Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00190446269106), received between May 6, 2021 and Jan 16, 2026 — classified as 4 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • No Apparent Adverse Event2
  • Detachment of Device or Device Component1
  • Difficult to Remove1
  • Material Deformation1
  • Mechanical Problem1
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • Foreign Body In Patient4
  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 16, 2026Detachment of Device or Device Component

    INSTRUMENT FAILURE - STEM INSERTER KNOB FOR SMALL SHELL BROKE DURING IMPACTING OF THE STEM, RESULTING IN THREADED END BREAKING OFF INTO THE STEM WITH IMPLANT. POLY WAS STILL ABLE TO BE PUT ON, LEAVING THE PIECE IN PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 24090870 ↗
  • InjuryAug 27, 2025Mechanical Problem; Mechanical Problem

    COMPLAINT - PATIENT WAS UNDERGOING A REVERSE SHOULDER ARTHROPLASTY. SURGEON WAS IMPACTING STEM WITH MALLET WHEN THE THREADED INSERTER BROKE OFF IN THE IMPLANT. THERE WAS A 20 MINUTE DELAY IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22909891 ↗
  • MalfunctionApr 8, 2024No Apparent Adverse Event; No Apparent Adverse Event

    INSTRUMENT FAILURE - INSTRUMENT BROKE, PIECELEFT IN PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 19068589 ↗
  • InjuryMar 5, 2024No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THERE WAS A 20 MINUTE DELAY IN SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 18838655 ↗
  • InjuryJun 1, 2023Material Deformation; Material Deformation

    INSTRUMENT FAILURE PART LEFT IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 17044425 ↗
  • MalfunctionDec 6, 2021Break

    INSTRUMENT FAILURE - THE TIP OF THE SCREW BROKE OFF WHEN INSERTING THE STEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12936843 ↗
  • MalfunctionOct 25, 2021Break; Difficult to Remove

    INSTRUMENT FAILURE : WHEN IMPACTING IN SMALL SHELL STEM, INSERTER KNOB BROKE. MORE SPECIFICALLY, THE THREADS THAT THREAD INTO STEM SNAPPED OFF. THEY COULDN'T BE REMOVED, SO WERE LEFT IN STEM. POLY WAS STILL ABLE TO FULLY SEAT INTO STEM.

    Patient problems: Foreign Body In Patient

    MAUDE report 12694743 ↗
  • MalfunctionMay 6, 2021Therapeutic or Diagnostic Output Failure

    INSTRUMENT FAILURE - DUE TO WHEN DISENGAGING THE STEM INSERTER FROM THE STEM, UPON REMOVAL, WE NOTICED SOME THREADS WERE STILL IN THE IMPLANT STEM. CASE WAS COMPLETED WITHOUT ANY HARM TO THE PATIENT OR TIMING. BROKEN INSTRUMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11782864 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →