Inc. · Glucose Oxidase, Glucose153 reports
- HemoCue® Glucose 201 Microcuvettes 110706Hemocue AB · Glucose Oxidase, Glucose2 recalls0 reports
- VITROS 1707801Ortho-Clinical Diagnostics, Inc. · GLUCOSE OXIDASE, GLUCOSE1 recall48 reports
- Prodigy AutoCode Blood Glucose Monitoring System 051885Prodigy Diabetes Care, LLC · Glucose Oxidase, Glucose93 reports
- OneTouch Ultra Blue 022-895LifeScan U.S. LLC · GLUCOSE OXIDASE, GLUCOSE1 recall7 reports
- OneTouch Ultra Blue 020-994LifeScan U.S. LLC · GLUCOSE OXIDASE, GLUCOSE1 recall2 reports
- Alere Cholestech LDX® 10-991Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall1 report
- Alere Cholestech LDX® 14-531Alere San Diego, Inc. · Glucose Oxidase, Glucose1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.