Skip to content

i-STAT 04P75-38Recalls & reported problems

Abbott Point Of Care Inc. · i-STAT 1 Analyzer, Immuno Ready · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00054749003778), received between May 27, 2022 and May 27, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (1 report).

13 other recalls cover devices cleared under the same FDA submission (K001387); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 27, 2022Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    ON (B)(6) 2022, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT I-STAT1 ANALYZER SN (B)(4) HAD A NOTE ON IT WHICH SAID "PLEASE LEAVE THIS IN DOCKING STATION TO CHARGE.". CUSTOMER STATES THAT IT WAS NOT CHARGING AND THEY WENT TO PICK IT UP THEY REALIZED IT WAS HOT TO THE TOUCH AND HAD A BURNING SMELL. THE ANALYZER AND RECHARGEABLE BATTERY WERE REPLACED AT NO CHARGE AND RETURNING FOR INVESTIGATION. APOC HAS DETERMINED THAT A COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT WHEN USING A GREEN NON-FUSED BATTERY CARRIER. HOWEVER, THE CUSTOMER STATES THAT DISPOSABLE LITHIUM BATTERIES IN A RED CARRIER WERE BEING USED AT THE TIME OF THE EVENT. THEREFORE THE ANALYZER IS UNLIKELY TO BECOME HOT TO TOUCH. THE PRODUCT WAS REPLACED AND IS BEING RETURNED FOR INVESTIGATION. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14521684 ↗

FDA clearance and classification

Cleared through 510(k) K001387 on Oct 5, 2000 (decision: Substantially Equivalent). Submitted by I-Stat Corp. as “I-Stat Portable Clinical Analyzer, Model 300”. The FDA review took 156 days. FDA record ↗

The FDA regulates product code CGA (glucose oxidase, glucose) under 21 CFR 862.1345 as a Class II device. 21 CFR 862.1345 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K001387. One submission often covers many models, so check the product and lot details in each record.

  • Z-1974-2012 · Sep 8, 2011 · Terminated
    i-Stat Portable Clinical Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2, an
  • Z-1977-2012 · Sep 8, 2011 · Terminated
    i-STAT1 Abaxis Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2, and the i-St
  • Z-1976-2012 · Sep 8, 2011 · Terminated
    i-STAT1 Abaxis Analyzer; Abbott Point of Care, Princeton, NJ 08540
  • Z-1975-2012 · Sep 8, 2011 · Terminated
    Fuso Analyzer 300F; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2, and the i-Stat 20
  • Z-1973-2012 · Sep 8, 2011 · Terminated
    iStat1 Analyzer Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2, PO2, and the i-Stat 200 a
  • Z-1978-2012 · Sep 8, 2011 · Terminated
    Military Kit Portable Clinical Analyzer ; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2,
  • Z-0598-2011 · Sep 20, 2010 · Terminated
    Military Kit Portable Clinical Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, p
  • Z-0597-2011 · Sep 20, 2010 · Terminated
    i-STAT Portable Clinical Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree 37 degree C tests for sodium, potassium, ionized calcium, hematocrit, pH, PCO2
  • Z-2462-2010 · Aug 20, 2010 · Terminated
    GCX Mountable Downloader-Recharger; (accessory to the i-STAT Portable Clinical Analyzer (model 300). The GCX Mountable Downloader-Recharger (GCX) is a connectivity component that i
  • Z-1133-2010 · Nov 6, 2009 · Terminated
    Martel Printer The Martel Printer is an accessory to the i-STAT Portable clinical Analyzer (model 300). The i-STAT Analyzer is used by trained medical professionals for running
  • Z-0865-2010 · Sep 29, 2009 · Terminated
    i-STAT HL7 Interface The HL7 Interface product is an accessory to the analyzer
  • Z-0812-2010 · Sep 9, 2009 · Terminated
    i-STAT 1 Analyzer, Abbott Point of Care Inc., Abbott Park, IL 60064 USA, Product of USA. The product is intended to be used by trained medical professionals for use with i-STAT
  • Z-2328-2009 · Apr 22, 2009 · Terminated
    Martel Printer Accessory to i-STAT Portable Clinical Analyzer, rechargeable battery pack. Printing results of clinical chemistry tests and test panels contained in the i-STAT t

Common questions about the i-STAT 04P75-38

Has the i-STAT 04P75-38 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00054749003778), based on the FDA recall file dated Nov 27, 2026. However, 13 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the i-STAT 04P75-38?

The FDA's MAUDE database lists 1 adverse event report for this device, received on May 27, 2022. That report lists device problem “Electrical /Electronic Property Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the i-STAT 04P75-38?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the i-STAT 04P75-38 FDA approved?

It was cleared through 510(k) K001387 on Oct 5, 2000. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the i-STAT 04P75-38, and is it still sold?

The FDA's device database lists Abbott Point Of Care Inc. as the company responsible for it (brand i-STAT, model 04P75-38). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other GLUCOSE OXIDASE, GLUCOSE devices

All GLUCOSE OXIDASE, GLUCOSE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page