Skip to content

NA 311.01Recalls & reported problems

Synthes GmbH · HANDLE WITH MINI QUICK COUPLING · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Synthes bone tap handle with a mini quick coupling, a surgical tool, has 106 reports, no deaths and 1 injury. Many reports describe the quick connection getting stuck or failing to hold screwdriver shafts, and some describe the handle breaking or chipping. Most reports note no clinical signs or patient consequences. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
106
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 106 adverse event reports that list this exact device (DI 10886982187192), received between Jan 18, 2017 and Mar 26, 2026 — classified as 1 injury, 105 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (75 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 2 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 4 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility75
  • Break26
  • Contamination /Decontamination Problem2
  • Crack2
  • Material Integrity Problem2
  • Mechanical Jam2
  • Appropriate Device Problem Term/Code Not Available1
  • Device Slipped1

Patient problems

  • No Clinical Signs, Symptoms or Conditions68
  • No Patient Involvement20
  • No Consequences Or Impact To Patient13
  • No Known Impact Or Consequence To Patient2
  • Not Applicable2
  • Insufficient Information1
  • No Code Available1

Recent adverse event reports

The 25 most recent of 106 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 26, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING STAFF IDENTIFIED THAT THE QUICK CONNECTION ON THE FRONT TIP OF THE DEVICE HAD BECOME STUCK AND COULD NO LONGER SUCCESSFULLY CONNECT WITH SCREWDRIVER SHAFTS AS INTENDED. THIS ITEM WAS DEEMED TO BE NO LONGER FUNCTIONAL AND WAS SUBSEQUENTLY DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24699986 ↗
  • MalfunctionMar 26, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT DURING MEDICAL DEVICE REPROCESSING STAFF IDENTIFIED THAT THE QUICK CONNECTION ON THE FRONT TIP OF THE DEVICE HAD BECOME STUCK AND COULD NO LONGER SUCCESSFULLY CONNECT WITH SCREWDRIVER SHAFTS AS INTENDED. THIS ITEM WAS DEEMED TO BE NO LONGER FUNCTIONAL AND WAS SUBSEQUENTLY DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24699979 ↗
  • MalfunctionMar 9, 2026Break; Device-Device Incompatibility

    IT WAS REPORTED THAT POSTOPERATIVELY IT WAS IDENTIFIED THAT THE DEVICE IS BROKEN AND THE HANDLE WON'T LATCH ONTO SCREWDRIVERS, THE SYNTHETIC WOOD CHIPPED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24545570 ↗
  • MalfunctionMar 9, 2026Break; Device-Device Incompatibility

    IT WAS REPORTED THAT POSTOPERATIVELY IT WAS IDENTIFIED THAT THE DEVICE IS BROKEN AND THE HANDLE WON'T LATCH ONTO SCREWDRIVERS, THE SYNTHETIC WOOD CHIPPED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24545563 ↗
  • MalfunctionDec 12, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2025, THE HANDLE NO LONGER GRIPS/RETAINS DRIVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23794683 ↗
  • MalfunctionDec 11, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON NOV 21, 2025, HANDLE WILL NOT HOLD DRIVER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23783271 ↗
  • MalfunctionJul 29, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE HANDLE WITH MINI QUICK COUPLING FROM THE LCP MODULAR MINI SET WILL NOT HOLD THE SCREWDRIVER SHAFT. THE COUPLING PART OF HANDLE WAS NOT WORKING. HANDLE WOULDN¿T HOLD THE THE POWER SCREWDRIVER. THERE WERE NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22641510 ↗
  • MalfunctionJul 23, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2025, THE SCREW DRIVERS AND ONE HANDLE ARE BROKEN, NEED REPLACEMENT ASAP. DRIVER TIPS ARE TWISTED AND HAVE DAMAGED THREADS. SCREW DRIVER HANDLE IS STUCK AND WILL NOT RELEASE. ITEMS ARE AVAILABLE FOR RETURN. NO DELAY OR PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22577332 ↗
  • MalfunctionJun 30, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE HANDLE NO LONGER LOCKS THE SCREW DRIVER IN PLACE PROPERLY OR TWISTS HANDLE PROPERLY TO TIGHTEN SCREW. ISSUE WAS IDENTIFIED POSTOPERATIVELY.

    Patient problems: Insufficient Information

    MAUDE report 22358018 ↗
  • MalfunctionMay 8, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2025, THE ITEM IS NOT FUNCTIONING PROPERLY. THE HANDLE WOULD NOT ACCEPT AND HOLD ONTO SCREWDRIVER TIPS. ISSUE WAS DISCOVERED DURING SURGERY; THERE WAS NO PATIENT INVOLVEMENT OR DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21986173 ↗
  • MalfunctionMay 6, 2025Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2025, THE REP WAS PROVIDING AN INSERVICE TO THE STAFF AND NOTICE THE SCREWDRIVER COUPLER WAS BROKEN. THERE WAS NO PATIENT STATUS/ OUTCOME / CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21969872 ↗
  • MalfunctionJan 16, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2025 THE DRIVER SHAFTS WILL NOT ENGAGE WITH QUICK CONNECT HANDLE. IT WAS FOUND POST-OP. THERE WAS NO PATIENT INVOLVEMENT OR DELAY IN CASE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21170231 ↗
  • MalfunctionJan 14, 2025Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE INSTRUMENTS BROKE IN DECONTAMINATION AND THE DAMAGE WAS NOTICED DURING INSPECTION. THE EVENT OCCURRED DURING STERILE PROCESSING, CALIBRATION AND FIELD INSPECTION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21150586 ↗
  • MalfunctionDec 31, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE METRO TRAY CAME BACK FROM THE CASE WITH THE SCREWDRIVER SHAFT WEDGED INTO THE HANDLE. IT IS NOT POSSIBLE TO REMOVE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21047920 ↗
  • MalfunctionDec 12, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2024 THE HANDLE STUCK TO DRIVESHAFT. THE EVENT OCCUR DURING STERILE PROCESSING AND FIELD INSPECTION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20920024 ↗
  • MalfunctionNov 21, 2024Device-Device Incompatibility; Illegible Information; Device-Device Incompatibility; Illegible Information

    IT WAS REPORTED THAT ON (B)(6) 2024, A HANDLE WITH MINI QUICK COUPLING WOULD NOT ENGAGE THE MINI SCREWDRIVERS IN THE MINI FRAG. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20751567 ↗
  • MalfunctionOct 22, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THE SCREWDRIVER HANDLES ARE LEAVING BROWN MARKS IN POUCHES AFTER STERILIZATION. THE HANDLES ARE CREATING DEBRIS. THE EVENT OCCURRED DURING FIELD INSPECTION AND STERILE PROCESSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20507490 ↗
  • MalfunctionAug 23, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE SCREWDRIVER HANDLE AND T6 DRIVER DID NOT ATTACH TOGETHER FOR SOME REASON. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT OUTCOME REPORTED. THIS REPORT INVOLVES ONE (1) HANDLE WITH MINI QUICK COUPLING. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20051496 ↗
  • MalfunctionAug 21, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT PATIENT UNDERWENT AN VETERINARY PROCEDURE ON (B)(6) 2024. DURING THE PROCEDURE, THE HANDLE WAS NOT TIGHTENING. BACK UP WAS AVAILABLE. SURGERY WAS COMPLETED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THIS REPORT IS FOR A HANDLE WITH MINI QUICK COUPLING. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20009918 ↗
  • MalfunctionAug 8, 2024Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2024, THE QC CONNECTION WAS BROKEN ON THE ITEM. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT. THIS REPORT INVOLVES ONE HANDLE WITH MINI QUICK COUPLING. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19939732 ↗
  • MalfunctionJun 12, 2024Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON UNKNOWN DATE THE HANDLE DOES NOT LATCH ONTO THE DRIVER. THE EVENT OCCUR DURING STERILE PROCESSING, FIELD INSPECTION AND CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT. NO OTHER INFORMATION WAS PROVIDED. THIS REPORT INVOLVES ONE (1) HANDLE WITH MINI QUICK COUPLING. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19524039 ↗
  • MalfunctionMar 5, 2024Break; Break

    IT WAS REPORTED THAT ON FEBRUARY 20, 2024 THE HANDLE IS BROKEN. THERE WAS NO PATIENT AND PROCEDURE INVOLVEMENT.THIS REPORT IS FOR ONE HANDLE WITH MINI QUICK COUPLING FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18839400 ↗
  • MalfunctionJan 19, 2024Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE DRIVER WAS BROKEN AND WAS UNABLE TO BE USED. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE HANDLE WITH MINI QUICK COUPLING. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18543378 ↗
  • MalfunctionSep 14, 2023Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, INSTRUMENT WAS REPORTED FOR REASON NOT SPECIFIED. THIS WAS A PRE OP CASE, SO THERE WAS NO PATIENT INVOLVEMENT. DURING THE SERVICE AND REPAIR EVALUATION IT WAS IDENTIFIED THAT THE INSTRUMENT COULD NOT HOLD THE BIT. THIS REPORT INVOLVES ONE (1) HANDLE WITH MINI QUICK COUPLING. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17745203 ↗
  • MalfunctionAug 31, 2023Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, WHILE RESETTING TRAYS THE SALES REP FOUND ISSUES WITH THE INSTRUMENTS IN QUESTION. THE TIP OF THE DEPTH GAUGE WAS FOUND TO BE BENT, THE QUICK CONNECT SCREWDRIVER WOULD NOT ENGAGE WITH DRIVER SHAFTS, AND THE QUICK CONNECT HANDLE WAS BROKEN. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS REPORT IS FOR A HANDLE WITH MINI QUICK COUPLING. THIS IS REPORT 1 OF 3 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17663740 ↗

FDA clearance and classification

The FDA regulates product code HWX (tap, bone) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 311.01

Has the NA 311.01 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10886982187192), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA 311.01?

The FDA's MAUDE database lists 106 adverse event reports for this device, received between Jan 18, 2017 and Mar 26, 2026. In those reports, the most common specific device problems are “Device-Device Incompatibility” (75), “Break” (26) and “Contamination /Decontamination Problem” (2); the most common patient problem is “Not Applicable” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 311.01?

None of the reports are classified as deaths, 1 as an injury and 105 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 311.01 FDA approved?

Devices of this type (product code HWX) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.

Who makes the NA 311.01, and is it still sold?

The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 311.01). Its record shows it as in commercial distribution.