DJO SURGICAL 804-06-310Recalls & reported problems
Encore Medical, L.P. · ALTIVATE REVERSE WEDGE GLENOID REAMER SLEEVE · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00190446843818), received between Jan 8, 2025 and Aug 7, 2025 — classified as 5 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (8 reports).
1 FDA recall record name this device's identifier, 1 still open.
7 other recalls cover devices cleared under the same FDA submission (K233481); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jul 18, 2025Class IIOpen, Classified
Their is a potential that the reamer may kick or bind up during or immediately prior to use.
Z-2548-2025 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Mechanical Problem8
- Adverse Event Without Identified Device or Use Problem4
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Bone Fracture(s)2
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INSTRUMENT FAILURE - REAMER BINDED WHILE WE WERE REAMING AND SURGEON BELIEVES IT CAUSED A CRACK IN THE GLENOID.
Patient problems: Bone Fracture(s)
MAUDE report 22745625 ↗InjuryJun 18, 2025Mechanical Problem; Mechanical Problem
COMPLAINT - THE REAMER WOULD BIND UP AND FREEZE DURING THE PROCEDURE. WE DISASSEMBLED AND CLEAN IT MULTIPLE TIMES AND IT CONTINUED TO BIND UP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22288618 ↗MalfunctionJun 13, 2025Mechanical Problem; Mechanical Problem
COMPLAINT - THE REAMERS WERE BINDING UP DUE TO BONE DEBRIS. THIS CAUSED A 15-20 MINUTE DELAY IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22207815 ↗MalfunctionJun 4, 2025Mechanical Problem; Mechanical Problem
COMPLAINT - IT WAS REPORTED THAT DURING OUR CASE WITH DR, WE WERE COMPLETING A REVERSE AUGMENTED GLENOID WHEN THE MAIN OFF AXIS REAMER AND ITS COMPONENTS SEEMINGLY BOUND UP ON EACH OTHER WHEN ATTEMPTING TO REAM FOR THE WEDGE BASEPLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22146567 ↗MalfunctionMay 21, 2025Mechanical Problem; Adverse Event Without Identified Device or Use Problem; Mechanical Problem; Adverse Event Without Identified Device or Use Problem
INSTRUMENT FAILURE - THE REAMER SEIZED UP WHILE REAMING, DESPITE ALL THE PRECAUTIONS BEING TAKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22066795 ↗MalfunctionMay 12, 2025Mechanical Problem; Mechanical Problem
INSTRUMENT FAILURE - THE REAMER SEIZED UP WHILE REAMING, DESPITE ALL THE PRECAUTIONS BEING TAKEN FRACTURING THE GLENOID DOWN THE MIDDLE.
Patient problems: Bone Fracture(s)
MAUDE report 22007089 ↗InjuryMay 7, 2025Mechanical Problem; Mechanical Problem
INSTRUMENT FAILURE - THE INTERNAL SHAFT ON THE WEDGE REAMER WAS BENT, WE PULLED RESET THE SECOND BOSS AND WEDGE REAMER AND STILL BINDING REAMER SHAFT, CAUSING A 30 MINUTE DELAY IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21980614 ↗InjuryMar 26, 2025Mechanical Problem; Mechanical Problem
INSTRUMENT FAILURE - WHEN THE DOCTOR WENT TO USE THE REAMER, THE REAMER WOULD GET BOUND UP BEFORE HE EVEN WAS ABLE TO PLACE IT IN THE PATIENT. WE OPENED UP ANOTHER SET TRIED WITH THE OTHER ONE AND IT DID THE SAME THING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21704302 ↗InjuryMar 26, 2025Mechanical Problem; Mechanical Problem
INSTRUMENT FAILURE - WHEN THE DOCTOR WENT TO USE THE REAMER, THE REAMER WOULD GET BOUND UP BEFORE HE EVEN WAS ABLE TO PLACE IT IN THE PATIENT. WE OPENED UP ANOTHER SET TRIED WITH THE OTHER ONE AND IT DID THE SAME THING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21704301 ↗MalfunctionMar 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INSTRUMENT FAILURE -REAMER WOULD NOT SPIN FREELY WITH FACING REAMER AND BOSS REAMER ENGAGE INTO IT. (IT BOUND), CAUSING A 25 MINUTE DELAY IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21670664 ↗InjuryMar 6, 2025No Apparent Adverse Event; No Apparent Adverse Event
PRIMARY SURGERY - THE OFFSET WEDGE REAMER ASSEMBLY LOCKED UP, THE INSTRUMENTS WERE DISASSEMBLED, CLEANED AND REASSEMBLED MULTIPLE TIME WITH THE SAME OUTCOME CAUSING A 20 MINUTE DELAY IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21542796 ↗MalfunctionJan 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INSTRUMENT FAILURE - THE REAMER SEIZED UP WHILE REAMING THE GLENOID DESPITE ALL THE PRECAUTIONARY STEPS BEING BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21108222 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K233481. One submission often covers many models, so check the product and lot details in each record.
- Z-2549-2025 · Jul 18, 2025 · Open, ClassifiedBrand Name: DJO SURGICAL Product Name: FA S Altivate Reverse Glenoid Tray Model/Catalog Number: None Software Version: NA Product Description: Enovis" shoulder devices are intended
- Z-2546-2025 · Jul 18, 2025 · Open, ClassifiedBrand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER HEAD Model/Catalog Number: 804-06-311 Product Description: Material: S.S/Nitronic, Non-Sterile
- Z-2547-2025 · Jul 18, 2025 · Open, ClassifiedBrand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER BOSS DRILL, 6.5mm Model/Catalog Number: 804-06-312 Product Description: Material: S.S., Non Sterile
- Z-2009-2025 · May 28, 2025 · Open, ClassifiedBrand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm Model/Catalog Number: 506-04-122 Software Version: NA Product Description: Encore Medical L.
- Z-2010-2025 · May 28, 2025 · Open, ClassifiedBrand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 18mm Model/Catalog Number: 506-04-118 Software Version: NA Product Description: Encore Medical L.
- Z-2008-2025 · May 28, 2025 · Open, ClassifiedBrand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm Model/Catalog Number: 506-04-126 Software Version: NA Product Description: Encore Medical L.
- Z-2011-2025 · May 28, 2025 · Open, ClassifiedBrand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 30mm Model/Catalog Number: 506-04-130 Software Version: NA Product Description: Encore Medical L.
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page