Skip to content

Brand Name: DJO SURGICAL recallZ-2548-2025 · Encore Medical, LP · initiated Jul 18, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Encore Medical, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Their is a potential that the reamer may kick or bind up during or immediately prior to use.

FDA root cause category: Device Design

What the company is doing

On July 29, 2025, Enovis issued a "Urgent Medical Device Recall Notification via Email to affected consignees. On August 4, 2025, Enovis issued an update to the recall notification via E-Mail requesting consignees rework affected trays and returned affected products. Enovis asked consignees to take the following actions: 1. Continue using the wedge reamer assembly while carefully inspecting and following the specific details in the supplied Quality Bulletin. 2. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 3. Please see the attached quality Bulletin 4. By 9/13/25, the firm will reach out regarding reworking and returning trays. 5. Acknowledge receipt of notification.

Recall details

Recall number
Z-2548-2025
Status
Open, Classified
Initiated by the company
Jul 18, 2025
Posted by the FDA
Sep 10, 2025
Recalling firm
Encore Medical, LP
Quantity
447 units
Product code
KWS
FDA submissions
K233481
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, FL, HI, IA, IL, IN, KS, KY, LA, MD, ME, NE, NJ, NY, OH, PR, RI, SC, TN, TX, UT, VA, WA, WI.

Product description

Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER SLEEVE Model/Catalog Number: 804-06-310 Product Description: Material: S.S/CoCrMo, Non-Sterile

Code information (models, lots, serial numbers)

Lot Code: All Lots GTIN: 00190446843818

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

20 devices cleared under K233481 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97412).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page