ENDOSKELETON® TC 5200-1001Recalls & reported problems

Medtronic Sofamor Danek, Inc. · Implant Holder · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00191375003793), received between Nov 28, 2025 and Sep 1, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem5
  • Break4
  • Failure to Align2

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 1, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50537535 ↗
  • MalfunctionApr 24, 2026Break

    INFORMATION WAS RECEIVED FROM THE HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT THE INSERTER SPRING WAS BROKEN AND THE AWL POINT WAS WORN DOWN DUE TO NORMAL USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24976964 ↗
  • MalfunctionApr 21, 2026Mechanical Problem; Failure to Align

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR ANTERIOR CERVICAL DISCECTOMY AND FUSION AND DIRECT LATERAL INTERBODY FUSION SPINAL THERAPY. IT WAS REPORTED THAT THE TC INSERTERS, THE INSERTER PLI 10 WAS REPORTED TO BE LOOSE ON THE HANDLE, THE INSTRUMENT PLI 20 COULD NOT BE DISASSEMBLED FOR CLEANING BECAUSE THE THREADS WERE STUCK. FOR THE DLIF INSERTER PLI 30, THE INSERTER THREADS WERE REPORTED TO BE CROSS-THREADED AND STUCK IN THE CANNULA. THERE WAS NO PATIENT INVOLVED. ADDITIONAL INFORMATION RECEIVED FROM THE MANUFACTURER REPRESENTATIVE THAT THE INSTRUMENTS ARE CROSS THREADED / CAN'T BE TAKEN APART FOR CLEANING NO PIECES BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24935542 ↗
  • MalfunctionMar 12, 2026Break; Failure to Align

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT THE INSERTER INNER SPRING BROKE OFF DURING CLEANING. FOR THE SECOND INSERTER, THERE WAS INABILITY TO UNTHREAD FOR CLEANING, AND FAILURE TO LATCH ONTO INTERBODY CAGES WHEN ATTEMPTED. THESE ISSUES WERE IDENTIFIED BY THE STERILE PROCESSING DEPARTMENT AT THE HOSPITAL PRIOR TO USE AND NOT DURING A PROCEDURE. NO PATIENT WAS INVOLVED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24584301 ↗
  • MalfunctionMar 11, 2026Break; Mechanical Problem

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR ANTERIOR CERVICAL DISCECTOMY AND FUSION SPINAL THERAPY AT LEVELS C3-5. IT WAS REPORTED THAT THE LOCKING MECHANISM ON THE SHAFT OF THE INSERTER FAILED. IT IS UNABLE TO ATTACH IMPLANTS IN ITS CURRENT CONDITION. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24574049 ↗
  • MalfunctionMar 11, 2026Break

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING POST-OP ANTERIOR CERVICAL DISCECTOMY AND FUSION. IT WAS REPORTED THAT INSERTER BROKE AND SPIRING LOST IN SPD UPON CLEANING THE INSTRUMENT. NO PATIENT WAS INVOLVED IN THIS EVENT. NO FURTHER COMPLICATIONS WERE REPORTED/ANTICIPATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24567785 ↗
  • MalfunctionDec 24, 2025Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA MANUFACTURING REPRESENTATIVE REGARDING A PRODUCT IDENTIFIED DURING AN EVENT. PROCEDURE INVOLVED IN THIS EVENT WAS ACDF (ANTERIOR CERVICAL DISCECTOMY AND FUSION). IT WAS REPORTED THAT THE INSERTER NO LONGER FUNCTIONS CORRECTLY. THE RELEASE MECHANISM WON'T FULLY RELEASE THE IMPLANT. THE SURGICAL TECH NOTICED THIS PRIOR TO USING THE INSTRUMENT. THE INSTRUMENT WAS REMOVED FROM THE FIELD. NO FURTHER COMPLICATIONS WERE REPORTED/ANTICIPATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23886556 ↗
  • MalfunctionNov 28, 2025Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING ANTERIOR CERVICAL DISCECTOMY AND FUSION SPINAL THERAPY WITH A PRE-OPERATIVE DIAGNOSIS OF C5-C6 HERNIATED NUCLEUS PULPOSUS. IT WAS REPORTED THAT THE IMPLANT DOESN'T RELEASE FROM INSERTER. THERE WERE NO PATIENT SYMPTOMS REPORTED. ADDITIONAL INFORMATION RECEIVED FROM THE MANUFACTURER REPRESENTATIVE THAT THE INSERTER WOULD NOT RELEASE THE IMPLANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23669302 ↗

Other DRIVER, PROSTHESIS devices

All DRIVER, PROSTHESIS devices →