NA M00505550Recalls & reported problems

Boston Scientific Corporation · Suture Cinch - Long · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
7
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00191506049188), received between Mar 3, 2026 and Aug 13, 2026 — classified as 7 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (8 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 3 reportsJun 2026: 2 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 14, 2026Open, Classified

    Device may fail to deploy the cinch implant and/or fail to cut the suture as intended.

    Z-2962-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Activation Failure8
  • Break8
  • Adverse Event Without Identified Device or Use Problem4
  • Failure to Cut4

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2026Break

    It was reported to Boston Scientific Corporation that a Suture Cinch - Long was used during an Endoscopic Sleeve Gastroplasty (ESG) procedure performed on (b)(6) 2026.During the procedure, the suture broke into 2 pieces off of the cinch catheter when applying tension before deploying. The procedure was completed with another Suture Cinch - Long of the same model.The patient's condition at the conclusion of the procedure was reported to be stable.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50244674 ↗
  • InjuryJul 27, 2026Break; Failure to Cut; Activation Failure

    It was reported to Boston Scientific that a Suture Cinch was utilized during an Endoscopic Sleeve Gastroplasty (ESG) procedure on (b)(6) 2026 for the treatment of obesity.During the procedure, the Suture Cinch handle did not deploy the cinch. The staff attempted to manually break the handle to manually deploy it however it did not work, the cinch did not cut the suture. The procedure was completed with another of the same device. There were no patient complications as a result of the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50003540 ↗
  • InjuryJul 6, 2026Break; Adverse Event Without Identified Device or Use Problem; Activation Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SUTURE CINCH WAS UTILIZED DURING AN ENDOSCOPIC SLEEVE GASTROPLASTY (ESG) PROCEDURE ON (B)(6) 2026. DURING THE PROCEDURE, THE HANDLE WAS CLOSED TO DEPLOY THE CINCH; HOWEVER, THE CINCH DID NOT DEPLOY. THE HANDLE BROKE, AND THE CINCH WIRE ALSO BROKE. THE NURSE USED A HEMOSTAT TO MANUALLY DEPLOY THE CINCH. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING THE ORIGINAL DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25719251 ↗
  • InjuryJun 29, 2026Failure to Cut; Adverse Event Without Identified Device or Use Problem; Activation Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SUTURE CINCH WAS UTILIZED DURING A GASTRIC PER-ORAL ENDOSCOPIC MYOTOMY (GPOEM) PROCEDURE ON (B)(6) 2026. DURING THE PROCEDURE, THE SUTURE CINCH HANDLE FAILED. THE PEEK PLUG AND COLLAR CAME TOGETHER; HOWEVER, THE HANDLE'S CUTTING FUNCTION COULD NOT BE ACTIVATED. AS A TROUBLESHOOTING MEASURE, THE USER MANUALLY FIRED THE DEVICE AND COMPLETED THE CUT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING THE ORIGINAL DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25650386 ↗
  • InjuryJun 2, 2026Break; Adverse Event Without Identified Device or Use Problem; Activation Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC THAT A SUTURE CINCH WAS UTILIZED DURING A COLONOSCOPY PROCEDURE ON (B)(6) 2026, FOR POLYP REMOVAL. DURING THE PROCEDURE, THE SUTURE CINCH WIRE BROKE WHILE THE TECHNICIAN WAS CLOSING THE HANDLE. THE STAFF HAD TO TROUBLESHOOT BY BREAKING THE HANDLE AND MANUALLY PULLING THE WIRE THE REST OF THE WAY WITH A CLAMP.PROCEDURE WAS COMPLETED WITH THE ORIGINAL DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25363980 ↗
  • InjuryMay 28, 2026Break; Adverse Event Without Identified Device or Use Problem; Activation Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SUTURING CINCH WAS USED IN AN ENDOSCOPIC SUBMUCOSAL DISSECTION (ESD) PROCEDURE ON (B)(6) 2026, FOR THE TREATMENT OF A COLONIC MASS. DURING THE PROCEDURE, THE SUTURING CINCH HANDLE SNAPPED UPON DEPLOYMENT AND THE CINCH WOULD NOT DEPLOY. TROUBLESHOOTING ATTEMPTS WERE MADE INCLUDING TRYING TO OPEN THE HANDLE AND JIGGLE IT. THE PHYSICIAN HAD TO CUT THE WIRE AND MANUALLY DEPLOY THE CINCH. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25321770 ↗
  • InjuryMay 26, 2026Break; Failure to Cut; Activation Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SUTURE CINCH WAS USED DURING A PROCEDURE ON (B)(6) 2026. DURING THE PROCEDURE, THE CINCH WIRE BROKE AND FAILED TO DEPLOY AND CUT THE SUTURE. THE PROCEDURE WAS COMPLETED WHEN THE PHYSICIAN MANUALLY DEPLOYED THE CINCH. THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25286025 ↗
  • InjuryMay 25, 2026Break; Failure to Cut; Activation Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SUTURE CINCH WAS USED DURING A PROCEDURE ON (B)(6) 2026. DURING THE PROCEDURE, THE CINCH WIRE BROKE AND FAILED TO DEPLOY AND CUT THE SUTURE. THE PROCEDURE WAS COMPLETED WHEN THE PHYSICIAN MANUALLY DEPLOYED THE CINCH. THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25272645 ↗
  • MalfunctionMar 3, 2026Break; Activation Failure

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A SUTURE CINCH - LONG WAS USED DURING AN ENDOSCOPIC MUCOSAL RESECTION (EMR) PROCEDURE ON (B)(6) 2025. DURING THE PROCEDURE, THE FOURTH HELIX WAS STUCK IN THE SHEATH AND FAILED TO DEPLOY. THE CINCH ALSO FAILED TO DEPLOY AND THE WIRE BENT. IT IS UNKNOWN IF THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24493268 ↗

Other Endoscopic tissue approximation device devices

All Endoscopic tissue approximation device devices →