Skip to content

Invictus 17100-18-1Recalls & reported problems

Alphatec Spine, Inc. · TROCAR TIP NITINOL GUIDEWIRE 18" · Class 1

All models of Invictus (96)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00190376220246), received between Nov 25, 2025 and Nov 25, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Material Separation1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 25, 2025Break; Material Separation; Break; Material Separation

    DURING ATTEMPTED SCREW PLACEMENT INTO THE L4/L5/S VERTEBRAL BODY, OSSIFICATION WAS ENCOUNTERED. PRIOR TO INSERTING THE SCREW, A TAP WAS ADVANCED OVER THE GUIDEWIRE. ON FLUOROSCOPY, THE GUIDEWIRE TIP APPEARED TO BE RETAINED WITHIN THE VERTEBRAL BODY. THE GUIDEWIRE WAS REMOVED FOR INSPECTION AND COMPARISON WITH A NEW GUIDEWIRE CONFIRMED THAT APPROXIMATELY 2MM OF THE TIP HAD SHEARED OFF. THE RETAINED FRAGMENT WAS EMBEDDED WITHIN THE OSSIFIED VERTEBRAL BODY AND COULD NOT BE SAFELY REMOVED. BASED ON THE SURGEON'S CLINICAL JUDGMENT, SCREW PLACEMENT PROCEEDED WITH THE GUIDEWIRE TIP LEFT IN SITU. BECAUSE THE FRAGMENT WAS SECURELY CONTAINED WITHIN THE OSSIFIED VERTEBRAL BODY, NO DISPLACEMENT WAS OBSERVED ON INTRAOPERATIVE OR POSTOPERATIVE IMAGING.

    Patient problems: Foreign Body In Patient

    MAUDE report 23647301 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page