Skip to content

DJO SURGICAL 804-06-318Recalls & reported problems

Encore Medical, L.P. · AltiVate Reverse Glenoid Cannulated Tap, 6.5mm · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00190446839200), received between Jul 25, 2025 and Jun 3, 2026 — classified as 5 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component4
  • Mechanical Problem4
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Foreign Body In Patient3

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 3, 2026Mechanical Problem

    REVISION SURGERY DUE TO DISTAL END OF TAP BROKE OFF IN PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25385996 ↗
  • MalfunctionFeb 25, 2026Detachment of Device or Device Component

    INSTRUMENT FAILURE - PATIENT HAD SIGNIFICANTLY SCLEROTIC BONE. SURGEON WAS ABLE TO SUCCESSFULLY ADVANCE THE CANNULATED TAP 30MM. WHEN ATTEMPTING TO BACK THE TAP OUT OF BONE, THE INSTRUMENT BROKE ALONG THE SKINNY GUIDE WIRE PORTION, BROKEN FRAGMENTED PIECES LEFT IN THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 24449313 ↗
  • InjuryFeb 6, 2026Detachment of Device or Device Component

    INSTRUMENT FAILURE - DISPOSABLE TAP BROKE IN PATIENTS BONE.

    Patient problems: Foreign Body In Patient

    MAUDE report 24285061 ↗
  • MalfunctionNov 4, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE CANNULATED TAP WAS ADVANCE OVER WIRE AND STARTED TO ADVANCE. THEN THE TAP STARTED SPINNING FREELY AND THE TIP OF THE TAP HAD BROKEN OFF INSIDE GLENOID. THE BROKEN PART WAS REMOVED BUT THIS CAUSED 20 MINUTES DELAY IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23472674 ↗
  • InjurySep 9, 2025Mechanical Problem; Mechanical Problem

    COMPLAINT - CANNULATED TAP BROKE OFF INTO GLENOID. THEN WHEN TRIED TO REMOVE BROKEN PART IT KEPT BREAKING INTO LIKE 15 PIECES SEEMED BRITTLE. THERE WAS A ONE AND HALF HOUR DELAY IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23012362 ↗
  • InjuryAug 19, 2025Mechanical Problem; Mechanical Problem

    COMPLAINT - SURGEON USED THE CANNULATED 6.5 TAP ON POWER OVER THE 2.4MM GUIDEWIRE. AS THE TAP WAS ADVANCED INTO THE BONE, WE HEARD A LOUD CRACK. THE ENTIRE TAP BROKE OFF IN THE PATIENT. THE PATIENT'S BONE WAS THE HARDEST BONE THE SURGEON HAS EXPERIENCED IN PRACTICE. THE SURGEON HAD TO USE A BROKEN SCREW REMOVAL SET TO REMOVE THE TAP. HE WAS PLANNING ON USING THE 6.5 WEDGED BASEPLATE, BUT THE HOLE BECAME TOO BIG DUE TO THE EXTRACTION. HE HAD TO USE AN 8.0 BASEPLATE, TAP, AND CENTRAL SCREW. THE FINAL OUTCOME WAS ACCEPTABLE. THERE WAS A 20-30 MINUTE DELAY IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22846553 ↗
  • InjuryJul 25, 2025Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component

    INSTRUMENT FAILURE: IT BROKE, BROKEN/FRAGMENTED PIECES LEFT IN PATIENT.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22620388 ↗

Other TAP, BONE devices

All TAP, BONE devices →

Sources and updates