Skip to content

Invictus 17109-055Recalls & reported problems

Alphatec Spine, Inc. · 5.5mm Tap, Cannulated · Class 1

All models of Invictus (96)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
22
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 00190376192079), received between May 9, 2022 and Aug 26, 2026 — classified as 5 injuries, 17 malfunctions.

The most frequently reported specific device problem is “Break” (17 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break17
  • Material Separation6
  • Fracture4
  • Damaged Thread1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • Foreign Body In Patient5
  • Insufficient Information2

Recent adverse event reports

The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2026Break; Material Separation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50456923 ↗
  • MalfunctionMar 10, 2026Break; Material Separation

    IT WAS REPORTED THAT THE TAP HAS A BROKEN TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24561094 ↗
  • MalfunctionFeb 12, 2026Break

    IT WAS REPORTED THAT THE TAP WAS DAMAGED AFTER GOING THROUGH THE WASH IN STERILE PROCESSING AT THE HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24326549 ↗
  • MalfunctionDec 30, 2025Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TAP HAS A BROKEN TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23930907 ↗
  • MalfunctionJul 18, 2025Break; Material Separation; Break; Material Separation

    DURING THE PROCEDURE, A CANNULATED TAP WAS INSERTED OVER THE GUIDE WIRE. WHILE ATTEMPTING TO MAINTAIN THE INTENDED TRAJECTORY, THE TAP FRACTURED, RESULTING IN THE THREADED PORTION BREAKING OFF AND REMAINING WITHIN THE PEDICLE. THE THREADED FRAGMENT WAS SUBSEQUENTLY REMOVED FROM THE PEDICLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22542612 ↗
  • MalfunctionJun 26, 2025Break; Break

    IT WAS REPORTED THAT THE TAP IS BROKEN.

    Patient problems: Insufficient Information

    MAUDE report 22341220 ↗
  • MalfunctionJun 26, 2025Break; Break

    IT WAS REPORTED THAT THE TAP IS BROKEN.

    Patient problems: Insufficient Information

    MAUDE report 22341149 ↗
  • InjuryNov 21, 2024Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT THE TIP OF THE TAP BROKE OFF INTO THE PATIENT'S BONE AND REMAINS THERE. NO FURTHER COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 20750470 ↗
  • InjuryOct 15, 2024Break; Material Separation; Break; Material Separation

    DURING A SPINAL PROCEDURE, IT WAS REPORTED THAT THE THREADED PORTION OF THE CANNULATED TAP BROKE OFF INSIDE THE PATIENT'S VERTEBRAL BODY. IT WAS UNABLE TO BE RETRIEVED FROM THE PATIENT. THERE WAS NO SURGICAL DELAY.

    Patient problems: Foreign Body In Patient

    MAUDE report 20452536 ↗
  • MalfunctionOct 8, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF THE TAP IS CRACKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20402499 ↗
  • MalfunctionJul 17, 2024Break; Break

    IT WAS REPORTED THAT THE DISTAL TIP OF THE TAP SHEARED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19766155 ↗
  • InjuryJul 11, 2024Break; Break

    IT WAS REPORTED THAT WHILE TAPPING A HOLE IN THE PATIENT'S BONE, THE TAP SHEARED OFF. THE SHEARED TIP REMAINS IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 19724520 ↗
  • MalfunctionJun 27, 2024Break; Break

    IT WAS REPORTED THAT THE TAP WAS RECEIVED WITH THE TIP BROKEN OFF. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19627532 ↗
  • InjuryJun 21, 2024Break; Break

    IT WAS REPORTED THAT THE TIP OF THE TAP BROKE OFF INTO THE PATIENT'S BONE. THE TIP REMAINS IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 19589834 ↗
  • MalfunctionJun 14, 2024Damaged Thread; Damaged Thread

    IT WAS REPORTED THAT THE DISTAL END OF THE TAP SPLAYED DURING A CASE. THERE WAS NO REPORT OF PATIENT INJURY DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19539133 ↗
  • MalfunctionApr 10, 2024Break; Break

    IT WAS REPORTED THAT A TAP BROKE DURING A SPINAL SURGERY PROCEDURE. ALL PIECES WERE RETRIEVED FROM THE PATIENT. THERE WAS NO REPORT OF PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19080147 ↗
  • MalfunctionMar 27, 2024Break; Break

    IT WAS REPORTED THAT THE TAP BROKE DURING A CASE. ALL PIECES WERE RETRIEVED. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED AS RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18986937 ↗
  • InjuryOct 16, 2023Fracture; Fracture

    WHILE PREPARING THE SCREW HOLE, THE DISTAL TIP OF THE TAP BROKE OFF IN THE PEDICLE AND COULD NOT BE RETRIEVED. IT WAS REPORTED THE PATIENT HAD INCREDIBLY DENSE, SCLEROTIC BONE. THIS IS REPORT 1 OF 2.

    Patient problems: Foreign Body In Patient

    MAUDE report 17945658 ↗
  • MalfunctionMay 5, 2023Fracture; Fracture

    IT WAS REPORTED THE TIP OF A TAP BENT DURING SURGERY. THERE WERE NO REPORTED PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16882317 ↗
  • MalfunctionOct 26, 2022Fracture; Fracture

    IT WAS REPORTED THE TIP OF A 5.5MM TAP BROKE INTO 3 PIECES DURING SURGERY. ALL OF THE PIECES WERE RETRIEVED FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15677586 ↗
  • MalfunctionSep 24, 2022Break; Break

    THE DISTAL TIP IS BROKEN OFF. NO PATIENT INVOLVEMENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15477294 ↗
  • MalfunctionMay 9, 2022Break; Break

    THE TIP OF A TAP SHEARED OFF IN THE PATIENT'S PEDICLE. IT WAS RETRIEVED FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14342256 ↗

Other TAP, BONE devices

All TAP, BONE devices →