NA 03.424.124Recalls & reported problems

Synthes GmbH · 2.4 VOLT(TM) TAP LOCKING SCREW TAP TO 90MM · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10886982335944), received between Feb 25, 2026 and Mar 18, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 6 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 25, 2026Class IIOpen, Classified

    A full lot swap between VOLT Locking Screw Tap 03.424.124 lot 68900P9 and Cortical Screw Tap 03.424.126 lot 68982P8, resulting in the bones taps having an incorrect thread.

    Z-1702-2026 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available7
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2026Material Integrity Problem; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT TAP Ø2.4 TAPPING UP TO 90 L130 WAS ASSOCIATED WITH RECALL. THE EVENT INVOLVED A SINGLE DEVICE FROM THIS LOT. THERE WAS NO REPORTED PATIENT CONSEQUENCE OR INVOLVEMENT, AND THE DEVICE WAS AVAILABLE FOR RETURN. THE COMPLAINT DEVICE WAS REMOVED FROM USE AS A PRECAUTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24628370 ↗
  • MalfunctionFeb 25, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED DURING A SET BUILDING PROCESS WHEN AN OPERATOR NOTICED THAT TWO PIECES WERE IDENTICAL BUT ETCHED WITH DIFFERENT PART NUMBERS. SPECIFICALLY, PARTS LABELED AS 03.424.124 WERE ETCHED ON 03.424.126 COMPONENTS, DESPITE DIFFERENCES IN THREAD SIZE. THERE WAS NO PATIENT INVOLVEMENT OR CLINICAL IMPACT, AS THE EVENT OCCURRED DURING SET-UP AND NOT DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442466 ↗
  • MalfunctionFeb 25, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED DURING A SET BUILDING PROCESS WHEN AN OPERATOR NOTICED THAT TWO PIECES WERE IDENTICAL BUT ETCHED WITH DIFFERENT PART NUMBERS. SPECIFICALLY, PARTS LABELED AS 03.424.124 WERE ETCHED ON 03.424.126 COMPONENTS, DESPITE DIFFERENCES IN THREAD SIZE. THERE WAS NO PATIENT INVOLVEMENT OR CLINICAL IMPACT, AS THE EVENT OCCURRED DURING SET-UP AND NOT DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442464 ↗
  • MalfunctionFeb 25, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED DURING A SET BUILDING PROCESS WHEN AN OPERATOR NOTICED THAT TWO PIECES WERE IDENTICAL BUT ETCHED WITH DIFFERENT PART NUMBERS. SPECIFICALLY, PARTS LABELED AS 03.424.124 WERE ETCHED ON 03.424.126 COMPONENTS, DESPITE DIFFERENCES IN THREAD SIZE. THERE WAS NO PATIENT INVOLVEMENT OR CLINICAL IMPACT, AS THE EVENT OCCURRED DURING SET-UP AND NOT DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442462 ↗
  • MalfunctionFeb 25, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED DURING A SET BUILDING PROCESS WHEN AN OPERATOR NOTICED THAT TWO PIECES WERE IDENTICAL BUT ETCHED WITH DIFFERENT PART NUMBERS. SPECIFICALLY, PARTS LABELED AS 03.424.124 WERE ETCHED ON 03.424.126 COMPONENTS, DESPITE DIFFERENCES IN THREAD SIZE. THERE WAS NO PATIENT INVOLVEMENT OR CLINICAL IMPACT, AS THE EVENT OCCURRED DURING SET-UP AND NOT DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442461 ↗
  • MalfunctionFeb 25, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED DURING A SET BUILDING PROCESS WHEN AN OPERATOR NOTICED THAT TWO PIECES WERE IDENTICAL BUT ETCHED WITH DIFFERENT PART NUMBERS. SPECIFICALLY, PARTS LABELED AS 03.424.124 WERE ETCHED ON 03.424.126 COMPONENTS, DESPITE DIFFERENCES IN THREAD SIZE. THERE WAS NO PATIENT INVOLVEMENT OR CLINICAL IMPACT, AS THE EVENT OCCURRED DURING SET-UP AND NOT DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442459 ↗
  • MalfunctionFeb 25, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED DURING A SET BUILDING PROCESS WHEN AN OPERATOR NOTICED THAT TWO PIECES WERE IDENTICAL BUT ETCHED WITH DIFFERENT PART NUMBERS. SPECIFICALLY, PARTS LABELED AS 03.424.124 WERE ETCHED ON 03.424.126 COMPONENTS, DESPITE DIFFERENCES IN THREAD SIZE. THERE WAS NO PATIENT INVOLVEMENT OR CLINICAL IMPACT, AS THE EVENT OCCURRED DURING SET-UP AND NOT DURING PATIENT USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24442457 ↗